Electrophoresis

Migration of charged solutes or particles in a liquid medium under the influence of an electric field, separation technique used primarily for separating and identifying serum proteins in blood

2
New cards

Iontophoresis is the

migration of small ions

3
New cards

zone electrophoresis is the

most common clinical application, which is used to seperate macromolecules, such as proteins, in a porus support medium it requires: charged molecules migrates as zones, migration takes place across a support medium, sample is mixed with a buffer, separated protein zones are generated on the support media known as electropherogram, zones are quantified using a desitometer

4
New cards

The ability to separate proteins is sbed on the caharacteristics of

amino acids, positive ions (cations) will migrate towards a cathode, negative ions (anions) will migrate toward an anode

5
New cards

Ampholyte (Zwitterion)

molecule that can become either negatively or positively charged based on the pH of the solution in which it resides

6
New cards

isoelectric point

the pH at which a molecule has a net charge of 0

7
New cards

An ampholyte will take a positive charge in a solution more acidic than its isoelectric point (pI) and

migrates toward the cathode

8
New cards

An ampholyte will take a negative charge in a solution more alkaline than its isoelectric point (pI) and

migrates toward the anode

9
New cards

Proteins migrates through gel from cathode to anode (+) and are separated into five bands

albumin, alpha 1 globulins, alpha 2 globulins, beta globulin, gamma globulin

10
New cards

Migratory inhabitation factor phenomenon

at a pH at 8.6, the gamma globulins move toward the cathode, despite the fact that they are negatively charged

11
New cards

rate of migration depends on

net electrical charge of the molecule, size and shape of the molecule, electrical field strength, propeties of supporting medium, temperature of the operation

12
New cards

The isoelectric point of an amino acid or protein can be defined as

pH at which the amino acid or protein has no net charge

13
New cards

proteins are separated into how many fractions in electrophoresis

five

14
New cards

In electrophoresis, the greater the charge of the molecule

the faster it will migrate

15
New cards

Power supply of the Electrophoresis apparatus

best to use a supply with a constant current opposed to constant voltage which can cause joule heat: flow of current through a medium that offers resistance

16
New cards

Work flow of Electrophoresis apparatus

caused by the movement of buffer into the support medium, moisture and heat generation during electrophoresis causes movement of the buffer into the gel, using a lead can prevent some of the loss of evaporation, methods that use excessive heat need a cooling system

17
New cards

Buffer of Electrophoresis apparatus

carries the applied current, establishes the pH at which electrophoresis is performed, determines the electrical charge on the solute, Ionic strength: the sum of all charges present, negative or positive, increasing the ionic strength will increases conductance of the support, size of the ionic cloud surrounding a charged molecule, yeilds slower migration rates and shaper band separation

18
New cards

Support media of Electrophoresis apparatus

cellulose acetate, agarose gel, poluacrylamide gel, starch gel, automated systems

19
New cards

Cellulose acetate

dry, brittle film composed of about 80% air space, used in isoelectric focusing

20
New cards

Agarose gel

used for separation of proteins in serum, urine, CSF, hemoglobin variants, isoenzymes, lipoproteins

21
New cards

Polyacrylamide gel

thermostable, transparent, strong, relatively chemically inert and uncharged, proteins are separated on the basis of molecular sleiving: separation according to both charge-to-mass ratio and molecular size

22
New cards

The component that carries the applied current and establishes the pH at which electrophoresis is preformed

buffer

23
New cards

what is the importance of the ionic strength of the buffer used in gel electrophoresis

a heat labile protein will become denatured if the ionic strength of the buffer is too high

24
New cards

the best support medium to use because it is thermostable, strong, transparent, and eliminates endomosis due to the lack of charge

polyacrylamide

25
New cards

Staining detection

Serum proteins in general: amido black, coomassie brillant blue, ponceu S, Isoenzymes: formazan, nitrotrezolium blue, liportein zones: fat red, oil red O, sudan black B, DNA fragments: ethidium bromide, CSF porteins: silver nitrate

26
New cards

Endomosis

prefernetial movement of water in one direction through an electrophoresis medium, low in AGE and PAGE, occurs due to selective binding of one type of charge on the surface of the medium

27
New cards

Electrolysis

chemical decomposition produced by passing an electric current through a liquid or solution containing ions

28
New cards

common problems in electrophoresis

discontinuites, unequal migration of samples across the gel, distorted protein zones, ireegularities in sample application, hemolyzed specimens, plasma instead of serum, split albumin

29
New cards

unequal migration of samples down the gel in an electrophoresis assay in an interference usually caused by

dirty electrodes causing uneven applications of the electrical field

30
New cards

A split albumin zone is likely due to

a patient having more than one type of albumin (bis-albuminina)

31
New cards

Isoelectric focusing: type of electrophoresis

the protiens migrate across a gel until they reach a zone where the pH of the gel matches the isoelectric point (pI) of the protein, the technique separates proteins in a pH gradient unti pH = pI

32
New cards

Immunofixation (IFE): type of electrophoresis

detects specific types of immunoglobulins or gamma and kappa chains

33
New cards

CApillary: type of electrophoresis

same as basic but tube is connected to a detector at its terminal end and to a high-voltage power supply via buffer reservoirs

34
New cards

Two-dimensional electrophoresis

1st: charge dependent isoelectric focusing 2nd: molecular weight-dependent electrophoresis, proteomics: the identification and quantification of proteins and their posttranslational modifications in a given system

35
New cards

Osmosis

process the constitutes movement of solvent across a membrane in response to difference in osmotic pressure across the two sides of membranes

36
New cards

Osmotic pressure

pressure required to stop osmosis through a semipermeable membrane between a solution and pure solvent, governs the movement of solvent (water) across membranes that separate two solutions

37
New cards

Osmole

1 gram (g) molecular weight (MW) of a substance divided by the number of particles into which it dissociates

38
New cards

Osmolality

Expresses concentrations related to mass of solvent

39
New cards

osmolarity

expresses concentration per volume of solution

40
New cards

major osmotic substances in normal plasma

sodium, chloride, glucose, urea (BUN)

41
New cards

Osmolality equation

mOsm/kg = 2[Na] + glucose/18 + urea/2.8

42
New cards

osmometry

a technique that measures the concentration of dissolved solute particles in a solution that contribute to the osmotic pressure

43
New cards

A solute dissolved in solvent causes the following changes in colligative properties

increased osmotic pressure, lowered vapor pressure, increased boiling point, lowered freezing point

44
New cards

freezing point osmometers

measurement of the freezing point of solution by supercooling the specimen to -7C

45
New cards

vapor pressure osmometers

correlation studies have shown more variation with vapor pressure osmometers, so freezing point osmometers are more commonly used

46
New cards

Chromatography

a group of separation techniques that separate analytes by differential distribution between a stationary phase and a mobile phase

47
New cards

Stationary phase

a solid or a liquid phase that interacts with components of the mobile phase

48
New cards

mobile phase

a gas or a liquid that flows in a chromatographic system and carries the sample past the stationary phase

49
New cards

as mobile phase flows past the stationary phase, the solutes may

reside only on the stationary phase (no separation), reside only in the mobile phase (migration with the mobile phase), distribute between the two pahses (Differential migration)

50
New cards

basis of separation

those solutes with high affinity for the stationary phase move slower, those with low affinity for the stationary phase will migrate faster

51
New cards

Planar chromatography

stationary phase is on a thin support such as paper or solid surface, used in identification of drugs in urine, analysis of amino acids and lipids in amniotic fluid

52
New cards

migration of a solute is expressed by its retention factor (Rf) value, the ratio of solute migration to solvent front migration

Rf = distance from application point to solute center/distance from application point to mobile phase front

53
New cards

which phase of chromatography carries the sample

mobile phase

54
New cards

thin layer and paper chromatography are two types of

Planar chromatography

55
New cards

Column chromatography

stationary phase is packed into a tube or coated on the inner surface of the tube, two types: gas and liquid chromatography, both can be connected to a mass spectrometer to identify the separated solutes

56
New cards

Gas chromatography

a gas mobile phase, or a carrier gas, is used to carry a mixture of volatile solutes through a column containing the stationary phase ex: nitrogen, helium, argon, sample must be volatile, stationary phase usually is a nonvolatile liquid coated or bonded to particles or the inner surface of the capillary, the carrier gas carries analytes to a detector in order of their elution and identified qualitatively by their renention time or by mass spec

57
New cards

liquid chromatography

separation is based on distribution of solutes between a liquid mobile phase and a stationary phase, High performance LC: stationary phase in LC consists of small-diameter particles and needs high pressure to pump liquid through column

58
New cards

Detector

a device that responds to the presence of analyte in the mobile phase, the magnitude of which is used to identify and quantify analytes, which produces an electronic singal that is plotted as a function of: time, distance or volume the resulting display is a chromatogram

59
New cards

Chromatogram

a plot of detector response used in column chromatography to the presence of analyte in the mobile phase as a function of time or mobile phase volume

60
New cards

retention time

time interval between specimen injection and solute reaching the detector

61
New cards

Resolutions (Rs)

measure of chromatographic separation, requires that two peaks have different elution times for the peak centers and sufficiently narrow bandwidths to eliminate or minimize overlap, incomplete separation occurs when the calculated value is Rs < 0.8, baseline separation is obtained when Rs is > 1.25

62
New cards

which component of a gas chromatograph introduces an aliquot of sample to be analyzed into the column

injector

63
New cards

incomplete separation occurs when the calculate value for Rs

is less then 0.8

64
New cards

Ion exchange chromatography: Cation-exchange particles

contain negatively charged groups in the stationary phase that binds positively charged components (cations) in the mobile phase

65
New cards

Ion exchange chromatography: anion-exchange particles

contain positively charged groups in the stationary phase that binds negatively charged components (anions) in the mobile phase

66
New cards

Ion exchange separates

amino acids, glycated hemoglobin, hemoglobin variants, oligonucleotides

67
New cards

Size exclusion chromatography: variety of materials are used as a stationary phase for size exclusion

Dextran, poluacrylamide, agarose, polystyrene-divinlybenzene, porous beads

68
New cards

Affinity chromatography

one component of highly specific molecular interaction pairs is immobilized in a stationary phase and is used to capture molecules from the mobile phase, interactions pairs are strong, highly specific and may include: enzyme inhibitor, hormone-receptor, antigen-antibody, aptamer-ligand

69
New cards

Qualitative and Quantitative analysis of chromatography: external calibration

reference solutions containing known quantities of analytes are analyzed in a separate run to establish detector responses per amount of analytes

70
New cards

Qualitative and Quantitative analysis of chromatography: internal calibration

reference solution of known analyte concentrations are prepared, and a constant amount of a different compound, the internal standard is added to each reference solution and to each sample

71
New cards

The differential distribution of solutes between two immiscible liquids is the basis for separtion for

partition

72
New cards

In column chromatoprahy, unknown analytes are identified as they pass through the detector, comparision of the ______ of the unknown analytes with that of the internal standard is the method used for unknown identification

Retention time

73
New cards

Mass spectrometry

the instrument produces a mass spectrum: the relative, abundance of each ion is plotted as a function of its mass-charge-ratio

74
New cards

The mass spectrometer displays data in 2 ways:

Mass spectrum: signal intensity vs m/z or Chromatogram: signal intensity vs time

75
New cards

Mass spectrometery does not measure mass but a m/z ratio

m/z = (M plus or minus n)/n

76
New cards

Mass analysis

the process by which ions are identified according to m/z ratios

77
New cards

Resolution (m/delta m)

the mass of a compound divided by its peak width at half maximum height (FWHM)

78
New cards

mass accuracy

how accurate the instrument is by comparing measured mass assignment for an ion to the expected mass

79
New cards

Sensitivity

when concentration of an analyte can be detected

80
New cards

time of flight

flight time: time to traverse the drift tube, t w sq rt(m/z)

81
New cards

Faraday cup

Ion abundance is so high that it would saturate the output of an electron multiplier

82
New cards

Migration rate in electrophoresis depends largely on the ratio of

net charge to size

83
New cards

SPE fractions at pH 8.6, fasted to slowest

Albumin, a1, a2, B, y

84
New cards

Disc electrophoresis is best described as separation using a

Discontinuous buffer system (stakingand separting gel)

85
New cards

Two dimensional electrophoresis separates by a column then by

molecular weight

86
New cards

method that uses antisera to type monoclonal immunoglobulins?

Immunofixation electrophoresis

87
New cards

A Northern blot detects

fragment RNA

88
New cards

A split (two headed) albumin zone is seen in

Bis-albuminemia

89
New cards

Separation by relative solubility in two immiscible liquids is

partition chromatography

90
New cards

Hydrogen bonding and hydrophobic interactions on a solid surface drive

adsorption chromatography

91
New cards

Porus beads separate solutes by size in solutions, this is an example of

size exclusion (gel filteration)

92
New cards

Immobilized antibody capturing its antigens is an example of

Affinity chromatography

93
New cards

example of a change that improves column efficiency

using smaller stationary phase particles

94
New cards

what change alters selectivity rather than efficiency

changing mobile phase composition

95
New cards

Fragmentation of ions at specific bonds reflects the ions

Physiochemical structure

96
New cards

trio that are the most common ion sources

ESI, APCI, MALDI

97
New cards

which ion source is most common for LC-MS

electrospray

98
New cards

for a less polar, thermaly stable analyte by LC-MS/MS use:

APCI

99
New cards

which MS/MS scan monitors precursor-product transitions for quantitation?

SRM/MRM

100
New cards

resolving large power is typical of which analyzer

obritrap