1/15
Looks like no tags are added yet.
Low magnification
x6-x10
• Lids
• Bulbar conjunctiva
• Cornea / limbus
• Tears / tear film
• Anterior chamber / iris
• Crystalline lens
Medium magnification
x20-x25
• Epithelium of cornea
• Stroma of cornea
• Endothelium of cornea
• Lens fits / surface of contact lenses and movement on blink
High magnification
x30-x40
• Epithelium: vacuoles, microcysts, dystrophies
• Stroma: striae and folds
• Endothelium: polymegethism, blebs, dystrophies
The purposes of slit lamp examination are:
To see whether the ocular tissue is safe to fit with contact lenses
To get base line data to monitor changes to the ocular tissue with lens wear
Medico-legal requirements
Diffuse illumination
Use a diffuser with a flip-up system to spread out light from the illumination system. If unavailable, use a reversed mirror or tissue.
Magnification: Low magnification.
Beam: Wide-angle beam.
Use: Allows an overall look at the eye and adnexa.
Optic section
Beam: Thin beam to illuminate a section of the cornea.
Magnification: Low to high magnification, giving a cross-section of the cornea.
Positioning: Illumination system must be at a right angle to the corneal surface to illuminate the separate layers of the cornea.
Angle: Vary the angle of illumination while keeping the observation system directly straight ahead.
Paralepidid
Beam: Broader beam with an average width of 1 mm.
Angle: Vary the angle of illumination.
Magnification: Low to high magnification.
Uses: Looking at corneal nerve fibres and punctate keratitis.
Conical beam
View: Allows us to view the full thickness of the cornea.
Beam: Circular, narrow and short beam.
Magnification: Medium magnification.
Room conditions: Must be carried out in a darkened room to see if there is any light scatter back at us, which shows inflammatory cells (if any) in the anterior chamber.
Angle: Observed at 45–60 degrees, allowing us to observe all areas.
What are the features and uses of indirect illumination?
Technique: Same as direct illumination, but looking to the side of the light beam, as a direct beam may bleach defects from view.
Positioning: Decouple the system to focus the observation system to the side of the direct beam. This helps us optimally focus the object of interest.
Uses: Looking at epithelial vesicles and erosions, iris pathology and sphincter.
In-direct illumination
2 types:
Used for epithelial vesicles and erosions. There are two types of systems:
Coupled - when illumination and observation system swing around the same pivot point
De-coupled - when the illumination system is moved to the side therefore it no longer meets the optimal focus of the observation system
Retro-illumination
Light bounces off the iris or fundus to back illuminate the cornea (Req’s de-coupled system)
Used for observing microcysts, vacuoles, cataracts, CL deposits
Moderately wide beam
Med to high mag
Specular Reflection
Angle of incidence of illumination is equal to angle of observation
Purkinjie image 1 - front surface of cornea
Purkinjie image 2 - endothelium, where there is a change of R.index from cornea to aqueous
Purkinjie image 3 - front surface of the lens
Purkinjie image 4 - back surface of crystalline lens
Medium to high magnification
Focus on area of interest and swing illumination beam until specular reflection achieved
Only visible through one eyepiece at a time
USES: look at tear film break up time and flow of debris, endothelium, crystalline lens (orange peel)

Tangential
Illumination at right angles to structure of interest
Low magnification
Wide angle beam but narrow
Used for iris freckles, general integrity of cornea and iris
Sclerotic scatter
Illumination slit located on limbus
You see halo of light in darkened room
Observation system de-coupled
Low magnification
Used for corneal opacities, corneal swelling, foreign bodies
Fluorescein examination
Assess staining of cornea & conjunctiva
Assess tear break up time
Assess tear meniscus height
Use Wratten filter to enhance image
Van Herick’s Technique
Low magnification Grd 4 - 1:1 open
Optic section with higher of 3-4mm Grd 3 - 1:2
60 degrees angle to observation Grd 2 - 1:4
Comparing width of cornea to anterior chamber Grd 1 - 1:<4
Grd 0 - closed
Conduct assessment also on the nasal side, if patient’s nose gets in the way lock the slit lamp at 60 degrees and move both systems approximately 5-10 degrees away from the nose, but ENSURE patient looks directly at the microscope before assessing the angle.