Sono Physics & Instrumentation (Real Time Imaging)

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/28

encourage image

There's no tags or description

Looks like no tags are added yet.

29 Terms

1
New cards

What is is called when images are displayed one frame at a time?

Static Scanning

2
New cards

What is a good way to remember static scanning vs. real time imaging?

Static scanning = US photograph

Real time imaging = US movie

3
New cards

What is the most important operational parameter associated with US movie? Define it.

Frame rate: is the system’s ability to create numerous frames each second

4
New cards

What is frame rate determined by?

  • Sounds speed in the medium

  • The depth of imaging

In clinical ultrasound, the maximum imaging depth determines frame rate


5
New cards

What are the units for frame rate?

Hertz (Hz)

Units of frame rate are images per second

6
New cards

What does temporal resolution pertain to and define it?

It pertains to “accuracy in time”

Describes the ability to precisely position moving structures from instant to instant

7
New cards

What is temporal resolution excellent and substandard?

Excellent: when a system produces many frames per second

Substandard: when it produces few frames per second

8
New cards

What is temporal resolution measured in?

hertz(Hz) or per second

9
New cards

What is the relationships for temporal resolution?

RELATED to frame rate

10
New cards

What is the relationship between frame rate and the time required to make a single image? What happens when they are multiplied together?

Frame rate and time for one frame (Tframe) are INVERSELY RELATED and are RECIPROCALS

Result: is 1

11
New cards

When an ultrasound system creates an image in 1/10th of a second, what is the frame rate? What does this equation illustrate?

Frame rate is 10 frames/seconds or 10Hz

It illustrates that as the time needed to make each image decreases, the frame rate increases

12
New cards

What are 2 Sonographer controlled settings of an ultrasound system that determines frame rate?

  • Imaging depth

  • Number of pulses in each pulse


13
New cards

How does imaging depth affect temporal resolution? How are the two related?

  • Shallow imaging increases frame rate and improves temporal resolution

  • Deeper imaging decreases frame rate and degrades temporal resolution

The two are INVERSELY RELATED


14
New cards

What is the mathematical equation for a single frame rate?

Tframe = # of pulses x PRP = 100 × 1/1000 second = 1/10 second

Frame Rate = 10Hz

15
New cards

How does the number of pulses in each image affect temporal resolution? What is the relationship between the two?

  • Higher frame rates are possible when each individual image in made with fewer pulses

  • Lower frame rates occur when each image is made with more pulses

Pulses per frame and frame rate and INVERSELY RELATED


16
New cards

What 3 factors affect the number of pulses needed to create an image?

  • Number of pulses per scan line (multi-focus vs. single focus)

  • Sector size

  • Lines per angle of sector (line density)


17
New cards

What occurs with multi-focusing? What occurs as a result of multi-focusing?

Multiple pulses are sent down each scan line and the number of pulses required to create an image increases

Multi-focusing decreases the frame rate, and diminishes temporal resolution

18
New cards

What are the 5 things that occur with single focus and the 5 things that occur with multi-focus?

Single

  • one pulse per scan line

  • Shorter T frame

  • Higher frame rate

  • Better temporal resolution

  • Poorer lateral resolution

Multi

  • many pulses per scan line

  • Longer T frame

  • Lower frame rate

  • Diminished temporal resolution

  • Improved lateral resolution


19
New cards

What is the main advantage of multiple focus zones?

It improves the accuracy of the images and each scan line has better lateral resolution bc it is narrow over a wide range of depths

20
New cards

What occurs when sector size expands? What is another name for sector size?

AKA: Field of view

With expansion, more pulses are required to create an image and temporal resolution decreases

21
New cards

What is the relationship for field of view/sector size? What do narrow or wide images result in?

Field of view is INVERSELY RELATED to frame rate?

Narrow images result in higher frame rates

Wider images result in lower frame rates

22
New cards

What are 4 things that happen with a narrow sector and 4 things that happen with a wide sector?

Narrow

  • fewer pulses per frame

  • Shorter T frame

  • Higher frame rates

  • Superior temporal resolution

Wide

  • more pulses per frame

  • Longer T frame

  • Lower frame rate

  • Inferior temporal resolution


23
New cards

What is line density? What does far and packed spacing mean?

Altering the space between sound beams

Far spacing is low line density

Closely packer in high line density

24
New cards

What occurs with low line density and high line density?

With low line density, few pulses create each image and the frame rate is high and so is temporal resolution

With high line density, the number of pulses increases, frame rate drops, and temporal resolution decreases

25
New cards

What is the main advantage of high line density?

The gaps between the lines are smaller, this improves the accuracy of each image

With high line density, each image contains more detail, known as improved spatial resolution

26
New cards

What are the 6 characteristics of low line density and the 6 characteristics of high line density?

Low Line Density

  • Widely spaced lines

  • Fewer pulses per frame

  • Shorter T frame

  • Higher frame rate

  • Higher temporal resolution

  • Poor spatial resolution

High Line Density

  • Tightly packed lines

  • More pulses per frame

  • Longer T frame

  • Lower frame rate

  • Low temporal resolution

  • Excellent spatial resolution


27
New cards

How are temporal resolution and image quality related?

There are many trade-offs in ultrasound: we have to decide which is more important… excellent image quality (superior photograph) or excellent temporal resolution (superior movie)

28
New cards

When trying to obtain a superior image of a stationary structure, what settings should be used? What are these settings consistent with?

  • Multiple focusing

  • Wide field of view

  • High line density

Consistent with low frame rates, wider images, and poor temporal resolution


29
New cards

When trying to obtain a superior movie or a rapid moving structure, what settings should you use? What are these settings consistent with?

  • Single focus

  • Narrow field of view

  • Low line density