Photographic Qualities
The photographic properties of an x-ray image that determine the visibility of detail, specifically brightness and contrast.
Geometric Qualities
The geometric properties of an x-ray image that govern spatial resolution (sharpness) and distortion (veracity of the image).
Contrast Resolution
The ability of an imaging system to display many different subject contrasts, enabling the visual differentiation of soft tissues such as the liver and kidneys on an abdominal radiograph (KUB).
Scatter Radiation
Unwanted exposure to the image receptor caused by Compton interactions that produces fog, provides no useful diagnostic information, and decreases image contrast.
Remnant Beam
The beam of image-forming x-rays that exits the patient to reach the image receptor, consisting of directly transmitted photons and Compton-scattered photons.
Influencing Factors of Scatter
The three primary variables that govern the production of scatter radiation and noise: kVp, field size, and patient thickness.
Collimation
The restriction of the useful x-ray beam field size, primarily used to reduce patient dose and secondarily used to reduce Compton scatter interactions to improve contrast resolution.
Tissue Compression
The mechanical reduction of anatomical part thickness, which decreases scatter production, lowers patient dose, and improves contrast resolution (such as in mammography and PA vs. AP abdomen).
Aperture Diaphragm
The simplest type of beam restrictor, consisting of a flat sheet of lead placed below the tube port with a fixed aperture designed for a specific image receptor size.
Cylinder / Cone
An extended circular metal beam-restricting device attached to the tube port whose extended end determines the diameter of the primary beam.
Cone Cutting
An artifact resulting from the mechanical misalignment of the x-ray tube, cone or cylinder, and image receptor, causing unintended cutoff of the primary beam.
Variable-Aperture Collimator
The most widely used beam-restricting device, equipped with two to three sets of adjustable lead shutters to produce variable square and rectangular radiation field sizes.
Off-Focus Radiation
Radiation produced by electrons striking areas of the x-ray tube outside the focal spot; controlled inside a collimator by the first-stage entrance shutters.
Collimator Light Field Coincidence
A regulatory compliance standard requiring the visible light localization field to match the actual x-ray beam within a variance of no more than 2% of the SID.
Positive Beam Limitation (PBL)
An automatic collimation system mandated in 1974 and deregulated in 1994 that senses the cassette size and automatically restricts the beam to match the image receptor dimensions.
Radiographic Grid
A scatter-cleanup device invented in 1913 by Gustav Bucky, measuring approximately 41 inch thick, that cleans up 80% to 90% of scatter before it reaches the image receptor.
Clinical Indications for Grid Use
Guidelines recommending grid use whenever the anatomical part thickness exceeds 10cm or when exposure techniques exceed 60kVp.
Interspace Material
The radiolucent section of a grid, made of aluminum or plastic fiber, that maintains precise spacing between adjacent radiopaque lead strips.
Grid Frequency
The number of lead grid lines per unit length, typically ranging from 25 to 45lines/cm.
Grid Ratio
The ratio of the height of the lead strips (h) to the distance separating them (D), defined mathematically as Grid Ratio=Dh.
Linear Grid
A grid design in which lead strips run parallel in only one longitudinal direction, allowing tube angulation along the length of the strips without causing cutoff.
Crossed (Cross-Hatch) Grid
A grid constructed by placing two linear grids together at right angles, providing superior scatter cleanup but prohibiting any central ray angulation.
Focused Grid
A grid designed with lead strips angled progressively toward the edges to match the divergence of the primary x-ray beam, thereby minimizing grid cutoff.
Convergence Line
An imaginary line formed by connecting the convergent points of angled lead strips along the length of a focused grid, which determines its focal distance.
Focal Range
The specific range of source-to-image distances (SIDs) over which a focused grid can be utilized without producing off-focus grid cutoff.
Parallel Grid
A grid whose lead strips are strictly parallel to each other throughout, possessing no angled strips, an infinite focal distance, and no specific focal range limitations.
Stationary Grid
A non-moving grid type, including wafer grids taped directly over the IR, grid cassettes with built-in front grids, and grid caps that allow the IR to slide behind them.
Potter-Bucky Diaphragm
A moving grid mechanism introduced in 1920 by Hollis E. Potter that moves during an exposure to blur out visible grid lines on the resulting image.
Reciprocating Grid
A motor-driven Bucky grid that shifts back and forth approximately 2cm perpendicular to the lead lines during an exposure to eliminate grid line visibility.
Oscillating Grid
An electromagnet-driven moving grid that moves in circular orbits of 2 to 3cm on all sides for 20 to 30seconds during an exposure.
Grid Conversion Factor (GCF) / Bucky Factor
A technical multiplier used to adjust mAs when adding, removing, or changing grid ratios, governed by the formula GCF2GCF1=mAs2mAs1.
Contrast Improvement Factor (k)
A numerical ratio describing a grid's effectiveness at increasing contrast, expressed as k=image contrast without gridimage contrast with grid, usually falling between 1.5 and 2.5.
Grid Cutoff
The unintended loss of primary transmitted photons reaching the image receptor caused by improper alignment or angling of the x-ray beam relative to the lead strips.
Off-Level Grid Cutoff
The most common grid cutoff error, occurring when the tube or grid is tilted across the lead strips, resulting in an overall uniform loss of exposure across the entire image.
Off-Center Grid Cutoff
A cutoff error occurring when the central ray is laterally displaced away from the center of a focused grid, causing an overall loss of exposure across the image.
Off-Focus Grid Cutoff
A cutoff error that occurs when the SID is positioned outside the specified focal range (either too close or too far), leading to underexposure along the peripheral edges of the radiograph.
Upside-Down Focused Grid Cutoff
A severe positioning error where a focused grid is oriented backward toward the tube, causing the beam divergence to oppose the strip angle, resulting in severe lateral edge cutoff while the center remains exposed.
Air-Gap Technique
A scatter reduction method utilizing a 10 to 15cm gap between the patient and image receptor; a 10 inch air gap yields scatter cleanup comparable to a 15:1 grid but increases magnification and requires an mAs increase of 10% per centimeter.
Virtual Grid
A digital image-processing algorithm that computationally models, estimates, and subtracts scatter noise to improve contrast without increasing patient dose or requiring physical grids.
As Grid Ratio ————, Likelihood of Grid Cutoff ————
increases; increases
What is the most common type of grid cutoff?
Off-level
There is a __% increase in kVp required when going from a stationary grid to a moving grid
15%
A 10” air gap has similar clean-up of a __:__ grid
15:1
When using the air-gap technique, mAs is increased by __% for every cm
10%