What is attenuation
The reduction of an x-ray beam as it passes through an object due to absorption
What factors affect attenuation
Density, thickness, and atomic number of the material
What is beam attenuation measured in
Hounsfield Units (HU)
What does 1 HU represent
0.1% difference between the linear attenuation coefficient of water
What is an algorithm in CT
A finite set of steps performed to solve a problem
What is image reconstruction
The process of using raw data to create an image by assigning CT numbers to pixels
What is interpolation
A mathematical method of estimating an unknown value using known values on either side
What is raw data
Data collected from detectors that has not yet been assigned Hounsfield Units
What is image data
Processed raw data in which each pixel has been assigned a Hounsfield Unit
What is convolved data
Data that has been filtered by a kernel during filtered back projection
What is the difference between raw data and image data
Image data has been processed and assigned HU values
What is reconstruction time
The time between the end of scanning and when the image is displayed
What is prospective reconstruction
An image produced during scanning
What is retrospective reconstruction
The process of creating a new image from raw data after scanning has finished
What is a ray in CT imaging
The path an x-ray beam takes from the tube to the detector
What is a ray sum
The measurement of how much beam attenuation occurs for each ray
What is a view
A complete set of ray sums
What is an attenuation profile
The account of attenuation properties for each ray sum correlated to its position
What is back projection
A summation technique that converts attenuation profiles into a matrix to form an image
What is filtered back projection
A method that applies a correction filter to improve image sharpness and edge accuracy
What is convolution
Applying a filter function to an attenuation profile
What are common convolution filters
Standard, smoothing, edge enhancement, bone, and detail filters
What is adaptive statistical iterative reconstruction (ASIR)
A reconstruction technique that uses noise profiles to enhance image clarity and suppress noise
How much can ASIR reduce patient dose
Up to 50%
What is the benefit of iterative reconstruction compared to filtered back projection
Provides similar image quality at lower patient dose
What is interpolation used for in spiral CT
To estimate unknown values between known data points before filtered back projection
What is the scan field of view (SFOV)
The area within the gantry from which raw data is collected
What does SFOV determine
The number of detectors collecting data for a scan
What is the display field of view (DFOV)
The portion of the collected data used to create the image; must be equal to or smaller than the SFOV
What happens if anatomy is outside the SFOV
It cannot be reconstructed
What does a smaller DFOV do
Acts like zoom, displaying a smaller area in greater detail
What is post-processing
Manipulating raw and image data after scanning to enhance visualization
What data is stored in PACS
Image data
What does image reformation or rendering use
Image data to produce images in different planes or 3D views
What can be changed with retrospective reconstruction
DFOV, image center, reconstruction algorithm, image incrementation, or slice thickness
What is the advantage of changing image incrementation
Reduces stair-step effect in multiplanar and 3D images
Why can't slice thickness be smaller than detector acquisition
Because the CT system cannot reconstruct data it didn't collect
Why isn't raw data stored
Too large and complex to maintain
What is multiplanar reformation (MPR)
A technique that produces new slices in different orientations such as sagittal, coronal, or oblique
What is curved planar reformation (CPR)
A reformation following the center of a tubular structure like vessels or colon
What data is required for MPR and 3D reformations
Image data with Hounsfield Units
Why must DFOV and image center match for MPR
Ensures identical alignment across planes
What is 3D reformation
A technique representing the full scan volume in one image
What is surface rendering (shaded surface display)
3D display using only the surface information within a defined HU threshold
What type of values does surface rendering use
Pixel values based on HU thresholds
What is volume rendering
Uses all voxel data, assigning opacity values to each voxel to create a semi-transparent 3D image
What is the difference between surface and volume rendering
Surface uses pixel surface data; volume uses all voxel data
What is maximum intensity projection (MIP)
A 2D image reconstructed from 3D data showing the highest attenuation values; used for angiography
What is minimum intensity projection (MinIP)
A display showing the lowest HU values, highlighting air-filled structures
What are quantitative measurements
Software-based calculations of volume, distance, and density within CT images
What is QCT used for
Quantitative CT bone densitometry; measures beam attenuation of bones to assess density
What is a CT number
The measurement of linear attenuation coefficient for a tissue, expressed in HU
What are CT numbers also known as
Hounsfield numbers
What are the standard Hounsfield Unit values
Water = 0, Air = -1000, Dense bone = +1000
What does pixel brightness correspond to
Hounsfield Unit value
Can attenuation of tissue be changed
Yes, by using contrast media
Why are Hounsfield Units clinically useful
They help identify and measure different tissues and pathologies
What is windowing
Adjusting displayed image contrast by manipulating window width (WW) and window level (WL)
What does window width (WW) represent
The range of CT numbers displayed
What does reducing window width do
Increases image contrast
What does a wide window width do
Creates a long gray scale with low contrast (used for bone imaging)
What does a narrow window width do
Creates a short gray scale with high contrast (used for soft tissue imaging)
What does window level (WL) represent
The center CT number of the window width; determines tissue brightness
Where should WL be set for bone imaging
Between +300 and +1000
Where should WL be set for fluid structures
Around 0
Where should WL be set for lung imaging
Around -500
How does increasing window level affect the image
Shifts the gray scale toward higher HU (brighter tissues)
How does decreasing window level affect the image
Makes the image appear darker
What happens if window width increases
Longer gray scale and lower contrast
What happens if window width decreases
Shorter gray scale and higher contrast
What does window level tell you
The tissue type of interest
What is region of interest (ROI)
A selected area on the image used for measuring HU values and standard deviation
What does standard deviation indicate in ROI analysis
Variation in HU within the selected region; closer to zero means more uniformity
Why is ROI size important
It should match the size of the structure being measured
What is distance measurement used for in CT
To measure size or distance between structures (e.g., aortic diameter)
What is image annotation
Adding text or markers on an image to convey additional information
What is a reference image
A scout or overview image showing slice locations relative to anatomy
What is image magnification
Temporarily enlarging a portion of an image for detailed viewing
What are two methods of image magnification
Changing DFOV (permanent change) and using magnifying glass (temporary zoom)
What is cine imaging
Continuous scanning without table movement; used to monitor contrast enhancement in vessels
When is cine imaging most commonly used
During arterial phase imaging to capture timing of contrast peaks