3 dimensions of each scan line
transverse, sagittal, elevational/thickness
sagittal plane artifact
axial resolution artifact
transverse plane artifact
lateral resolution artifact
elevational plane artifact
elevation/slice thickness artifact
axial resolution names
longitudinal, axial, range, radial, depth; LARRD
lateral resolution names
lateral, angular, transverse, azimuthial (LATA)
mirror artifact
fake strucure is deeper than actual structure; happens bc of reflection in a strong reflector; considered reverberation but is actually a reflection artifact
refraction artifact
both fake & real structures at the same depth (side by side); happens bc of difference in prop speed between 2 mediums
axial resolution
determined by SPL (pulse length) & wavelength; ability to differentiate structures parallel to beam (in front of each other)
lateral resolution
determined by beam width; ability to differentiate structures side by side (perpendicular to beam)
enhancement
beam doesn’t attenuate through fluid-filled structure, causing a very strong beam past it (hyperechoic, ↓att)
shadowing
happens w/ stones, calcified objects; beam gets absorbed (gyperechoic structures), leading to ↑att, weak beam, signal doesn’t go past stone
slice/elevational thickness artifact
thickness of cystic structure wall is bigger than crystal thickness, so part of the wall shows up in the structure; fix w/ harmonics or move probe; fill-in structure
speckle
noise/low-level echoes that makes pictures blurry/grainy; fix w/ harmonics shows up like birght & dark reflections/spots
harmonics
picks up noise/low-level echoes, out-of-phase echoes & fuses them together to create strong signals; double freqeuncy returns; ↑detail, ↑contrast resolution
clutter
type of noise in spectral doppler; will show at baseline; fix w/ filter
ghosting
type of noise in color doppler; flashes of color outside vessel; fix w/ filter or gain
how to avoid artifacts
move probe/change angle; use harmonics, TGC; filter; spatial compounding; freqeucny compounding; temporal compounding (persistance)
spatial compounding
steers beam, sends signals in multiple directions to avoid shadowing from bones/ribs
temporal compounding
persistance; mkaes image linger/will move in slomo; more commong in vascular; can also speed picture up
frequency compounding
uses ↑f & ↓f pulses together to get advantages of both (↑detail, ↑penetration)
how to fix slice thickness artifact
harmonics, use probe w/ higher elevation
constructive/in-phase interference
strong reflectors, appear bright; takes in-sync waves with diff amplitudes & meshes them together to create stronger echoes
destructive/out-of-phase interference
weak reflectors, dark grey/hypoechoic; takes out-of-sync waves & puts them together to create an average wave from both
reverberation
occurs w/ gas or metal structures; signal is copied over & over but eventually disappears bc of att; comet tail & ring down don’t attenuate, stays strong
mechanical probe artifacts
side lobes; fix w/ apodization or subdicing
electronic probe artifacts
grating lobes; fix w/ apodization & subdicing
subdicing
stimulates crystals in phases/gradually
apodization
reduces side crystal strength & redirects it to middle crystals
artifacts caused by false assumptions
speed error artifact, refraction, range ambiguity
speed error artifact
machine assumes beam always stravels at 1.54 mm/usec but it sligthly varies; a structure appears cut off/seperated; slower speed = appears deeper; faster speed = appears closer
refraction assumption
machine assumes beam travels only in a straight line; makes a copy of structure next to actual structure bc signal got refracted but still returns to transducer
range ambiguity artifact
machine sends 2nd pulse before recieving 1st pulse; when 1st pulse finally returns, machine assumes it is the 2nd pulse returning fast & places structure closer than it actually is
organ texture
organs are smooth; soeecjle doesn’t accurately represent texture, which is why its an artifact