1/93
Looks like no tags are added yet.
arterial, venous, capillary, arterial lines (art-lines/A-lines)
4 specimen types for blood gas analysis
radial, ulnar, femoral, brachial, temporal
What are the 5 arterial sites of collection?
radial artery
For blood gas analysis, what is the recommended site of collection?
temporal artery
For blood gas analysis in children, what is the recommended site of collection?
heparinized whole blood
For blood gas analysis, what is the recommended type of specimen?
heparin (green top / heparinized syringe / dry base heparin (DBH))
For blood gas analysis, what is the recommended anticoagulant that should be used?
patient is septic
An arterial specimen should be used for CO2 and O2 measurements, except for what specific case?
venous
If a patient is septic, what is the recommended site for blood draw?
lactic acid
if patient is septic, use venous blood because you'll want to correlate O2 levels with __________
capillary
mix of arterial and venous blood (oxy/deoxy)
bright red
Arterial vs Venous
Color of arterial blood
dark maroon
Arterial vs Venous
Color of venous blood
arterial; deeper
Arterial vs Venous
Which is more difficult to collect and why?
arterial
Arterial vs Venous
Has pulse
More painful due to more nerves
Prolonged bleeding
Paresis (temp loss of sensation)
Arterial blood sample
pH: increased
PO2: increased
PCO2: decreased
Venous blood sample
pH: decreased
PO2: decreased
PCO2: increased
Allen's Test
Perfusion check:
- fist for 30 sec
- pressure on radial and ulnar artery
- let go of ulnar artery, open fist
- perfusion in 8 sec = good site for BGA

pre-analytical errors
- ordering and ID patient
- anticoagulant used
- mixing
- air bubbles
- exposure to air
- dilution
- time
- temperature
- transport
increase O2
decrease CO2
increase pH
How are the O2, CO2 and pH affected by bubbles in a sample?
increase O2 (rushes in)
decrease CO2 (rushes out)
increase pH
How are the O2, CO2 and pH affected by air exposure?
specimen can be diluted with infusion fluid if taken out of a port improperly
When might a blood gas specimen be diluted?
draw 2 tubes; discard tube 1 & test tube 2
How can we avoid a diluted sample when drawing from an arterial line?
room temp within 15 min
If oxygen or oxygen sat is measured, what is ideal temp and time of analysis?
stable for 30 min at RT
If oxygen or oxygen sat are NOT measured, what is temp and time of analysis?
ice water for 1 hour
If there is a delay in testing (>30 min), what is temp and time for analysis?
ice water slows glycolysis (slows production of CO2)
How does ice water affect blood gas samples?
speeds up glycolysis (increases CO2 / decreases O2 & pH)
How does warm temps affect blood gas samples?
hand-delivered
How are blood gas samples transported to the lab?
reject & request new sample
The lab receives an arterial blood sample at 0800 for blood gas testing for an ICU patient. Sample has been delivered to the lab in a plastic syringe, on ice. Sampling time was 0615. Sample has visible sedimentation. What would you do?
reject & request new sample
The lab receives an arterial blood sample from the emergency department. Blood gas testing is requested. The sample was transported by pneumatic tube within 10 minutes from sample collection. You notice an air bubble in the syringe. What would you do?
increase pH, decrease pCO2 , increase pO2
Upon receiving a room temperature arterial specimen for blood gas analysis collected 25 minutes ago, you observe that the cap on the syringe is missing. You reject the specimen, and base your decision on the premise that the following parameters have been altered:
analytical errors
- improper or lack of maintenance
- inadequate reagents
- poor or lack of calibration
- failure to perform QC or recognize QC problems
- non-instrument aspects (injections of a short sample)
- failure to interpret instrument flags or warnings that something may be wrong
postanalytical errors
involves all aspects of reporting pt results
- poor turn-around-times
- transcription mistakes
- reporting results without age-adjusted reference ranges, correct units, or rounding of data
- reporting results on wrong patient
- failure to interpret instrument flags or warnings that something may be wrong
pH, PCO2, PO2
electrolytes
total hemoglobin concentration
Which components are most commonly measured by blood gas analyzers
Electrodes
measures pH, PCO2, and PO2
Ion-Selective Electrodes (ISE)
measures electrolytes (associated with acid/base balance)
spectrophotmeter
measures total hemoglobin concentration (and bilirubin)
Bicarbonate
total CO2
total O2 content
O2 saturation
base excess
Which components are most commonly calculated by blood gas analyzers
80-110 mmHg
PO2 reference range
35-45 mmHg
PCO2 reference range
7.35-7.45
pH reference range
94-100%
SaO2 (saturated O2) reference range
22-26 mEq/L
HCO3- (bicarbonate) reference range
decreases 1 mmHg/year
What happens to blood gas parameters as we age (60+)
potentiometric reaction
What type of reaction is used by the blood gas analyzer to measure pH?
measures potential generated
What does pH reaction measure?
glass membrane that selects for H+
Indicator electrode membrane for pH
made of calomel or Ag/AgCl with saturated AgCl salt bridge
Reference electrode membrane for pH
potentiometric reaction
What type of reaction is used by the blood gas analyzer to measure PCO2?
measures potential generated
What does PCO2 reaction measure?
compound membrane (modified pH electrode that selects for CO2 gas)
Indicator electrode membrane for PCO2
outside is a Teflon or silicon semi-permeable membrane that lets CO2 in
What is the outside indicator electrode membrane for PCO2?
inside is a H2O/bicarb buffer that produces H+
What is the inside indicator electrode membrane for PCO2?
amperometric reaction
What type of reaction is used by the blood gas analyzer to measure PO2?
measures change in current
What does PO2 reaction measure?
Polarographic Clarke electrode with a semipermeable membrane that allows O2 to diffuse through
Indicator electrode membrane for PO2
silver wire
anode of Clarke electrode (PO2)
platnium wire
cathode of Clarke electrode (PO2)
reduced
When O2 diffuses through, it is (oxidized / reduced) at the cathode in a AgCl/phosphate buffer solution
SO2
calculated parameter derived from the measurement of oxyhemoglobin vs total hemoglobin by spectrophotometric methods using a cooximeter
OxyHgb / Total Hgb
interferences with Electrode measurement of blood gasses
1. carryover between specimens; avoided with wash step
2. pH electrode, and some other gases needs to be maintained at 37C
3. excess protein build-up on pH electrode/membrane will decrease sensitivity
Proteolytic enzymes (breakdown proteins)
Replace membrane
Using disposable membranes
Periodic cleaning
4 actions taken to reduce interference in electrode measurement of blood gasses
base excess
the amount of H+ ions that would be required to return the pH of blood to 7.34 if the PCO2 were adjusted to normal
calculated figure which provides an estimate of the metabolic component of the acid-base (CO2-HCO3-) balance
bicarbonate concentration and pH
What two things is base excess based on?
positive base excess
the addition of a base that raises the buffer content, resulting in a high pH (too alkaline) and metabolic alkalosis
negative base excess
a decreased base from either not enough bicarbonate or too much acid in the blood, resulting in a low pH (acidic) and metabolic acidosis
Anion Gap
(Na+ + K+) - (Cl- + HCO3-)
increased
Too much acid results in a (increased / normal / decreased) anion gap
normal; chloride would have been swapped in
Bicarbonate leaked out results in a (increased / normal / decreased) anion gap
carboxyhemoglobin
oxyhemoglobin
deoxyhemoglobin
methemoglobin
A cooximeter is a type of narrow bandpass spectrophotometer that is capable of measuring what 4 hemoglobins simultaneously?
A hemolyzing agent is added to release all hgb then each of the 4 hgbs are measured at their specific wavelength using a calibration curve
How does a cooximeter work?
sodium hydrosulfite
What is the reducing agent utilized in the cooximeter?
other parameters measured by a Cooximeter
ctHb (Concentration of total Hgb)
sO2 (Oxygen Saturation)
Fractions of each type of hgb, along with HgbF
Concentration of total bilirubin (conj & unconj)
Pulse Oximetry
determines the O2 content in the blood
light passes through a site on the body and spectrophotometric readings are taken in the red visible light and infrared spectrums
oxyhemoglobin
Pulse Oximetry
absorbs more infrared / allows more red light to pass through
deoxyhemoglobin
Pulse Oximetry
absorbs more red light / allows more infrared light to pass through
non-invasive, inexpensive, and compact
detects hypoxemia earlier than visual detention of cyanosis
Advantages (2) of Pulse Oximeters
not as reliable as ABG measurement
Disadvantage of Pulse Oximeters
arterial blood gas (ABG) test
preferred method for measuring O2 in your blood
1. Barometric pressure checks
2. Temperature checks
2 important maintenance checks for ABG instrumentation
internal barometer
Barometric pressure check
many BGA have an __________ that is used to correct for changes in barometric pressure
barometer should be checked for accuracy on a periodic basis
temperature dependent
Temperature checks
accurate measurement of ABGs is __________
acceptable tolerance for temperatures: +/- 0.1-0.2C
commercially prepared CO2 gasses at two different concentrations
What two calibrators are used for the 2-point calibration of PCO2 in the ABG analyzer?
"no atmospheric" air as one calibrator and "room air" as the other calibrator
What two calibrators are used for the 2-point calibration of PO2 in the ABG analyzer?
uses two electrolyte buffer solutions of known composition and pH, usually at 6.8 and 7.4 both at 37C
What two calibrators are used for the 2-point calibration of pH in the ABG analyzer?
calibration should occur at least every 6 months per CLIA
How often does calibration occur for an ABG analyzer?
1. Two-point calibration every 8hrs
2. One-point calibration every 4hrs
3. PO2 and PCO2 - should be initiated every 30 min (automated)
Blood gasses are so unstable that calibration must occur (3):
calilbration verification
required to confirm calibration & linearity of pH, PO2, PCO2
materials purchased form a reliable manufacture & usually consists of 4-5+ vials of assayed material
acidosis
alkalosis
normal
What are the 3 levels of quality control used for blood gas analysis?
aqueous buffered fluids
blood-based or fluorocarbon-based
tonometer
What are the QC materials for ABG made of/from? (3)
aqueous buffered fluids
QC material that contains a known gas mixture appropriate for each level of control
- contained in sealed ampules/vials
- before injection, vials need to be mixed well to make homogenous solution
blood-based or fluorocarbon-based control
QC material that mimics human whole blood better than aqueous based
- most preferred
- more expensive
tonometer
QC material that is used to prepare whole-blood controls
- equilibrates whole-blood samples using known gas mixtures
1. Cost of tonometer
2. Time required to achieve equilibrium (hours)
3. Inability to control pH measurement with tonometered blood samples
3 disadvantages of using a tonometer for blood gas QC