Define magnification and resolution, and give the resolution limits of the human eye and light microscope.
Magnification: how much an object can be enlarged (Total magnification = eyepiece magnification × objective lens magnification). Resolution: the ability to distinguish between two closely spaced objects. Minimum resolution of the human eye: ~0.2 mm; limit of resolution of a standard light microscope: ~0.2 µm.
Outline the basic optical path and components of a compound light microscope.
Light source -> condenser lens (focuses light onto the specimen) -> specimen -> objective lens -> eyepiece (ocular lens) -> retina / detector.
Describe how dead tissue specimens are prepared and stained for bright-field light microscopy.
Tissues are fixed in formalin and embedded in paraffin wax (or frozen in a block) to make them rigid, cut into 1–10 µm sections with a microtome, mounted on a glass slide, and stained (e.g., H&E stain or enzyme-linked immunohistochemistry) to reduce light amplitude and create contrast.
Describe how living cells are observed without staining under a light microscope.
Phase contrast or Differential Interference Contrast (DIC) microscopy alters the microscope optics to shift the phase of light passing through transparent cells, converting phase differences into amplitude changes (creating contrast without killing/staining cells; often combined with timelapse digital video imaging).
Explain the principle of fluorescence and the optical layout of a fluorescence microscope.
A fluorophore absorbs excitation light at a shorter, higher-energy wavelength (e.g., 488/588 nm), undergoes non-radioactive decay, and emits a photon at a longer, lower-energy wavelength (e.g., 525 nm). Optics: light source -> excitation filter (selects 1 wavelength) -> dichroic mirror (reflects excitation light down through objective onto specimen, while allowing longer emitted fluorescence to pass straight up) -> emission filter (cuts out unwanted light) -> ocular/detector.
List the three main sources of fluorescence used to label cells.
Describe different methods for improving image collection and resolution in light microscopy.
Computer software image enhancement; Confocal microscopy (uses a laser and pinhole apertures to reject out-of-focus light, capturing sharp optical slices for 3D reconstruction); TIRF (indirect excitation near surfaces); Atomic Force Microscopy (tip measures surface forces); Super-resolution fluorescence microscopy (bypasses light diffraction limit down to ~20 nm resolution).
Outline the basic structure and preparation requirements of a Transmission Electron Microscope (TEM).
Uses an electron gun, magnetic condenser lens, ultra-thin dehydrated specimen section (cannot view living cells), magnetic objective lens, and projector lens onto a viewing screen or photographic film.
Outline the basic structure and sample preparation of a Scanning Electron Microscope (SEM).
Uses an electron gun, condenser lens, beam deflector + scan generator (scans beam across specimen), and objective lens; a detector collects electrons scattered from the surface of whole cells/small organisms (usually coated in a heavy metal like gold) to form an image on a video screen.
Compare Light Microscopy (LM), Transmission EM (TEM), and Scanning EM (SEM) by resolution and image type.
LM: limit of resolution ~0.2 µm; views live or fixed/stained 2D sections or fluorescent 3D stacks (colour/greyscale). TEM: electron wavelength ~0.004 nm (vastly higher resolution than LM); produces high-resolution 2D greyscale cross-sections of internal ultrastructure (e.g., RER, mitochondria) in dead, ultra-thin dehydrated slices. SEM: limit of resolution ~10 nm; produces detailed 3D-like greyscale surface topography images of whole cells or organisms.
Describe specialised electron microscopy preparation techniques (shadowing, freeze-fracture, negative staining, cryo-EM