1/12
Vocabulary practice flashcards covering the biomechanical study on FiberTape® Cerclage versus stainless steel wire for sternal closure, including material properties, testing parameters, and mechanical outcomes.
Median sternotomy
The primary surgical approach used in cardiothoracic surgery to gain efficient and unobstructed access to the heart.
Deep sternal wound infection
A severe potential complication following sternal dehiscence that carries a mortality rate as high as 47%\text{.}
Bone cut-through
A complication where physiological loads from respiration or coughing concentrate on the small surface area of metal wire, causing it to cut through bone and potentially leading to dehiscence and infection.
FiberTape® cerclage system
A 100%\text{ nonmetallic, suture-based sternal closure solution featuring a flat construct that provides broad footprint compression and a larger implant-to-bone contact area.}
Ultra-high-molecular-weight polyethylene (UHMWPE)
The material from which FiberTape cerclage is manufactured, rendering it 100%\text{ radiolucent, MR safe, soft without sharp edges, and suitable for patients with metal allergies.}
Clinical failure
The point during static tension mechanical testing defined by the formation of a 3mm\text{ gap in the sternal construct.}
Static tension loading rate
The displacement rate of 10mm/s\text{ applied to test constructs fixed to custom sternal metal fixtures (N=6) to measure ultimate load and gap formation.}
Mean ultimate load (FiberTape® cerclage)
1616.5±77.3N\text{, which is statistically significantly higher than that of stainless steel wire (P≤.001).}
Mean ultimate load (Stainless steel wire)
856.2±168.2N\text{ as determined during static tensile loading to failure.}
Mean load at 3mm\text{ gap formation (FiberTape® cerclage)}
833.6±176.0N\text{, significantly exceeding the load capacity of stainless steel wire (P=.001).}
Mean load at 3mm\text{ gap formation (Stainless steel wire)}
518.2±15.5N\text{ recorded at the defined clinical failure threshold.}
Ductile mechanical behavior (Stainless steel wire)
A mechanical profile characterized by elongation, low modulus of resilience, and plastic deformation or stiffness loss occurring around or before the clinical failure point.
Brittle-like mechanical behavior (FiberTape® cerclage)
A mechanical response characterized by higher mechanical strength, minimal stretching, and a high modulus of resilience prior to failure.