94 Terms

1
New cards

arterial, venous, capillary, arterial lines (art-lines/A-lines)

4 specimen types for blood gas analysis

2
New cards

radial, ulnar, femoral, brachial, temporal

What are the 5 arterial sites of collection?

3
New cards

radial artery

For blood gas analysis, what is the recommended site of collection?

4
New cards

temporal artery

For blood gas analysis in children, what is the recommended site of collection?

5
New cards

heparinized whole blood

For blood gas analysis, what is the recommended type of specimen?

6
New cards

heparin (green top / heparinized syringe / dry base heparin (DBH))

For blood gas analysis, what is the recommended anticoagulant that should be used?

7
New cards

patient is septic

An arterial specimen should be used for CO2 and O2 measurements, except for what specific case?

8
New cards

venous

If a patient is septic, what is the recommended site for blood draw?

9
New cards

lactic acid

if patient is septic, use venous blood because you'll want to correlate O2 levels with __________

10
New cards

capillary

mix of arterial and venous blood (oxy/deoxy)

11
New cards

bright red

Arterial vs Venous

Color of arterial blood

12
New cards

dark maroon

Arterial vs Venous

Color of venous blood

13
New cards

arterial; deeper

Arterial vs Venous

Which is more difficult to collect and why?

14
New cards

arterial

Arterial vs Venous

Has pulse

More painful due to more nerves

Prolonged bleeding

Paresis (temp loss of sensation)

15
New cards

Arterial blood sample

pH: increased

PO2: increased

PCO2: decreased

16
New cards

Venous blood sample

pH: decreased

PO2: decreased

PCO2: increased

17
New cards

Allen's Test

Perfusion check:

- fist for 30 sec

- pressure on radial and ulnar artery

- let go of ulnar artery, open fist

- perfusion in 8 sec = good site for BGA

<p>Perfusion check:</p><p>- fist for 30 sec</p><p>- pressure on radial and ulnar artery</p><p>- let go of ulnar artery, open fist</p><p>- perfusion in 8 sec = good site for BGA</p>
18
New cards

pre-analytical errors

- ordering and ID patient

- anticoagulant used

- mixing

- air bubbles

- exposure to air

- dilution

- time

- temperature

- transport

19
New cards

increase O2

decrease CO2

increase pH

How are the O2, CO2 and pH affected by bubbles in a sample?

20
New cards

increase O2 (rushes in)

decrease CO2 (rushes out)

increase pH

How are the O2, CO2 and pH affected by air exposure?

21
New cards

specimen can be diluted with infusion fluid if taken out of a port improperly

When might a blood gas specimen be diluted?

22
New cards

draw 2 tubes; discard tube 1 & test tube 2

How can we avoid a diluted sample when drawing from an arterial line?

23
New cards

room temp within 15 min

If oxygen or oxygen sat is measured, what is ideal temp and time of analysis?

24
New cards

stable for 30 min at RT

If oxygen or oxygen sat are NOT measured, what is temp and time of analysis?

25
New cards

ice water for 1 hour

If there is a delay in testing (>30 min), what is temp and time for analysis?

26
New cards

ice water slows glycolysis (slows production of CO2)

How does ice water affect blood gas samples?

27
New cards

speeds up glycolysis (increases CO2 / decreases O2 & pH)

How does warm temps affect blood gas samples?

28
New cards

hand-delivered

How are blood gas samples transported to the lab?

29
New cards

reject & request new sample

The lab receives an arterial blood sample at 0800 for blood gas testing for an ICU patient. Sample has been delivered to the lab in a plastic syringe, on ice. Sampling time was 0615. Sample has visible sedimentation. What would you do?

30
New cards

reject & request new sample

The lab receives an arterial blood sample from the emergency department. Blood gas testing is requested. The sample was transported by pneumatic tube within 10 minutes from sample collection. You notice an air bubble in the syringe. What would you do?

31
New cards

increase pH, decrease pCO2 , increase pO2

Upon receiving a room temperature arterial specimen for blood gas analysis collected 25 minutes ago, you observe that the cap on the syringe is missing. You reject the specimen, and base your decision on the premise that the following parameters have been altered:

32
New cards

analytical errors

- improper or lack of maintenance

- inadequate reagents

- poor or lack of calibration

- failure to perform QC or recognize QC problems

- non-instrument aspects (injections of a short sample)

- failure to interpret instrument flags or warnings that something may be wrong

33
New cards

postanalytical errors

involves all aspects of reporting pt results

- poor turn-around-times

- transcription mistakes

- reporting results without age-adjusted reference ranges, correct units, or rounding of data

- reporting results on wrong patient

- failure to interpret instrument flags or warnings that something may be wrong

34
New cards

pH, PCO2, PO2

electrolytes

total hemoglobin concentration

Which components are most commonly measured by blood gas analyzers

35
New cards

Electrodes

measures pH, PCO2, and PO2

36
New cards

Ion-Selective Electrodes (ISE)

measures electrolytes (associated with acid/base balance)

37
New cards

spectrophotmeter

measures total hemoglobin concentration (and bilirubin)

38
New cards

Bicarbonate

total CO2

total O2 content

O2 saturation

base excess

Which components are most commonly calculated by blood gas analyzers

39
New cards

80-110 mmHg

PO2 reference range

40
New cards

35-45 mmHg

PCO2 reference range

41
New cards

7.35-7.45

pH reference range

42
New cards

94-100%

SaO2 (saturated O2) reference range

43
New cards

22-26 mEq/L

HCO3- (bicarbonate) reference range

44
New cards

decreases 1 mmHg/year

What happens to blood gas parameters as we age (60+)

45
New cards

potentiometric reaction

What type of reaction is used by the blood gas analyzer to measure pH?

46
New cards

measures potential generated

What does pH reaction measure?

47
New cards

glass membrane that selects for H+

Indicator electrode membrane for pH

48
New cards

made of calomel or Ag/AgCl with saturated AgCl salt bridge

Reference electrode membrane for pH

49
New cards

potentiometric reaction

What type of reaction is used by the blood gas analyzer to measure PCO2?

50
New cards

measures potential generated

What does PCO2 reaction measure?

51
New cards

compound membrane (modified pH electrode that selects for CO2 gas)

Indicator electrode membrane for PCO2

52
New cards

outside is a Teflon or silicon semi-permeable membrane that lets CO2 in

What is the outside indicator electrode membrane for PCO2?

53
New cards

inside is a H2O/bicarb buffer that produces H+

What is the inside indicator electrode membrane for PCO2?

54
New cards

amperometric reaction

What type of reaction is used by the blood gas analyzer to measure PO2?

55
New cards

measures change in current

What does PO2 reaction measure?

56
New cards

Polarographic Clarke electrode with a semipermeable membrane that allows O2 to diffuse through

Indicator electrode membrane for PO2

57
New cards

silver wire

anode of Clarke electrode (PO2)

58
New cards

platnium wire

cathode of Clarke electrode (PO2)

59
New cards

reduced

When O2 diffuses through, it is (oxidized / reduced) at the cathode in a AgCl/phosphate buffer solution

60
New cards

SO2

calculated parameter derived from the measurement of oxyhemoglobin vs total hemoglobin by spectrophotometric methods using a cooximeter

OxyHgb / Total Hgb

61
New cards

interferences with Electrode measurement of blood gasses

1. carryover between specimens; avoided with wash step

2. pH electrode, and some other gases needs to be maintained at 37C

3. excess protein build-up on pH electrode/membrane will decrease sensitivity

62
New cards

Proteolytic enzymes (breakdown proteins)

Replace membrane

Using disposable membranes

Periodic cleaning

4 actions taken to reduce interference in electrode measurement of blood gasses

63
New cards

base excess

the amount of H+ ions that would be required to return the pH of blood to 7.34 if the PCO2 were adjusted to normal

calculated figure which provides an estimate of the metabolic component of the acid-base (CO2-HCO3-) balance

64
New cards

bicarbonate concentration and pH

What two things is base excess based on?

65
New cards

positive base excess

the addition of a base that raises the buffer content, resulting in a high pH (too alkaline) and metabolic alkalosis

66
New cards

negative base excess

a decreased base from either not enough bicarbonate or too much acid in the blood, resulting in a low pH (acidic) and metabolic acidosis

67
New cards

Anion Gap

(Na+ + K+) - (Cl- + HCO3-)

68
New cards

increased

Too much acid results in a (increased / normal / decreased) anion gap

69
New cards

normal; chloride would have been swapped in

Bicarbonate leaked out results in a (increased / normal / decreased) anion gap

70
New cards

carboxyhemoglobin

oxyhemoglobin

deoxyhemoglobin

methemoglobin

A cooximeter is a type of narrow bandpass spectrophotometer that is capable of measuring what 4 hemoglobins simultaneously?

71
New cards

A hemolyzing agent is added to release all hgb then each of the 4 hgbs are measured at their specific wavelength using a calibration curve

How does a cooximeter work?

72
New cards

sodium hydrosulfite

What is the reducing agent utilized in the cooximeter?

73
New cards

other parameters measured by a Cooximeter

ctHb (Concentration of total Hgb)

sO2 (Oxygen Saturation)

Fractions of each type of hgb, along with HgbF

Concentration of total bilirubin (conj & unconj)

74
New cards

Pulse Oximetry

determines the O2 content in the blood

light passes through a site on the body and spectrophotometric readings are taken in the red visible light and infrared spectrums

75
New cards

oxyhemoglobin

Pulse Oximetry

absorbs more infrared / allows more red light to pass through

76
New cards

deoxyhemoglobin

Pulse Oximetry

absorbs more red light / allows more infrared light to pass through

77
New cards

non-invasive, inexpensive, and compact

detects hypoxemia earlier than visual detention of cyanosis

Advantages (2) of Pulse Oximeters

78
New cards

not as reliable as ABG measurement

Disadvantage of Pulse Oximeters

79
New cards

arterial blood gas (ABG) test

preferred method for measuring O2 in your blood

80
New cards

1. Barometric pressure checks

2. Temperature checks

2 important maintenance checks for ABG instrumentation

81
New cards

internal barometer

Barometric pressure check

many BGA have an __________ that is used to correct for changes in barometric pressure

barometer should be checked for accuracy on a periodic basis

82
New cards

temperature dependent

Temperature checks

accurate measurement of ABGs is __________

acceptable tolerance for temperatures: +/- 0.1-0.2C

83
New cards

commercially prepared CO2 gasses at two different concentrations

What two calibrators are used for the 2-point calibration of PCO2 in the ABG analyzer?

84
New cards

"no atmospheric" air as one calibrator and "room air" as the other calibrator

What two calibrators are used for the 2-point calibration of PO2 in the ABG analyzer?

85
New cards

uses two electrolyte buffer solutions of known composition and pH, usually at 6.8 and 7.4 both at 37C

What two calibrators are used for the 2-point calibration of pH in the ABG analyzer?

86
New cards

calibration should occur at least every 6 months per CLIA

How often does calibration occur for an ABG analyzer?

87
New cards

1. Two-point calibration every 8hrs

2. One-point calibration every 4hrs

3. PO2 and PCO2 - should be initiated every 30 min (automated)

Blood gasses are so unstable that calibration must occur (3):

88
New cards

calilbration verification

required to confirm calibration & linearity of pH, PO2, PCO2

materials purchased form a reliable manufacture & usually consists of 4-5+ vials of assayed material

89
New cards

acidosis

alkalosis

normal

What are the 3 levels of quality control used for blood gas analysis?

90
New cards

aqueous buffered fluids

blood-based or fluorocarbon-based

tonometer

What are the QC materials for ABG made of/from? (3)

91
New cards

aqueous buffered fluids

QC material that contains a known gas mixture appropriate for each level of control

- contained in sealed ampules/vials

- before injection, vials need to be mixed well to make homogenous solution

92
New cards

blood-based or fluorocarbon-based control

QC material that mimics human whole blood better than aqueous based

- most preferred

- more expensive

93
New cards

tonometer

QC material that is used to prepare whole-blood controls

- equilibrates whole-blood samples using known gas mixtures

94
New cards

1. Cost of tonometer

2. Time required to achieve equilibrium (hours)

3. Inability to control pH measurement with tonometered blood samples

3 disadvantages of using a tonometer for blood gas QC