X-rays
Form of electromagnetic radiation
Short wavelength, high-frequency
10^-8 - 10^-13
Considered wave or photon
How x-rays formed at target
electrons decelerated by target
Kinetic energy of electrons becomes energy of photons
Why there is minimum wavelength
As electrons are accelerated through p.d. V, each electron reaches anode with kinetic energy Ek = eV
Most energetic photon uses all of energy in interaction hfmax = eV
eV = hc/wavelengthmin
some ke dissipates as thermal energy
Wavelength inversely proportional to photon energy
Max photon energy when all ke = photon energy, therefore minimum wavelength
How X-rays are produced
Fast-moving x-rays rapidly decelerated
As electrons slow down, kinetic energy is transformed to photons of EM radiation
Gamma rays produced in radioactive decay
gamma photon emitted by decaying nucleus after alpha or beta particle
X-ray tube anodes
Cathode (heated filament that electrons are emitted from) and anode (rotating anode is made of hard metal like tungsten)

Why rotate anode
Most of time, not all ke converted to photons, a lot converted to heat
Rotate to allow cool down by radiating heat to surroundings COULD JSE WATER

How x-ray tubes work
Voltage is supplied to electrodes
Voltage accelerates beam of electrons between gap (kinetic energy is eV of supplied p.d)
Electrons strike anode at high speed, lose kinetic energy in form of X-ray photons
Electron is attracted to nucleus of atom in anode, changing speed and direction(decelerating)
Decelerate loses energy by emitting EM radiation(photons)
Equation of maximum frequency
Energy gained through acceleration of electron = eV
E is maximum energy photon can have
maximum X-ray frequency calculated with e = hf
fmax = eV/h
How to eliminate low energy X-rays
Lowest energy X-rays don’t have enough energy to penetrate body, and no effect on image, but still increases dose to patient
NEEDS TO BE FILTERED
Aluminium absorbers
Controlling x-ray intensity
power per unit area
Rate of energy transfer per unit cross sectional area
Increasing current in tube increases intensity
Greater current (more electrons per second) collides with anode, meaning more x-rays per second
More intense = image formed in shorter time
Collimated beam
Thin outer casing window - allows X-rays to emerge into space outside of tube
Control width of X-ray with metal tubes outside window to absorb X-rays (aluminium)
Produces parallel sided beam called collimated beam
Attenuation
gradual decrease of intensity of EM radiation as it passes through a medium
CT scanning
Rotate around one section - x-ray tube rotates around patient through one slice/section
Record many angles - detectors opposite tube record x-ray images of that section from many angles
Computer combines images to give 2D image of section
Patient is moved along axis and process is repeated for many sections
Computer combines 2D images to form 3D images
X-ray attenuation equation
𝐼=𝐼0𝑒^−𝜇𝑥
I = intensity
x = thickness
djskdf is attenuation coefficient of material
Half-thickness
Thickness of material that will reduce transmitted intensity of an X-ray beam of particular frequency to half value
AIM of radiographers
reduce patients exposure to harmful x-ray as much as possible
Improve contrast (difference in bright and dark)
Harm of x-rays
Ionising, damage living tissue, causing mutation and leading growth of cancerous tissue
Reducing x-ray dosage
Intensifier screens
contains phosphor, emit visible light when absorbs x-ray photons
X ray film between two screens, screens emit light photons - light photons blacken screen
Image intensifiers for digital
X-ray strikes phosphor screen, produces visible light photons
Photons release electrons from photo cathode (photoelectric effect)
Electrons accelerated and focused by positively charged anode to strike screen, give out visible light, viewed via television camera, viewed with device
Contrast
Good contras - clear difference in blackening of film
Determined by hardness of x-ray (hard = high energy, low wavelength)
Improving contrast
Contrast media:
materials like barium are good absorbers of x-rays, used to show edges of soft tissue in x-ray images
Have high atomic number, atoms have many electrons (more absorbing)
Injected or swallow pill for area of interest
CT scanner
Used for 3d, depth shown
Patient lies in vertical ring of x-ray detector
X-ray tube rotates around ring, exposes patient to fan shape beam of x-rays from all directions
Detectors opposite tube send electronic records to computer
Computer software builds up 3d image of patient
Radiographer views images of slices through the patient on computer screen
USe xrays
Scan in sections
Many angles
Image of each section is 2D
scan repeatedly over sections
Combine images of sections to produce 3D scan
Advantages of CT scan
Produce images that show 3D relationships between different tissues
Can distinguish tissues with similar densities (attenuation coefficients)
Dangers of CT scans
high exposure to x-ray and ionizing radiation
More expensive
Why patient should hold breath in CT scan
Movement means images from different angles don’t match, image of section is blurred
Increasing p.d effect on x-ray beam
Electron gains more energy
Min wavelength decreases E = hc/wav
Penetration increases (as electrons have more energy)
Intensity increases
Controlling intensity of beam
I = rate of energy transfer per unit cross sectional area
Increase current in X-ray tube to increase intensity
more electrons per second - more photons per second
Image formed faster
Increasing tube current effect effect on X-ray beams
More electrons per second hit target (I = rate of flow of electrons)
More electrons = more photons per second (increase intensity)
Wavelength unchanged (no change in energies)
Exposure time decreased as image forms faster from higher intensity (more photons per second)
Aluminum filters effect on x-ray beams
Absorbs low energy photons (decreasing amounts of soft x-rays)
Patient dosage of x-rays decrease
Collimator effect on x-ray beams
Lead plates
Limit beam to area being imaged
Narrower parallel beam
Decreases scattering and dosage
Effect of rotating anode
Spreads heat out
Focal-spot temperature decreases, exposes fresh tungsten under beam
Why are xrays used to produce images of internal body structure that have good contrast
Different attenuation coefficients between bone and tissue
Trasnmitted intensities are differe,t cause different degrees of blackening therefore high contrast
Why continuous spectrums of energies of xray photons are produced
Electrons decelerate when they hit target metal to produce xray photons
There are a range of decelerations, photon energy is dependent on deceleration magnitude
Therefore range of energies
Sharpness definition
Ease of which edges can be distinguished
Contrast definition
Difference in degrees of blackening from different intensities
Comparing attenuation coefficients to explain contrast
Similar attenuation coefficient (factor 3.5 between detected intensities considered too similar)
Similar attenuation = lower contrast