What is the function of an anesthesia machine
delivers inhalation anesthesia and oxygen while removing gases
What are the 3 things an anesthesia machine must do
deliver oxygen at a controlled rate, vaporize liquids to gas, and removes exhaled gas
What is very important to insure proper function of an anesthesia machine
Maintenance
What are the components of an anesthesia machine
oxygen storage
pressure reducing valve
flowmeter
flush valve
vaporizer
inspiratory valve
negative pressure valve
corrugated breathing tube
expiratory tube
adjustable pressure relief valve
manometer
rebreathing bag
carbon dioxide absorber
scavenging system
What are the components of the oxygen source
Cylinder, two types of tanks E and H, color coated green, at 2200psi pressure, 100% o2, have a yoke and tank pressure gauge
What is the difference between an E and H oxygen tank
E is smaller and attached to the machine, H is larger and chained to a wall
What is the yoke part of a oxygen tank
a U-shaped metal clamp or connector that attaches a pressure regulator or flowmeter directly to the valve of a medical oxygen cylinder
What does the pressure reducing valve do
reduces pressure leaving the oxygen tank to 40-50 psi
What does the flowmeter do
reduces pressure to 15 psi, regulates o2 flow, (the floating ball tells you how much oxygen is given in L/min or ml/min)
What should you avoid with the flowmeter
Excess tightening
What is the flush valve
flushes valve, pure o2 only and vaporizer is bypassed
When should you not use the flush valve
Do not use with non-rebreather and do not push with pop off valve closed
What are the important function of the vaporizer
Temp, gas flow rate, back pressure
What are the parts of a vaporizer
inlet- o2 enters, outlet- o2 and anesthetic gas exits (circular system)
What are the vaporizers other functions
Holds liquid anesthetic, controlled delivery, is calibrated, precision v.s non-precision
what should be done with the vaporizer for good maintenance
serviced annually or per manufacturer recommendations and only use the liquid anesthetic the machine is calibrated for
What are the types of unidirectional valves
Inspiratory/expiratory, unidirectional, flutter valve (inhale/exhale)
what does an unidirectional valve do
insures one way direction of gas flow
What does the negative pressure relief valve do
Safety device that allows room air to enter the system
what is corrugated tubing/y piece
tubing comes together to form the y-piece
What is the weight range for a shorter/smaller diameter tubing
7-20kgs
What is the weight range for a longer/larger diameter tubing
>20kgs
What is the function of the Adjustable pressure relief valve (pop-off valve)
in the open postion it prevents pressure build up
What are the 3 flow rate techniques for the pop-off valve
wide, partial, and closed when bagging only
What is the most important thing to remember about the pop-off valve
The patient can not breath when the pop-off valve is closed
What are the functions of the manometer
pressure gauge and reads pressure in the system (cmH2O or mmHg units)
What is the normal resting pressure with the pop-off valve completely open
should be zero on the manometer
What pressure should you never exceed on the manometer when bagging a patient
20cmH2O
What are the function of the rebreathing/reservoir bag
patients re-breath exhaled gas and the bag deflates on inspiration and inflates on expiration
How do you find tidal volume
10ml/kg
What are 2 equations you can use to find the rebreathing bag/reservoir bag size
5x tidal volume or 30 x weight in pounds
what is the function of the carbon dioxide absorber
exhaled CO2 goes to the CO2 absorber or scavenger to track usage with ph indicator from a chemical reaction caused by the heat and water produced from the Co2 changing the color
What are the 2 types of CO2 absorbers
baralime (barium hydroxide lime) or sodalime (sodium hydroxide lime)
How much space does the carbon dioxide absorber leave
½ inch in air space at the top
what are the functions of the scavenging system
evacuates waste gas through active or passive systems
what is the difference between active and passive scavenging systems
Active is mechanical usually through a vacuum to the outside v.s passive that relies on gravity (f-air canister or hose to outside wall vent)
When should you leak test an anesthesia machine
before every procedure, after a hose or bag change, after changing the co2 absorber
How do you run a leak test
close the pop-off valve and occlude the breathing circuit then flush to 20-40mmHg making sure to observe the meter
why is leak testing important
reduces and eliminates gas leaks into the environment
what are the 2 types of Breathing circuits
rebreathing and nonrebreathing
rebreathing circuit
recirculation of expired gas, should be used in animals >7kg, universal f-circuit, corrugated hose
Non rebreathing circuit
no gas is rebreathed, should be used in animals <7kg, has low gas resistance and high oxygen flow, bypasses CO2 absorber and manometer, the patient can breath normally and receive fresh, new gas with minimal effort
Rebreathing oxygen flow rate
Induction: 50-100mL/kg/min
• Maintenance: 20-40mL/kg/min
• Recovery: 50-100mL/kg/min
Just need enough to match tidal
volume
Non rebreathing oxygen flow rate
200ml/kg/min for all phases
Precision vs. non-precision vaporizer
precision is temp controlled so the higher the temp the more gas is made
Flow of the anesthesia machine (Andrew)
O2 from tank goes to pressure release valve to flow meter to oxygen flush valve to vaporizer to unidirectional valve to inspiratory valve to expiatory valve to negative pressure relief valve, through corrugated tubing and y piece, to expartory tube, to adjustable pressure relief valve (pop off valve), nanometer, revisor bag, soda lime, scavenger system