Leak Test Video:

  • When a manometer reads 2,000 psi it DOES NOT ensure that the oxygen tank is OPEN

    • What happens is that someone can close out the oxygen tank & NOT FLUSH OUT the oxygen that is trapped in the system - which would make the reading continue to read HIGH even though the tank is SHUT

    • To verify DEPRESS the O2 FLUSH

      • When the O2 flush is press & the needle starts DROPPING that means PRESSURE was TRAPPED

      • When the O2 flush is press & the needle stays UP that means the oxygen tank is OPEN & ready to go

      • When done with anesthetic machine CLOSE the oxygen tank & RELEASE oxygen trapped in the system w/ O2 FLUSH

  • Common areas for leaks:

    • Reservoir bag - around the rim

    • Breathing circuit

    • Connectors/hoses of the machine

    • Screw valves

    • Flowmeter

    • Inlet & outlet of vaporizer

  • Steps to perform leak test:

    • Plug up any possible escape of O2

    • CLOSE POP-OFF

      • This closes the scavenge so there is NO leak of O2

      • FLUSH the system w/ oxygen (Press O2 FLUSH), to fill the reservoir bag

      • Look at manometer to ensure that its OVER 40 cm of water of pressure

  • OPEN FLOWMETER to about 0.2 L/min

    • Ensures there’s NO LEAK in the flowmeter system

  • Count to 10 seconds to ensure the needle of manometer DOES NOT drop & bag DOES NOT DEFLATE

  • When completed leak test CLOSE FLOWMETER & OPEN POP-OFF & DON’T CLOSE oxygen tank


2
New cards

Leak Test Steps - Rebreathing:

  1. CLOSE POP-OFF

  2. CLOSE patient end (mouth piece) w/ finger

  3. Press O2 FLUSH until a pressure reading of 40 cm H2O registers on the MANOMETER

    • Reservoir bag SHOULD INFLATE

    • Count to 20-40 seconds to see if the needle of manometer DOES NOT DECLINE

      • If the manometer DECLINES could indicate a LEAK BUT to ensure put FLOWMETER to 0.2L/min

        • If the manometer GOES UP again means NO LEAK & good to go

  4. When the FLOWMETER was used CLOSE it FIRST

  5. WITHOUT removing your finger from patient end, OPEN POP-OFF

    • Reservoir bag SHOULD DEFLATE

  6. REMOVE thumb from patient end

  7. Completely OPEN POP-OFF valve

  8. DO NOT TURN OFF OXYGEN TANK when completed leak test


3
New cards

Leak Test - Non-rebreathing:

  • DO NOT USE O2 FLUSH - USE FLOWMETER

  1. HOLD CLOSE POP-OFF & patient end at SAME TIME

  2. Use FLOWMETER & put it to 4L/min until a pressure reading of 40 cm H2O registers on the MANOMETER

    • After the 40 cm H2O have been reach in the manometer put the flowmeter to ZERO

    • Reservoir bag SHOULD BE INFLATED

    • Count to 20-40 seconds to see if the needle of manometer DOES NOT DECLINE

      • If the manometer DECLINES could indicate a LEAK BUT to ensure put FLOWMETER to 0.2L/min

        • If the manometer GOES UP again means NO LEAK & good to go

  3. When the FLOWMETER was used CLOSE it FIRST

  4. WITHOUT removing your finger from patient end, OPEN POP-OFF - by releasing finger off the pop-off

    • Reservoir bag SHOULD DEFLATE

  5. COMPLETED leak test

  6. DO NOT CLOSE OXYGEN TANK


4
New cards

Additional Leak Test Information:

  • Rebreathing circuit leak-test:

    • CLOSE POP-OFF

    • CLOSE patient end of circuit w/ THUMB or PALM

    • PRESS FLUSH until a pressure reading of 40 cm H2O (should stay at 40 for 20-40 seconds)

    • Reservoir bag should INFLATE

    • OPEN POP-OFF

  • Non-rebreathing leak-test:

    • CLOSE patient end while HOLDING POP-OFF CLOSE

    • Use FLOW METER to INFLATE the reservoir bag

    • Release the POP-OFF to OPEN


  • Between switching from the REBREATHING CIRCUIT to NON-REBREATHING CIRCUIT DO NOT TURN OFF the O2 TANK


  • ONLY DO a LEAK TEST everytime you CHANGE CIRCUITS


  • How to open oxygen tank:


5
New cards

Vaporizer:

  • When filling anesthetic gas FILL it to the LINE

    • Keep the vaporizer at least ONE-HALF FULL at ALL times

  • OVERFILLING will result in ANESTHETIC OVERDOSE

  • UNDERFILLING will lead to an INABILITY to keep patient anesthetized

  • To fill the vaporizer unscrew cap located on the SIDE of vaporizer

  • When anesthetic liquid is LEAKING - tighten pin in the inside w/ cap

  • Precision vaporizer are COLORED CODED:

    • Isoflurane = PURPLE

    • Sevoflurane = YELLOW


6
New cards

Scavenge:

  • We use a PASSIVE SCAVENGING SYSTEM

    • Rely on GRAVITY to remove the gases from the system

  • The canister SHOULD be WEIGHTED w/ a gram scale BEFORE its first use

    • AFTER each use, the canister SHOULD be REWEIGHED

  • Once the canister has GAINED 50 grams from the initial weight it SHOULD be DISPOSED - b/c it will NO LONGER ABSORB the WASTE GAS


7
New cards

CO2 Absorber:

  • To see if the CO2 absorber is exhausted check GRANULES

    • Soft & crumbles = FRESH granules

    • Hard & dry = OLD granules

      • May exhibit a color change from WHITE to PURPLE

  • Granules SHOULD be CHANGED REGULARLY - to prevent the patient from REBREATHING toxic levels of CO2

  • When filling the granules to the FILL LINE - place screw in hole so it DOESN’T get blocked w/ granules & causes it to not attach to the circuit valves

    • Don’t OVER TIGHTEN screw as it can BREAK


8
New cards

O2 Tank:

  • E-tank

    • If you hear leak (hissing sound) in tank - CLOSE IT

      • Could mean tank is not attached well to YOKE

    • To remove tank put your foot down below tank & unscrew it to remove it

    • Then LINE PINS from the TANK to YOKE to fix the leaking sound

  • H-tank

    • To swap tank use wrench & screw off b/c chained against wall