Comprehensive vocabulary flashcards covering image quality, scatter production, beam-restricting devices, grid design, alignment errors, and technical conversions from the RADT 1065 lecture.

44 Terms

1
New cards

Photographic Qualities

The photographic properties of an x-ray image that determine the visibility of detail, specifically brightness and contrast.

2
New cards

Geometric Qualities

The geometric properties of an x-ray image that govern spatial resolution (sharpness) and distortion (veracity of the image).

3
New cards

Contrast Resolution

The ability of an imaging system to display many different subject contrasts, enabling the visual differentiation of soft tissues such as the liver and kidneys on an abdominal radiograph (KUB).

4
New cards

Scatter Radiation

Unwanted exposure to the image receptor caused by Compton interactions that produces fog, provides no useful diagnostic information, and decreases image contrast.

5
New cards

Remnant Beam

The beam of image-forming x-rays that exits the patient to reach the image receptor, consisting of directly transmitted photons and Compton-scattered photons.

6
New cards

Influencing Factors of Scatter

The three primary variables that govern the production of scatter radiation and noise: kVpkVp, field size, and patient thickness.

7
New cards

Collimation

The restriction of the useful x-ray beam field size, primarily used to reduce patient dose and secondarily used to reduce Compton scatter interactions to improve contrast resolution.

8
New cards

Tissue Compression

The mechanical reduction of anatomical part thickness, which decreases scatter production, lowers patient dose, and improves contrast resolution (such as in mammography and PA vs. AP abdomen).

9
New cards

Aperture Diaphragm

The simplest type of beam restrictor, consisting of a flat sheet of lead placed below the tube port with a fixed aperture designed for a specific image receptor size.

10
New cards

Cylinder / Cone

An extended circular metal beam-restricting device attached to the tube port whose extended end determines the diameter of the primary beam.

11
New cards

Cone Cutting

An artifact resulting from the mechanical misalignment of the x-ray tube, cone or cylinder, and image receptor, causing unintended cutoff of the primary beam.

12
New cards

Variable-Aperture Collimator

The most widely used beam-restricting device, equipped with two to three sets of adjustable lead shutters to produce variable square and rectangular radiation field sizes.

13
New cards

Off-Focus Radiation

Radiation produced by electrons striking areas of the x-ray tube outside the focal spot; controlled inside a collimator by the first-stage entrance shutters.

14
New cards

Collimator Light Field Coincidence

A regulatory compliance standard requiring the visible light localization field to match the actual x-ray beam within a variance of no more than 2%2\% of the SID.

15
New cards

Positive Beam Limitation (PBL)

An automatic collimation system mandated in 19741974 and deregulated in 19941994 that senses the cassette size and automatically restricts the beam to match the image receptor dimensions.

16
New cards

Radiographic Grid

A scatter-cleanup device invented in 19131913 by Gustav Bucky, measuring approximately 14 inch\frac{1}{4}\text{ inch} thick, that cleans up 80%80\% to 90%90\% of scatter before it reaches the image receptor.

17
New cards

Clinical Indications for Grid Use

Guidelines recommending grid use whenever the anatomical part thickness exceeds 10 cm10\,\text{cm} or when exposure techniques exceed 60 kVp60\,\text{kVp}.

18
New cards

Interspace Material

The radiolucent section of a grid, made of aluminum or plastic fiber, that maintains precise spacing between adjacent radiopaque lead strips.

19
New cards

Grid Frequency

The number of lead grid lines per unit length, typically ranging from 2525 to 45 lines/cm45\,\text{lines/cm}.

20
New cards

Grid Ratio

The ratio of the height of the lead strips (hh) to the distance separating them (DD), defined mathematically as Grid Ratio=hD\text{Grid Ratio} = \frac{h}{D}.

21
New cards

Linear Grid

A grid design in which lead strips run parallel in only one longitudinal direction, allowing tube angulation along the length of the strips without causing cutoff.

22
New cards

Crossed (Cross-Hatch) Grid

A grid constructed by placing two linear grids together at right angles, providing superior scatter cleanup but prohibiting any central ray angulation.

23
New cards

Focused Grid

A grid designed with lead strips angled progressively toward the edges to match the divergence of the primary x-ray beam, thereby minimizing grid cutoff.

24
New cards

Convergence Line

An imaginary line formed by connecting the convergent points of angled lead strips along the length of a focused grid, which determines its focal distance.

25
New cards

Focal Range

The specific range of source-to-image distances (SIDs) over which a focused grid can be utilized without producing off-focus grid cutoff.

26
New cards

Parallel Grid

A grid whose lead strips are strictly parallel to each other throughout, possessing no angled strips, an infinite focal distance, and no specific focal range limitations.

27
New cards

Stationary Grid

A non-moving grid type, including wafer grids taped directly over the IR, grid cassettes with built-in front grids, and grid caps that allow the IR to slide behind them.

28
New cards

Potter-Bucky Diaphragm

A moving grid mechanism introduced in 19201920 by Hollis E. Potter that moves during an exposure to blur out visible grid lines on the resulting image.

29
New cards

Reciprocating Grid

A motor-driven Bucky grid that shifts back and forth approximately 2 cm2\,\text{cm} perpendicular to the lead lines during an exposure to eliminate grid line visibility.

30
New cards

Oscillating Grid

An electromagnet-driven moving grid that moves in circular orbits of 22 to 3 cm3\,\text{cm} on all sides for 2020 to 30 seconds30\,\text{seconds} during an exposure.

31
New cards

Grid Conversion Factor (GCF) / Bucky Factor

A technical multiplier used to adjust mAs when adding, removing, or changing grid ratios, governed by the formula GCF1GCF2=mAs1mAs2\frac{\text{GCF}_1}{\text{GCF}_2} = \frac{\text{mAs}_1}{\text{mAs}_2}.

32
New cards

Contrast Improvement Factor (kk)

A numerical ratio describing a grid's effectiveness at increasing contrast, expressed as k=image contrast with gridimage contrast without gridk = \frac{\text{image contrast with grid}}{\text{image contrast without grid}}, usually falling between 1.51.5 and 2.52.5.

33
New cards

Grid Cutoff

The unintended loss of primary transmitted photons reaching the image receptor caused by improper alignment or angling of the x-ray beam relative to the lead strips.

34
New cards

Off-Level Grid Cutoff

The most common grid cutoff error, occurring when the tube or grid is tilted across the lead strips, resulting in an overall uniform loss of exposure across the entire image.

35
New cards

Off-Center Grid Cutoff

A cutoff error occurring when the central ray is laterally displaced away from the center of a focused grid, causing an overall loss of exposure across the image.

36
New cards

Off-Focus Grid Cutoff

A cutoff error that occurs when the SID is positioned outside the specified focal range (either too close or too far), leading to underexposure along the peripheral edges of the radiograph.

37
New cards

Upside-Down Focused Grid Cutoff

A severe positioning error where a focused grid is oriented backward toward the tube, causing the beam divergence to oppose the strip angle, resulting in severe lateral edge cutoff while the center remains exposed.

38
New cards

Air-Gap Technique

A scatter reduction method utilizing a 1010 to 15 cm15\,\text{cm} gap between the patient and image receptor; a 10 inch10\text{ inch} air gap yields scatter cleanup comparable to a 15:115:1 grid but increases magnification and requires an mAs increase of 10%10\% per centimeter.

39
New cards

Virtual Grid

A digital image-processing algorithm that computationally models, estimates, and subtracts scatter noise to improve contrast without increasing patient dose or requiring physical grids.

40
New cards

As Grid Ratio ————, Likelihood of Grid Cutoff ————

increases; increases

41
New cards

What is the most common type of grid cutoff?

Off-level

42
New cards

There is a __% increase in kVp required when going from a stationary grid to a moving grid

15%

43
New cards

A 10” air gap has similar clean-up of a __:__ grid

15:1

44
New cards

When using the air-gap technique, mAs is increased by __% for every cm

10%