implementing imaging standard

image critique

  • requires knowledge of all areas of radiography

  • needs logical, organized approach

  • critical thinking and problem solving

  • critical aspect in digital imaging systems


2
New cards

diagnostic image

  • what is a diagnostic quality image?

    • no practical technical definition

  • larger number of variables involved

    • makes radiography art and science

  • requires application of scientific knowledge to analyze image


3
New cards

important items to remember

  • radiologists not trained to create and analyze quality of image

  • radiographers responsible for image analysis


4
New cards

analytical process

  • overview of entire image

    • search for differences

    • search for patterns and cues

  • narrow search

    • seek and formulate hypothesis

    • holds cues that do not fit hypothesis in reserve


5
New cards

identifying an imaging problem

  • definition of image quality must take into account clinical question

  • must use systematic approach

    • avoid just glancing at image


6
New cards

effective image critique method

3 step process:

  • classification of image

  • determination of cause of problem

  • recommendation of corrective action


7
New cards

classification of image

  • diagnostic acceptance limits

    • optimal diagnostic quality

  • images outside acceptance limits must be further critiqued


8
New cards

determination of cause of the problem

classified into one of three categories:

  • technical factors

  • procedural factors

  • equipment malfunction


9
New cards

technical factor problems

  • two types: visibility of detail and geometric problem with detail

  • visibility of detail problem

    • difficulty with visibility of image

      • density/image receptor exposure and contrast

      • image detectors receiving too little or too much radiation, high or low subject contrast

  • geometric problem with detail itself

    • recorded detail

      • focal spot size, SID, or OID

    • distortion

      • shape, size


10
New cards

procedural factor problem

  • it involves patient positioning or patient preparation for the x-ray examination

  • patient position:

    • tube/part/image receptor alignment

  • patient preparation

    • removal of radiopaque objects

    • proper bowel cleansing for GI exams

      • by providing clear instructions to patient


11
New cards

equipment malfunction

  • requires proper maintenance

  • can avoid with effective quality control program


12
New cards

recommendation of corrective action

  • critical knowledge

    • critiquing acceptable, but suboptimal images

    • repeat analysis

    • recording of successful results

  • personal technique booklet


13
New cards

exposure system selection

  • forefathers of technique:

    • Ed C. Jerman (1920s)

    • Arthur Fuchs (1940s)

  • radiographic exposure

    • very complex process

    • best technique systems manipulate one variable while holding others constant

  • goal of any exposure system:

    • consistency

  • exposure systems often incorporated into pre-programmed control consoles

    • suggested starting points

    • should be revised by radiographer when necessary

  • each system has advantages and disadvantages

  • variables change according to anatomical part thickness

  • important to use calipers when not utilizing AEC


14
New cards

calipers

  • devices used to measure part thickness

  • important QC measure when not using AEC

  • improper use can result in repeated exposures


15
New cards

part measurement

  • two main methods:

    • central ray entrance and exit

    • thickest part

  • same approach must be followed by all department radiographers


16
New cards

exposure systems

3 systems:

  • fixed kVp

  • variable kVp

  • anatomically programmed radiography


17
New cards

fixed kVp systems

  • developed by Fuchs during World War II (1943)

  • theoretical support from literature and professionals

  • aka: fixed kVp/variable mAs technique chart

    • selecting an optimal kVp and adjusting mAs for variations in body part

    • optimal kVp: a kVp that is high enough to ensure proper penetration. In the chart, it is indicated for each body part

  • advantages: (optimal kVp uses reduce in mAs)

    • reduction in patient dose

    • provides more information within image

    • increases consistency in image receptor exposure

    • lengthens exposure latitude

    • reduces x-ray tube wear

    • allows for shorter exposure times

      • thus reduces patient motion

    • easier to memorize

  • disadvantage:

    • higher amount of scatter results in reduces image contrast


18
New cards

variable kVp systems

  • developed by Jerman (1925)

  • first exposure “method”

  • aka variable kVp/fixed mAs technique chart

    • changes kVp based on anatomical body part thickness whereas mAs in maintained

  • used when kVp range was limited by equipment generators

    • permits finer adjustments of kVp than mA or time

  • advantages:

    • allows for smaller changes in exposure for small changes in part thickness

    • higher contrast

      • can affect perceived image resolution

  • disadvantages:

    • more patient dose

    • higher repeat rate

    • tube wear


19
New cards

historical development

  • early radiographic practice is +2 kVp for each additional cm of tissue

  • basic variable kVp formula:

    • (2 kVp x part cm) + 30 = new kVp

    • not very accurate

      • kVp often too low

      • multiple scales were needed to penetrate high-density body parts

  • mAs values determined through trial and error

    • kVp was used from scales according to patient measurements

    • compensation made for body habitus

    • known pathological conditions considered


20
New cards

anatomically programmed radiography (APR) systems

  • commonly used computerized exposure control

  • control console provides choices of anatomical parts with suggested exposure factors

  • factors can be modified for non-routine conditions

  • when combined with AEC

  • APR systems incorporate AEC devices

  • selection of exposure time (mAs) eliminated

  • must still choose mA, kVp, and distance

  • requires learning when to use and when to avoid