1/48
Comprehensive vocabulary flashcards covering nucleic acid extraction, quantification, chromatography, protein chemistry, PCR, electrophoresis, and analytical terminology.
Four Stages of Column-Based Extraction
lysis (releases nucleic acids), binding (captures them), washing (removes contaminants), and elution (releases purified nucleic acids).
Chaotropic Agent
A substance that disrupts noncovalent interactions and protein structure; guanidinium salts are common examples.
Guanidinium Thiocyanate
A chaotropic agent that denatures proteins, helps inactivate nucleases, and promotes nucleic acid binding under suitable conditions.
Dry Spin
A centrifugation step before elution that removes residual alcohol, which can inhibit downstream enzymatic reactions such as PCR.
Low-Salt Elution Buffer
A buffer with low ionic strength that rehydrates nucleic acids and weakens adsorption, allowing them to leave the silica matrix.
EDTA (in Extraction Buffers)
A reagent that chelates divalent metal ions required by many nucleases; excess amounts can inhibit downstream enzymes.
Proteinase K
An enzyme that digests proteins, including many nucleases and nucleic acid-associated proteins.
DNase (in RNA Extraction)
An enzyme used for removing contaminating genomic DNA.
RNase (in DNA Extraction)
An enzyme used for removing contaminating RNA.
Guanidinium Bleach Incompatibility
A safety hazard where mixing guanidinium-containing waste with bleach generates highly toxic gases.
Spectrophotometer
An instrument that measures light absorbance or transmittance at selected wavelengths.
Fluorometer
An instrument that measures light emitted by fluorescent molecules after excitation.
NanoDrop
An instrument that quantifies nucleic acids by measuring UV absorbance at 260nm and estimating concentration using an appropriate conversion factor.
Qubit
A fluorometric quantification platform in which fluorescent dyes selectively bind a target class of nucleic acid and fluorescence is compared with standards.
A260/A280 Ratio
An approximate purity indicator with typical values of approximately 1.8 for pure DNA and 2.0 for pure RNA; low values can suggest protein or phenol contamination.
A260/A230 Ratio
An approximate purity indicator with typical values of approximately 2.0–2.2 for pure nucleic acids; low values may indicate salts, guanidinium, phenol, or other contaminants.
Beer–Lambert Law
The law expressed as A=εlc, stating that absorbance is proportional to concentration and optical path length within the linear range.
A260=1 Path Length Equivalents
At a 1cm path length, an absorbance of A260=1 corresponds to approximately 50μg/mL for dsDNA or 40μg/mL for RNA, assuming pure material.
Affinity Chromatography
A separation method based on specific binding between a target and an immobilized ligand.
Ion-Exchange Chromatography
A separation method based on reversible electrostatic interactions between molecules and a charged resin.
Size-Exclusion Chromatography (SEC)
A separation technique based on hydrodynamic size using porous beads, generally without target binding, where larger molecules elute first.
Isoelectric Point (pI)
The pH at which a protein's net electrical charge is zero; the protein carries a net positive charge below this value and a net negative charge above it.
Cation-Exchange Resin
A chromatography resin that is negatively charged and binds positively charged molecules.
Anion-Exchange Resin
A chromatography resin that is positively charged and binds negatively charged molecules.
Protein A and Protein G
Affinity chromatography ligands that generally bind the Fc region of IgG, with binding affinity depending on species and subclass.
pKa
The pH at which acid and conjugate base concentrations are equal; buffering capacity is generally strongest near this value.
PCR
Amplification of a specific DNA sequence using primers, polymerase, and thermal cycling.
RT-PCR
Reverse transcription PCR, in which RNA is converted to cDNA prior to PCR amplification.
qPCR
Quantitative real-time PCR, which monitors target amplification using fluorescence.
RT-qPCR
Reverse transcription followed by quantitative real-time PCR to measure RNA-derived targets.
Ct or Cq
The cycle number at which fluorescence crosses a defined threshold; lower values generally indicate more starting target.
qPCR Melt Curve
A thermal dissociation profile for dye-based assays where melting peaks indicate product specificity or the presence of primer-dimers.
No-Template Control (NTC)
A qPCR reaction prepared without a template, used to detect reagent contamination or nonspecific amplification.
No-RT Control
An RNA sample processed without reverse transcriptase, used to detect genomic DNA contribution.
100% qPCR Efficiency
A reaction efficiency condition where the target approximately doubles each cycle, causing a tenfold dilution to shift the Cq by about 3.32 cycles.
Agarose Gel Resolution
Electrophoretic separation property where higher agarose concentrations yield smaller pores and better resolution of smaller fragments, while lower agarose suits larger fragments.
Accuracy
The closeness of a measured value to the true or reference value.
Precision
The agreement among repeated measurements.
Analytical Sensitivity
The change in assay response per unit change in analyte concentration.
Analytical Specificity
The ability of an assay to measure the intended analyte without interference from other substances.
Limit of Detection (LOD)
The lowest amount of an analyte that is reliably distinguishable from background under defined conditions.
Limit of Quantification (LOQ)
The lowest amount of an analyte that can be measured with acceptable accuracy and precision.
Repeatability
Precision evaluated under the same operating conditions over a short time interval.
Reproducibility
Precision evaluated under changed conditions, especially between different laboratories.
Robustness
The capacity of an analytical method to remain unaffected by small, deliberate changes in operating conditions.
Recovery Efficiency
The recovered target amount divided by the starting target amount, expressed as a percentage.
RPM vs. RCF
RPM is rotational speed, whereas RCF is centrifugal acceleration relative to gravity and depends on rotor radius.
Pore Size vs. MWCO
Pore size describes actual physical openings, whereas molecular weight cut-off (MWCO) is an approximate molecular mass retention rating for ultrafiltration.