1/38
Looks like no tags are added yet.
transducer
any device that converts one form of energy into another
transducers perform 2 functions
transmission and reception
cable
provides an electrical connection between the crystal and the u/s system
housing (case)
protects the internal components of the transducer from damage, also insulates the patient from electrical shock
acoustic insulator
thin barrier of cork or rubber that prevents vibrations in the case from inducing an electrical voltage in the crystal of the transducer (blue and inside on the diagram)
electrical shield
prevents electrical signals in the air from causing noise on the images (orange and outer on diagram)
backing material
bonded to the back of the active element
reduce crystal ringing, improves axial resolution by decreasing PD and SPL
piezoelectric crystal
converts voltage into mechanical, acoustic waves and acoustic waves back to voltage
matching layer
minimized the acoustic impedance mismatch between crystal and tissue
lens
helps to focus the beam, not on every transducer
The impedance of a transducer active element is 1.8 MRayls, and the impedance of the skin is 1.3 MRayls. What is an acceptable impedance for the matching layer?
1.526 MRayls, in between 1.8 and 1.3
piezoelectric effect
the ability of crystals to generate an electrical voltage when mechanically stressed or deformed, when returning sound waves strike the crystal
reverse piezoelectric effect
the ability of crystals to deform and generate sound waves when an electrical voltage is applied causing the crystal to vibrate and emit an acoustic wave
poling (polarization)
the process of placing the manufactured crystal in an extremely high temperatire then exposing it to a strong electrical field
curie point/ temperature
temperature the crystal is polarized
operating frequency and propagation speed (related to crystal material) relationship
directly related
operating frequency and crystal thickness relationship
inversely related
operating frequency is _______ to the electrical drive signal in CW
equal
NZL (mm) =
D²(mm) fo (MHz/ 6
Huygens’ Principle
large active element acts as millions of tiny sound sources, each producing a V-shaped wavelet that contributes to the overall sound wave
fixed focusing
external with a lens in front of the crystal
internal through crystal curvature

phased array focusing
alternating electrical impulses
only possible with multi-element transducers (not on single crystal transducers)
arrays
multiple active elements, many separate elements connected to its own electronic circuitry (kerfs)
1D arrays
active elements arranged in a single row
creates traditional 2D images
1.5D Arrays
electronic elevation focusing
center elements alone for shallow focus or both center and outer can be used for deeper elevation focus
2D arrays
multiple elements in a square pattern
can produce 3D and 4D imaging
phasing
use of nanosecond time delays in electrical pulses used to steer and focus the sound beam
interpolation
a pre-processing image function, automatically estimates and fills in missing data or blank pixels between diverging scan lines, averages the amplitude of echoes of nearby scan-lines
array sequencing
turning on of groups of elements in succession
mechanical transducer
only 1 active element
sector shaped image
mechanical steering and focusing
damaged PZT = loss of image
mechanical annular array
concentric element rings
variable focus by turning on/off the rings
produces sector shaped images
damaged PZT = horizontal dropout
mechanically steered, electronically focused
linear sequential array
large, rectangular footprint
120-250 elements in line
uses sequencing to create scan lines
manual steering
PZT damage = dropout vertical
linear sequential to wide scan
produces a larger fov
sequencing for middle part
phasing for wing production
linear phased array
used for 2D and Doppler
rectangular image
200-300 elements
steered and focused electronically by phasing
fixed elevation focus and steer
curved linear arary
2D and Doppler
curved image by sequencing
electronic focusing by phasing
fixed elevation focus
mechanical steering
vector array
combo of linear sequential and linear phased
smaller footprint
trapezoidal pie shape
electronic steering and focusing
sector phased array
2D, M-mode, Doppler
sector shaped (pie), small footprint
steered electronically
variable transmit focus via phasing
damaged crystal = inconsistent/ erratic beam steering and focusing
3D imaging
created from 2D array
checkerboard array (2-9k elements)
info gathered from 3 planes
volume elements
aka voxels