1/15
Looks like no tags are added yet.
Calibration of dental x-ray equipment can be performed by a dentist, dental hygienist, or dental assistant.
False. A qualified technician must calibrate dental x-ray equipment to ensure consistent x-ray machine performance and production of diagnostic radiographs.
Annual tests for dental x-ray machines can be performed by a dentist, dental hygienist, or dental assistant.
True
For quality control purposes, each new box of unopened film should be tested for film freshness and fogging before it is used.
True
After processing, fresh film that has been properly stored and protected will appear...
Clear with a slight blue tint
After performing the screen-film contact test, a wire mesh image of uniform density appears. These results indicate...
Adequate film-screen contact
When functioning properly, a viewbox should emit a uniform and intense light
False. It should emit a uniform and subdued light when functioning properly.
One of the most critical areas of quality control that requires daily monitoring is...
Film processing
The coin test is used to check...
Proper safelighting. The coin test is an easy way to check the light-tightness of the darkroom. The lighting in the darkroom is a critical step in proper film processing, and the coin test determines whether the safelight is of proper wattage and distance from the work area.
The following must be closely monitored with manual processing techniques
Thermometer, timer, water bath temperature, developer, and fixer solutions
On the average, processing solutions should be changed....
Every 3-4 weeks
On the average, processing solutions should be replenished...
Daily
Fresh films and fresh chemicals must be used when preparing reference radiographs
True
A reference radiograph is used to check...
Film densities of radiographs that are processed daily
The densities seen on the daily image appear lighter than the densities seen on the reference radiograph; this result indicates that...
The developer solution is too weak
The clearing test is used to monitor...
Fixer solution that removes unexposed silver halide crystals on the film to result in a clearer dental radiograph image
Regardless of who actually exposes the patient to radiation, the dentist is ultimately responsible for the diagnostic quality of all dental images
True