31 Terms

1
New cards

What permits the gantry frame to rotate continuously?

Slip rings

2
New cards

Removes low energy x-ray photons, minimizes patient exposure, and produces a more uniform beam

Filtering

3
New cards

What restricts the x-ray beam to a specific area and reduces scatter radiation?

Collimation

4
New cards

How efficient are xenon gas detectors?

60-87%

5
New cards

How efficient are solid state detectors?

99-100%

6
New cards

What is the average gantry diameter?

70 cm

7
New cards

A pencil beam is a characteristic of which generation CT scanner?

1st generation

8
New cards

A fan beam is a characteristic of which generation CT scanner?

2nd, 3rd, 4th, and 6th generation

9
New cards

A cone beam is a characteristic of which generation CT scanner?

7th generation

10
New cards

What is the CT table primarily made up of?

Carbon graphite

11
New cards

Which generation CT scanner was a specific cardiac scanner?

5th generation

12
New cards

Axial scanning is used on which CT scanner generation?

1st, 2nd, 3rd, and 4th generation

13
New cards

Volumetric scanning is used on which CT scanner generation?

6th and 7th generation

14
New cards

What do we ideally want our pitch to be?

1

15
New cards

If the pitch is greater than 1, what occurs?

Under sampling, reduced exposure, missing data information

16
New cards

What are the two types of slip rings?

High and low voltage

17
New cards

What does scatter radiation do?

Reduces image quality and increases patient radiation dose

18
New cards

What is another name for a solid-state detector?

Scintillation detector

19
New cards

The process of obtaining information to create images is known as what?

Data acquisition

20
New cards

What are the two common focal spot sizes?

.5 mm and 1 mm

21
New cards

What does the generator do?

Produces high voltage and transmits it to the x-ray tube

22
New cards

What are the two types of collimators?

Pre-patient collimators and post-patient collimators (pre-detector)

23
New cards

A group of detectors is known as what?

Detector array

24
New cards

What is determined by collimation (beam width) and detector configuration?

Slice thickness

25
New cards

The ratio of the distance the table travels per 360 degree rotation is known as what?

Pitch

26
New cards

What is the pitch formula?

Pitch= I/S(B)

I= table increment (mm/sec.)

S= slice thickness (mm)

B= beam collimation

27
New cards

What determines the range of exposure techniques?

The power capacity of the generator

28
New cards

The table movement per (360 degree) rotation divided by beam width is known as what?

Beam pitch

29
New cards

What is the beam pitch formula?

Beam pitch= I/BW

I= table increment (mm/sec.)

BW= Beam width

30
New cards

How do you find the beam width?

Beam width= number of slices per rotation X slice thickness

31
New cards

If the pitch is less than 1, what occurs?

Over sampling, increased patient dose