Lecture 9 Resolution and instrumentation

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/77

encourage image

There's no tags or description

Looks like no tags are added yet.

78 Terms

1
New cards

The 3 aspects of resolution

detail, contrast, and temporal

2
New cards

Detail resolution

The ability to distinguish the presence of 2 structures that lie close to one another

3
New cards

3 types of detail resolution

axial resolution, lateral resolution, and elevational resolution

4
New cards

What is poor detail resolution

if 2 reflectors are not separated sufficiently, they produce overlapping (not distinct) echoes. If separate echoes are not generated, the reflectors will be merged together on the display.

5
New cards

What is better detail resolution

the lower, the better

6
New cards

Axial resolution

The minimum reflector separation required along the direction of the sound traveled to produce 2 separate echoes(reflections)

7
New cards

What does axial resolution decide?

It determines how close together (front to back) two separate interfaces can be before they can no longer be identified as two separate echoes.

8
New cards

The equation of axial resolution

axial resolution = one half the spatial pulse length AR(mm) = SPL (mm) / 2

9
New cards

if the distance between 2 interfaces is less than ½ SPL

only one elongated echo will appear

10
New cards

What is the relationship between axial resolution &SPL?

Directly related. Increasing SPL, increasing AR (poorer AR)

11
New cards

To improve Aixal resolution, we must

decrease SPL

12
New cards

To decrease SPL, we need to

decrease wavelength, so increase frequency

13
New cards

we cannot decrease the number of cycles to improve axial resolution because

cycles/pulse is built into the ultrasound system (backing materials)

14
New cards

What is the relationship between axial resolution and frequency?

Inversely related. The higher frequency, the lower axial resolution (better)

15
New cards

Synonyms of Axial Resolution

Longitudinal resolution; Range resolution, Depth resolution

16
New cards

LARD

Longitudinal, Aixal, Range, Depth

17
New cards

Lateral resolution

Minimum reflector separation required in the direction perpendicular to the sound beam that can produce 2 separate echoes.

18
New cards

What does lateral resolution determine?

it determines how close together, side by side, two separate interfaces can be before they can no longer be resolved as 2 separate echoes when scanning across the reflectors.

19
New cards

The equation of Lateral resolution

Lateral resolution is equal to beam diameter. LR (mm) = beam diameter (mm)

20
New cards

What is the relationship between beam diameter and lateral resolution?

The smaller the beam diameter, the smaller the lateral resolution, the better

21
New cards

What will happen if beam diameter is smaller than the distance between 2 interfaces

2 echoes will be seen

22
New cards

True or false: lateral resolution varies with depth.

True. Because beam diameter varies with depth.

23
New cards

Where is lateral resolution the best?

Lateral resolution is the best in the focal zone since the sound beam is the narrowest at this location.

24
New cards

beam diameter is also known as:

pulse diameter, beam width, pulse width

25
New cards

Synonyms of lateral resolution

Angular resolution, Transverse resolution, Azimuthal resolution

26
New cards

LATA

Lateral resolution, Angular Resolution, Transverse resolution, Azimuthal resolution

27
New cards

How can we improve lateral resolution

Reduce the beam diameter.

28
New cards

How can we reduce the beam diameter?

Focus; use smaller crystal diameter, and increase frequency (when sufficiently far from the transducer face)

29
New cards

True or false: Diagnostic sonography transducers usually have better axial resolution than lateral resolution.

True

30
New cards

What is the relationship between AR and LR in the focal region?

AR and LR are comparable in the focal region

31
New cards

What is beam former?

Beam former sends digitized echo voltage signals to the signal processor

32
New cards

What does beam former consist of?

Pulser (Pulse generator), echo delays, T/R switch, amplifier, A-D-C, summer

33
New cards

What is the pulser?

Pulser is part of the beam former, and it produces electrical pulses that are sent to the transducer for conversion to sound energy.

34
New cards

What will alert the receiver and the memory that a sound pulse is being produced to prevent interference along the sound path?

Master synchronizer

35
New cards

What does the pulser control?

it controls output power transferred to the patient, PRF (1 voltage pulse for 1 pulse of ultrasound), amplitude, intensity of the ultrasound pulse

36
New cards

What is the relationship between voltage pulse amplitude and usltrasound pulse amplitude and intensity?

The greater the voltage pulse amplitude, the greater the amplitude and intensity of the ultrasound pulse.

37
New cards

What is ALARA principle?

As low as reasonably achievable

38
New cards

How does the voltage amplitude affect the sonographic image?

Stronger voltages will produce stronger pulses and therefore stronger echoes. low voltage = weak ultrasound power = weak echoes = dark images. High voltage =strong ultrasound power = strong echoes = bright image

39
New cards

What are the two functions of the transducer?

Transmitter: stimulated by the electric voltage from pulser and produce the ultrasound and sends the sound into patient. Receiver: ultrasound waves return as echoes that are transformed back into electrical information that can be processed by ultrasound system

40
New cards

What does T/R switch do?

it diverts the voltages from the beam former in the correct direction

41
New cards

what does T/R switch do during transmission and receiving?

During transmission it sends voltages to transducer; during receiving sends to the receiver via the Beam Former.

42
New cards

What does receiver do?

The receiver processes the voltage into information regarding position and orientation of the reflectors.

43
New cards

There are 5 receiver functions:

Amplification, compensation, compression, demodulation, rejection

44
New cards

What is the receiver?

The receiver is the component of the ultrasound system that processes the electronic signals produced by the transducer during reception for display.

45
New cards

What are the two functional parts of the receiver?

  1. beam former

  2. signal processor


46
New cards

what functions of the receiver are part of signal processing?

Compression, demodulation, and rejection

47
New cards

What is amplification?

it means to increase the small voltages to larger ones suitable for processing and storage and makes it louder or brighter

48
New cards

what does amplification deal with?

Amplification deals strictly with echoes, does not affect output power.

49
New cards

Which knobo is related to amplification?

Receiver overall “gain“ amplifies

50
New cards

What is gain?

Gain is the ratio of output power to signal or electrical power input

51
New cards

What is compensation?

Compensation means to equalize all returning echoes regardless of depth

52
New cards

What is the relationship between amplitude and path depth?

Attenuation is related to path depth so echoes returning from a greater depth will have a lower amplitude than those returning from a shallow depth.

53
New cards

What is the difference between compensation and amplification?

Compensation is selective amplification: the greater the depth, the greater amplification needed.

54
New cards

What is lateral gain control?

adjust the gain on the image side to side, common in echocardiography

55
New cards

What is compensation also called?

Time gain compensation, Depth Gain Compensation, swept gain, gain compensation

56
New cards

Why is TGC important?

The TGC controls are an important aspect of obtaining a good diagnostic image. The wrong settings can hide abnormalities or create false pathology.

57
New cards

What is rejection also called?


Filtering, suppression, or threshold

58
New cards

What is rejection?

eliminates the smaller amplitude voltage pulses produced by weaker reflections and electronic noise

59
New cards

Where do the weaker reflections come from?

from multiple scattering from within the tissue

60
New cards

True or false: Weaker reflections contribute no useful information and may interfere with observation of useful information (speckle artifact)

True

61
New cards

Is rejection controlable by the sonographer?

Yes

62
New cards

What is demodulation?

it converts the echo voltage from Radio Frequency to Video form using rectification and smoothing

63
New cards

What is demodulation also known as?

Detection or envolope detection

64
New cards

Is demodulation controlable by the sonographer?

No

65
New cards

What is rectification?

turn all negative voltages into positive ones

66
New cards

What is smoothing?

putting an envolope around the bumps to even out the rough edges

67
New cards

What is compression/dynamic range?

The process of decreasing the differences between the smallest and largest amplitudes

68
New cards

is compression controllable by the sonographer?

Yes

69
New cards

What does compression do?

it amplifies weak inputs more than stronger ones while maintaining the relationships between them so the largest is still the largest and the smallest is still the smallest

70
New cards

What is dynamic range?

It is the ratio of the largest amplitude to the smallest amplitude that a system can handle.

71
New cards

Dynamic range is expressed in

dB

72
New cards

What is the typical dynamic range?

0-80 dB (DMS)

73
New cards

What is the relationship between impedance differences and dynamic range?

If the impedance differences are small, a low dynamic range is necessary. Creates more contrast on the image. If impedance differences are large, a higher dynamic range is necessary, Smoother image is created

74
New cards

what do the 5 functions do the echoes?

Manipulate the echoes before image creation

75
New cards

What is the only function not in tech’s control?

Demodulation

76
New cards

What is Analog-to-Digital Converters (ADC) ?

ADC convert the analog (proportional) voltages representing echoes to numbers for digital signal processing and storage.

77
New cards

True or false: some ultrasound systems place ADC closer to the transducer.

True

78
New cards