Chapter 5: The X-Ray Tube

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
full-widthPodcast
1
Card Sorting

1/34

flashcard set

Earn XP

Description and Tags

Flashcards covering the nine learning objectives for Chapter 5: The X-Ray Tube, including tube housing, internal components, cathode/anode design, the line-focus principle, the anode heel effect, electrical paths, and safe operational guidelines.

35 Terms

1
New cards

What is the maximum permissible leakage radiation escaping the protective housing of an x-ray tube under regulatory limits?

Less than 100 mR/hr100\,\text{mR/hr} (0.0258 millicoulomb/kg/h0.0258\,\text{millicoulomb/kg/h}) at a distance of 1 meter1\,\text{meter}.

2
New cards

How do the oil bath and cooling fans protect the x-ray tube within its housing?

The oil bath draws heat away from the tube, and cooling fans circulate air around the assembly to dissipate heat and prevent thermal damage.

3
New cards

Why should the high-voltage electrical cables never be used as handles to maneuver the tube?

Using them as handles can damage the equipment and create a safety risk to the radiographer.

4
New cards

What are the three principal internal components of a general-purpose electronic vacuum x-ray tube?

A positive anode, a negative cathode, and an induction motor.

5
New cards

Why must the glass or metal envelope maintain a vacuum during x-ray tube operation?

Without a vacuum, air electrons could join the electron stream, causing arcing and damaging the tube.

6
New cards

What are glass envelopes typically made of, and what advantage do metal envelopes provide?

Glass envelopes are generally made of heat-resistant borosilicate glass; metal envelopes help maintain a constant electrical potential between the electron stream and enclosure to prevent arcing.

7
New cards

What is the target window of an x-ray tube enclosure and what is its typical size?

A specially designed thinner section of the enclosure (usually about 5 cm5\,\text{cm} square) that minimally absorbs the useful x-ray beam at the exit point.

8
New cards

What process occurs when electrical current heats the selected filament to incandescence?

Electrons are boiled off the filament through thermionic emission.

9
New cards

What is the space charge in an x-ray tube?

The electron cloud that forms around the filament after electrons are released by thermionic emission.

10
New cards

What is the space-charge effect?

A self-limiting condition that occurs when the electron cloud reaches a size where it becomes difficult for additional electrons to be emitted.

11
New cards

What speed do electrons reach when traveling across the 1–3 cm1\text{–}3\,\text{cm} gap from cathode to anode?

About half the speed of light.

12
New cards

How is a stationary anode constructed, and why are its exposure techniques limited?

It consists of a tungsten button embedded in a copper rod; because electrons continually strike the same small area, rapid heat buildup limits exposure techniques and risks target damage.

13
New cards

What materials comprise the disc and shaft of a rotating anode?

A molybdenum base/core coated with tungsten, mounted on a copper shaft with a molybdenum core.

14
New cards

What three properties make tungsten the ideal target material for the anode?

A high melting point (about 3400∘C3400^\circ\text{C}), high thermal conductivity, and a high atomic number (7474).

15
New cards

Why is rhenium sometimes added to the tungsten target of a rotating anode?

To increase thermal capacity and tensile strength.

16
New cards

Why is molybdenum used in the core of the rotating anode disc and shaft?

It is lightweight and has low thermal conductivity, which slows heat transfer toward the rotor bearings.

17
New cards

Where are the stator and rotor of the induction motor located relative to the tube enclosure?

The stator is located outside the tube enclosure, while the rotor is located inside the enclosure.

18
New cards

By what mechanism does the induction motor rotate the anode without direct electrical connection to the rotor?

Electromagnetic mutual induction.

19
New cards

What are the typical anode rotation speeds in general-purpose and specialty x-ray tubes?

Approximately 3,400 rpm3,400\,\text{rpm} for general-purpose tubes and up to 10,000 rpm10,000\,\text{rpm} for specialty tubes.

20
New cards

What is the purpose of the line-focus principle?

It uses an angled anode target face so a large actual focal spot can be maintained for heat dissipation while producing a small effective focal spot for image detail.

21
New cards

How do the actual focal spot and effective focal spot differ in definition?

The actual focal spot is the area on the anode target actually bombarded by electrons, while the effective focal spot is the x-ray beam area as seen from the patient's perspective.

22
New cards

What range of anode target angles is commonly used in general-purpose tubes, and which angle is most common?

Target angles commonly range from 7∘7^\circ to 18∘18^\circ, with 12∘12^\circ being the most common.

23
New cards

What causes the anode heel effect?

Because the anode target is angled, x-ray photons produced on the anode side must pass through part of the target (the heel), causing some photons to be weakened and absorbed.

24
New cards

What relative beam intensities are found on the anode and cathode sides due to the heel effect?

Approximately 75%75\% intensity on the anode side and 120%120\% intensity on the cathode side.

25
New cards

What are the physical dimensions of a cathode filament coil?

Usually 7–15 mm7\text{–}15\,\text{mm} long and 1–2 mm1\text{–}2\,\text{mm} wide.

26
New cards

Why is 1%–2%1\%\text{–}2\% thorium added to the tungsten cathode filament?

To increase thermionic emission and extend filament life.

27
New cards

What material is the focusing cup made of, and how does it focus the electrons?

It is made of nickel and uses a strong negative charge to electrostatically repel and gather electrons into a cloud aimed at the anode target.

28
New cards

To which circuits do the filaments and focusing cup connect?

The filaments connect to the filament circuit (sharing a common ground wire), while the focusing cup connects to the secondary circuit.

29
New cards

What two events occur during the rotor preparation phase of the two-phase exposure switch?

The induction motor accelerates the rotor to operating speed, and the selected filament is energized to achieve the desired thermionic emission.

30
New cards

What path does voltage follow along the high-voltage circuit in a high-frequency generator before reaching the tube?

From the autotransformer, through capacitor banks and the inverter circuit, through the step-up transformer, and then through a rectifier bank.

31
New cards

What are the three fundamental requirements for x-ray production?

(1) A large potential difference to give electrons kinetic energy (kVp\text{kVp}), (2) a source/quantity of electrons (mA\text{mA}), and (3) a target for interaction (the anode).

32
New cards

What are the three major causes of x-ray tube failure?

(1) Frequent use of very high or maximum exposure factors, (2) low but very long exposure factors keeping the tube at high temperatures, and (3) filament overload from prolonged excessive heating or arcing.

33
New cards

What three heat transfer processes dissipate thermal energy from the x-ray tube assembly?

Conduction through heat-tolerant materials, radiation from the anode to the oil bath, and convection into the room through cooling fans.

34
New cards

What is the formula used to calculate Heat Units (HU\text{HU})?

HU=kVp×mA×s×c\text{HU} = \text{kVp} \times \text{mA} \times s \times c, where ss is exposure time in seconds and cc is the generator correction factor.

35
New cards

What are the generator correction factors (cc) for single-phase, three-phase 6 pulse, three-phase 12 pulse, and high-frequency units?

Single-phase = 1.01.0; three-phase, 6 pulse = 1.351.35; three-phase, 12 pulse = 1.411.41; high-frequency = 1.451.45.