exposure
number of ionizations in air
how is exposure measured?
Coulombs per kilogram (C/kg)
air kerma
energy of ionizations in air
how is air kerma measured?
gray (or joules/kg)
what is air kerma useful for?
measuring tube output, leakage radiation and scatter
absorbed dose
the amount of energy absorbed per unit mass
how is absorbed dose measured?
gray (or joules/kg)
what is absorbed dose useful for?
predicting short term radiation injuries
equivalent Dose (EqD)
takes into consideration the type of ionizing radiation that was absorbed
How is equivalent dose determined?
absorbed dose x Wr (weight factor)
what is equivalent Dose measured in?
sieverts
effective dose
takes into consideration the relative radiation sensitivities of different tissue types
what is effective dose used for?
to predict the risk of long term radiation induced injuries
what is effective dose measured in?
sieverts
whole body exposure limits
5 rem/yr (50 mSv)
who is personnel monitoring provided to?
workers who accumulate 1/10 of recommended dose limit
where should devices be worn?
collar, waist, or ring
dosimetry
measurement of ionizing radiation doses to personnel
what detector is used in OSL badges?
aluminium oxide
how is an OSL processed?
using laser light to cause luminescence
disadvantage of OSL
expensive
what detector is used in TLD badges?
lithium fluoride crystals
how are TLDs processed?
released stored energy when heated as visible light
disadvantage of TLDs
sensitive to heat
disadvantage of pocket dosimeters
poor accuracy, must be recharged
purpose of dosimetry report
provide personnel with regular feedback reguarding protection practices
primary barriers
protect against direct exposure to the primary beam
secondary barriers
scatter and leakage radiation
cardinal principles of radiation control
time, distance, shielding
beam limitation
1/2 inch of collimation, field size never larger than IR
how can we minimize dose w technique?
highest kVp, lowest mAs
filteration
removes low energy photons
where is a filter placed?
between tube and patient
where are grids placed?
between the patient and IR
what grid ratio should be used?
the lowest possible
when is the fetus most radiosensitive?
8-15 weeks of gestation
pregnancy dose limit
0.5 rem if declared