implementing imaging standard
image critique
requires knowledge of all areas of radiography
needs logical, organized approach
critical thinking and problem solving
critical aspect in digital imaging systems
diagnostic image
what is a diagnostic quality image?
no practical technical definition
larger number of variables involved
makes radiography art and science
requires application of scientific knowledge to analyze image
important items to remember
radiologists not trained to create and analyze quality of image
radiographers responsible for image analysis
analytical process
overview of entire image
search for differences
search for patterns and cues
narrow search
seek and formulate hypothesis
holds cues that do not fit hypothesis in reserve
identifying an imaging problem
definition of image quality must take into account clinical question
must use systematic approach
avoid just glancing at image
effective image critique method
3 step process:
classification of image
determination of cause of problem
recommendation of corrective action
classification of image
diagnostic acceptance limits
optimal diagnostic quality
images outside acceptance limits must be further critiqued
determination of cause of the problem
classified into one of three categories:
technical factors
procedural factors
equipment malfunction
technical factor problems
two types: visibility of detail and geometric problem with detail
visibility of detail problem
difficulty with visibility of image
density/image receptor exposure and contrast
image detectors receiving too little or too much radiation, high or low subject contrast
geometric problem with detail itself
recorded detail
focal spot size, SID, or OID
distortion
shape, size
procedural factor problem
it involves patient positioning or patient preparation for the x-ray examination
patient position:
tube/part/image receptor alignment
patient preparation
removal of radiopaque objects
proper bowel cleansing for GI exams
by providing clear instructions to patient
equipment malfunction
requires proper maintenance
can avoid with effective quality control program
recommendation of corrective action
critical knowledge
critiquing acceptable, but suboptimal images
repeat analysis
recording of successful results
personal technique booklet
exposure system selection
forefathers of technique:
Ed C. Jerman (1920s)
Arthur Fuchs (1940s)
radiographic exposure
very complex process
best technique systems manipulate one variable while holding others constant
goal of any exposure system:
consistency
exposure systems often incorporated into pre-programmed control consoles
suggested starting points
should be revised by radiographer when necessary
each system has advantages and disadvantages
variables change according to anatomical part thickness
important to use calipers when not utilizing AEC
calipers
devices used to measure part thickness
important QC measure when not using AEC
improper use can result in repeated exposures
part measurement
two main methods:
central ray entrance and exit
thickest part
same approach must be followed by all department radiographers
exposure systems
3 systems:
fixed kVp
variable kVp
anatomically programmed radiography
fixed kVp systems
developed by Fuchs during World War II (1943)
theoretical support from literature and professionals
aka: fixed kVp/variable mAs technique chart
selecting an optimal kVp and adjusting mAs for variations in body part
optimal kVp: a kVp that is high enough to ensure proper penetration. In the chart, it is indicated for each body part
advantages: (optimal kVp uses reduce in mAs)
reduction in patient dose
provides more information within image
increases consistency in image receptor exposure
lengthens exposure latitude
reduces x-ray tube wear
allows for shorter exposure times
thus reduces patient motion
easier to memorize
disadvantage:
higher amount of scatter results in reduces image contrast
variable kVp systems
developed by Jerman (1925)
first exposure “method”
aka variable kVp/fixed mAs technique chart
changes kVp based on anatomical body part thickness whereas mAs in maintained
used when kVp range was limited by equipment generators
permits finer adjustments of kVp than mA or time
advantages:
allows for smaller changes in exposure for small changes in part thickness
higher contrast
can affect perceived image resolution
disadvantages:
more patient dose
higher repeat rate
tube wear
historical development
early radiographic practice is +2 kVp for each additional cm of tissue
basic variable kVp formula:
(2 kVp x part cm) + 30 = new kVp
not very accurate
kVp often too low
multiple scales were needed to penetrate high-density body parts
mAs values determined through trial and error
kVp was used from scales according to patient measurements
compensation made for body habitus
known pathological conditions considered
anatomically programmed radiography (APR) systems
commonly used computerized exposure control
control console provides choices of anatomical parts with suggested exposure factors
factors can be modified for non-routine conditions
when combined with AEC
APR systems incorporate AEC devices
selection of exposure time (mAs) eliminated
must still choose mA, kVp, and distance
requires learning when to use and when to avoid