Patient Protection Before Exposure
• Proper prescribing of dental images
• The use of equipment that complies with provincial & federal radiation guidelines.
When should image (x-ray) be taken?
only prescibe when its necessary
Who can prescribe dental radiograph?
Dentist
Hygienist
Denturist
How many images should a patient require?
The # of films and frequency should be assessed on an individual basis.
There is no predetermined number!
What are Guidelines for Prescribing Radiograph
New or a recall patient?
Risk factors for caries?
History of periodontal disease?
Assess growth and development?
Age of the patient?
• Primary, mixed or permanent dentition?
• Dentulous or edentulous?
Requirements For Dental X-ray Equipment (protecting the patient prior to exposure
Filtration
Collimation
PID
Filtration
The function of the aluminum filter is to filter out soft x-rays.
Those soft x-rays are harmful to the patient and do not add any diagnostic value to the image.
Filtration results in a higher energy and more penetrating beam.
What is the requirements for filtration?
hinger than 70 kvp you need 2.5 mm pf aluminum
70 kvp or low you need 1.5 mm of aluminum
Collimation
The lead collimator restricts the size and shape of the x-ray beam
which reduces the patient’s exposure
A rectangular collimator exposes ___ less tissue than a circular collimato
60%
Requirements from Health Canada for collimation is to restrict the beam size to:
7cm for the circular collimation.
38.5cm2 for the rectangular collimation.
What PID length is preferred?
longer is preferred because of less divergence
Patient protection during exposure
• Thyroid collar & Lead apron
• Beam Alignment Devices
• Exposure factors
• Proper techniques
What protects the person form scatter radiation?
Thyroid Collars and Lead Aprons
Beam Alignment Devices / Receptor Holder
Stabilize - less chance for movement.
Patient does not have to hold the recepto
Exposure factors
• 60-80 kV
• 6-8 mA
• Shortest exposure time possible to create a diagnostic image
What are the proper techniques?
The radiographer must be knowledgeable in all the techniques.
Being organized reduces the chances of a double exposure or a retake (additional exposure to the patient
Protecting the Operator
Avoid the primary beam:
Never hold a receptor in place for a patient.
Never hold the tubehead during exposure
In Canada, according to Safety Code 30, the minimum distance
is?
3 meters
Radiation Monitoring
Dental x-ray equipment in Alberta must be registered with the
Alberta Dental Association & College (ADA&C)
On-site radiation monitoring must be done by qualified service
personnel an Authorized Radiation Protection Agency.
(There are 4 ARPA in Alberta)
Once equipment is registered, full compliance verification is
required every 5 years.
Each owner must provide the ADA & C annual confirmation
of facility registration, including evidence of dosimetry service
and management of Quality Assurance.
Failure to meet the requirements of this program may result in
suspension of the practice!
What is the ONLY way to monitor personnel exposure to ionizing radiation
Dosimeters
Exposure Limits
MPD = maximum permissible dose
✓ for radiation workers is 50 mSv/year.
✓ for pregnant radiation workers is 4 mSv (during pregnancy)
✓ For general public or students in training is 1 mSv/year
Radiation Safety Legislation
◼Federal
Safety Code 30 (Health Canada)
◼Provincial
Radiation Protection Act
Radiation Protection Regulation
What is the ALARA principle?
As Low As Reasonably Achievable