Electrophoresis
Migration of charged solutes or particles in a liquid medium under the influence of an electric field, separation technique used primarily for separating and identifying serum proteins in blood
Iontophoresis is the
migration of small ions
zone electrophoresis is the
most common clinical application, which is used to seperate macromolecules, such as proteins, in a porus support medium it requires: charged molecules migrates as zones, migration takes place across a support medium, sample is mixed with a buffer, separated protein zones are generated on the support media known as electropherogram, zones are quantified using a desitometer
The ability to separate proteins is sbed on the caharacteristics of
amino acids, positive ions (cations) will migrate towards a cathode, negative ions (anions) will migrate toward an anode
Ampholyte (Zwitterion)
molecule that can become either negatively or positively charged based on the pH of the solution in which it resides
isoelectric point
the pH at which a molecule has a net charge of 0
An ampholyte will take a positive charge in a solution more acidic than its isoelectric point (pI) and
migrates toward the cathode
An ampholyte will take a negative charge in a solution more alkaline than its isoelectric point (pI) and
migrates toward the anode
Proteins migrates through gel from cathode to anode (+) and are separated into five bands
albumin, alpha 1 globulins, alpha 2 globulins, beta globulin, gamma globulin
Migratory inhabitation factor phenomenon
at a pH at 8.6, the gamma globulins move toward the cathode, despite the fact that they are negatively charged
rate of migration depends on
net electrical charge of the molecule, size and shape of the molecule, electrical field strength, propeties of supporting medium, temperature of the operation
The isoelectric point of an amino acid or protein can be defined as
pH at which the amino acid or protein has no net charge
proteins are separated into how many fractions in electrophoresis
five
In electrophoresis, the greater the charge of the molecule
the faster it will migrate
Power supply of the Electrophoresis apparatus
best to use a supply with a constant current opposed to constant voltage which can cause joule heat: flow of current through a medium that offers resistance
Work flow of Electrophoresis apparatus
caused by the movement of buffer into the support medium, moisture and heat generation during electrophoresis causes movement of the buffer into the gel, using a lead can prevent some of the loss of evaporation, methods that use excessive heat need a cooling system
Buffer of Electrophoresis apparatus
carries the applied current, establishes the pH at which electrophoresis is performed, determines the electrical charge on the solute, Ionic strength: the sum of all charges present, negative or positive, increasing the ionic strength will increases conductance of the support, size of the ionic cloud surrounding a charged molecule, yeilds slower migration rates and shaper band separation
Support media of Electrophoresis apparatus
cellulose acetate, agarose gel, poluacrylamide gel, starch gel, automated systems
Cellulose acetate
dry, brittle film composed of about 80% air space, used in isoelectric focusing
Agarose gel
used for separation of proteins in serum, urine, CSF, hemoglobin variants, isoenzymes, lipoproteins
Polyacrylamide gel
thermostable, transparent, strong, relatively chemically inert and uncharged, proteins are separated on the basis of molecular sleiving: separation according to both charge-to-mass ratio and molecular size
The component that carries the applied current and establishes the pH at which electrophoresis is preformed
buffer
what is the importance of the ionic strength of the buffer used in gel electrophoresis
a heat labile protein will become denatured if the ionic strength of the buffer is too high
the best support medium to use because it is thermostable, strong, transparent, and eliminates endomosis due to the lack of charge
polyacrylamide
Staining detection
Serum proteins in general: amido black, coomassie brillant blue, ponceu S, Isoenzymes: formazan, nitrotrezolium blue, liportein zones: fat red, oil red O, sudan black B, DNA fragments: ethidium bromide, CSF porteins: silver nitrate
Endomosis
prefernetial movement of water in one direction through an electrophoresis medium, low in AGE and PAGE, occurs due to selective binding of one type of charge on the surface of the medium
Electrolysis
chemical decomposition produced by passing an electric current through a liquid or solution containing ions
common problems in electrophoresis
discontinuites, unequal migration of samples across the gel, distorted protein zones, ireegularities in sample application, hemolyzed specimens, plasma instead of serum, split albumin
unequal migration of samples down the gel in an electrophoresis assay in an interference usually caused by
dirty electrodes causing uneven applications of the electrical field
A split albumin zone is likely due to
a patient having more than one type of albumin (bis-albuminina)
Isoelectric focusing: type of electrophoresis
the protiens migrate across a gel until they reach a zone where the pH of the gel matches the isoelectric point (pI) of the protein, the technique separates proteins in a pH gradient unti pH = pI
Immunofixation (IFE): type of electrophoresis
detects specific types of immunoglobulins or gamma and kappa chains
CApillary: type of electrophoresis
same as basic but tube is connected to a detector at its terminal end and to a high-voltage power supply via buffer reservoirs
Two-dimensional electrophoresis
1st: charge dependent isoelectric focusing 2nd: molecular weight-dependent electrophoresis, proteomics: the identification and quantification of proteins and their posttranslational modifications in a given system
Osmosis
process the constitutes movement of solvent across a membrane in response to difference in osmotic pressure across the two sides of membranes
Osmotic pressure
pressure required to stop osmosis through a semipermeable membrane between a solution and pure solvent, governs the movement of solvent (water) across membranes that separate two solutions
Osmole
1 gram (g) molecular weight (MW) of a substance divided by the number of particles into which it dissociates
Osmolality
Expresses concentrations related to mass of solvent
osmolarity
expresses concentration per volume of solution
major osmotic substances in normal plasma
sodium, chloride, glucose, urea (BUN)
Osmolality equation
mOsm/kg = 2[Na] + glucose/18 + urea/2.8
osmometry
a technique that measures the concentration of dissolved solute particles in a solution that contribute to the osmotic pressure
A solute dissolved in solvent causes the following changes in colligative properties
increased osmotic pressure, lowered vapor pressure, increased boiling point, lowered freezing point
freezing point osmometers
measurement of the freezing point of solution by supercooling the specimen to -7C
vapor pressure osmometers
correlation studies have shown more variation with vapor pressure osmometers, so freezing point osmometers are more commonly used
Chromatography
a group of separation techniques that separate analytes by differential distribution between a stationary phase and a mobile phase
Stationary phase
a solid or a liquid phase that interacts with components of the mobile phase
mobile phase
a gas or a liquid that flows in a chromatographic system and carries the sample past the stationary phase
as mobile phase flows past the stationary phase, the solutes may
reside only on the stationary phase (no separation), reside only in the mobile phase (migration with the mobile phase), distribute between the two pahses (Differential migration)
basis of separation
those solutes with high affinity for the stationary phase move slower, those with low affinity for the stationary phase will migrate faster
Planar chromatography
stationary phase is on a thin support such as paper or solid surface, used in identification of drugs in urine, analysis of amino acids and lipids in amniotic fluid
migration of a solute is expressed by its retention factor (Rf) value, the ratio of solute migration to solvent front migration
Rf = distance from application point to solute center/distance from application point to mobile phase front
which phase of chromatography carries the sample
mobile phase
thin layer and paper chromatography are two types of
Planar chromatography
Column chromatography
stationary phase is packed into a tube or coated on the inner surface of the tube, two types: gas and liquid chromatography, both can be connected to a mass spectrometer to identify the separated solutes
Gas chromatography
a gas mobile phase, or a carrier gas, is used to carry a mixture of volatile solutes through a column containing the stationary phase ex: nitrogen, helium, argon, sample must be volatile, stationary phase usually is a nonvolatile liquid coated or bonded to particles or the inner surface of the capillary, the carrier gas carries analytes to a detector in order of their elution and identified qualitatively by their renention time or by mass spec
liquid chromatography
separation is based on distribution of solutes between a liquid mobile phase and a stationary phase, High performance LC: stationary phase in LC consists of small-diameter particles and needs high pressure to pump liquid through column
Detector
a device that responds to the presence of analyte in the mobile phase, the magnitude of which is used to identify and quantify analytes, which produces an electronic singal that is plotted as a function of: time, distance or volume the resulting display is a chromatogram
Chromatogram
a plot of detector response used in column chromatography to the presence of analyte in the mobile phase as a function of time or mobile phase volume
retention time
time interval between specimen injection and solute reaching the detector
Resolutions (Rs)
measure of chromatographic separation, requires that two peaks have different elution times for the peak centers and sufficiently narrow bandwidths to eliminate or minimize overlap, incomplete separation occurs when the calculated value is Rs < 0.8, baseline separation is obtained when Rs is > 1.25
which component of a gas chromatograph introduces an aliquot of sample to be analyzed into the column
injector
incomplete separation occurs when the calculate value for Rs
is less then 0.8
Ion exchange chromatography: Cation-exchange particles
contain negatively charged groups in the stationary phase that binds positively charged components (cations) in the mobile phase
Ion exchange chromatography: anion-exchange particles
contain positively charged groups in the stationary phase that binds negatively charged components (anions) in the mobile phase
Ion exchange separates
amino acids, glycated hemoglobin, hemoglobin variants, oligonucleotides
Size exclusion chromatography: variety of materials are used as a stationary phase for size exclusion
Dextran, poluacrylamide, agarose, polystyrene-divinlybenzene, porous beads
Affinity chromatography
one component of highly specific molecular interaction pairs is immobilized in a stationary phase and is used to capture molecules from the mobile phase, interactions pairs are strong, highly specific and may include: enzyme inhibitor, hormone-receptor, antigen-antibody, aptamer-ligand
Qualitative and Quantitative analysis of chromatography: external calibration
reference solutions containing known quantities of analytes are analyzed in a separate run to establish detector responses per amount of analytes
Qualitative and Quantitative analysis of chromatography: internal calibration
reference solution of known analyte concentrations are prepared, and a constant amount of a different compound, the internal standard is added to each reference solution and to each sample
The differential distribution of solutes between two immiscible liquids is the basis for separtion for
partition
In column chromatoprahy, unknown analytes are identified as they pass through the detector, comparision of the ______ of the unknown analytes with that of the internal standard is the method used for unknown identification
Retention time
Mass spectrometry
the instrument produces a mass spectrum: the relative, abundance of each ion is plotted as a function of its mass-charge-ratio
The mass spectrometer displays data in 2 ways:
Mass spectrum: signal intensity vs m/z or Chromatogram: signal intensity vs time
Mass spectrometery does not measure mass but a m/z ratio
m/z = (M plus or minus n)/n
Mass analysis
the process by which ions are identified according to m/z ratios
Resolution (m/delta m)
the mass of a compound divided by its peak width at half maximum height (FWHM)
mass accuracy
how accurate the instrument is by comparing measured mass assignment for an ion to the expected mass
Sensitivity
when concentration of an analyte can be detected
time of flight
flight time: time to traverse the drift tube, t w sq rt(m/z)
Faraday cup
Ion abundance is so high that it would saturate the output of an electron multiplier
Migration rate in electrophoresis depends largely on the ratio of
net charge to size
SPE fractions at pH 8.6, fasted to slowest
Albumin, a1, a2, B, y
Disc electrophoresis is best described as separation using a
Discontinuous buffer system (stakingand separting gel)
Two dimensional electrophoresis separates by a column then by
molecular weight
method that uses antisera to type monoclonal immunoglobulins?
Immunofixation electrophoresis
A Northern blot detects
fragment RNA
A split (two headed) albumin zone is seen in
Bis-albuminemia
Separation by relative solubility in two immiscible liquids is
partition chromatography
Hydrogen bonding and hydrophobic interactions on a solid surface drive
adsorption chromatography
Porus beads separate solutes by size in solutions, this is an example of
size exclusion (gel filteration)
Immobilized antibody capturing its antigens is an example of
Affinity chromatography
example of a change that improves column efficiency
using smaller stationary phase particles
what change alters selectivity rather than efficiency
changing mobile phase composition
Fragmentation of ions at specific bonds reflects the ions
Physiochemical structure
trio that are the most common ion sources
ESI, APCI, MALDI
which ion source is most common for LC-MS
electrospray
for a less polar, thermaly stable analyte by LC-MS/MS use:
APCI
which MS/MS scan monitors precursor-product transitions for quantitation?
SRM/MRM
resolving large power is typical of which analyzer
obritrap