1
New cards

transducer

any device that converts one form of energy into another

2
New cards

transducers perform 2 functions

transmission and reception

3
New cards

cable

provides an electrical connection between the crystal and the u/s system

4
New cards

housing (case)

protects the internal components of the transducer from damage, also insulates the patient from electrical shock

5
New cards

acoustic insulator

thin barrier of cork or rubber that prevents vibrations in the case from inducing an electrical voltage in the crystal of the transducer (blue and inside on the diagram)

6
New cards

electrical shield

prevents electrical signals in the air from causing noise on the images (orange and outer on diagram)

7
New cards

backing material

bonded to the back of the active element

reduce crystal ringing, improves axial resolution by decreasing PD and SPL

8
New cards

piezoelectric crystal

converts voltage into mechanical, acoustic waves and acoustic waves back to voltage

9
New cards

matching layer

minimized the acoustic impedance mismatch between crystal and tissue

10
New cards

lens

helps to focus the beam, not on every transducer

11
New cards

The impedance of a transducer active element is 1.8 MRayls, and the impedance of the skin is 1.3 MRayls. What is an acceptable impedance for the matching layer?

1.526 MRayls, in between 1.8 and 1.3

12
New cards

piezoelectric effect

the ability of crystals to generate an electrical voltage when mechanically stressed or deformed, when returning sound waves strike the crystal

13
New cards

reverse piezoelectric effect

the ability of crystals to deform and generate sound waves when an electrical voltage is applied causing the crystal to vibrate and emit an acoustic wave

14
New cards

poling (polarization)

the process of placing the manufactured crystal in an extremely high temperatire then exposing it to a strong electrical field

15
New cards

curie point/ temperature

temperature the crystal is polarized

16
New cards

operating frequency and propagation speed (related to crystal material) relationship

directly related

17
New cards

operating frequency and crystal thickness relationship

inversely related

18
New cards

operating frequency is _______ to the electrical drive signal in CW

equal

19
New cards

NZL (mm) =

D²(mm) fo (MHz/ 6

20
New cards

Huygens’ Principle

large active element acts as millions of tiny sound sources, each producing a V-shaped wavelet that contributes to the overall sound wave

21
New cards

fixed focusing

external with a lens in front of the crystal

internal through crystal curvature

22
New cards
<p>phased array focusing</p>

phased array focusing

alternating electrical impulses

only possible with multi-element transducers (not on single crystal transducers)

23
New cards

arrays

multiple active elements, many separate elements connected to its own electronic circuitry (kerfs)

24
New cards

1D arrays

active elements arranged in a single row

creates traditional 2D images

25
New cards

1.5D Arrays

electronic elevation focusing

center elements alone for shallow focus or both center and outer can be used for deeper elevation focus

26
New cards

2D arrays

multiple elements in a square pattern

can produce 3D and 4D imaging

27
New cards

phasing

use of nanosecond time delays in electrical pulses used to steer and focus the sound beam

28
New cards

interpolation

a pre-processing image function, automatically estimates and fills in missing data or blank pixels between diverging scan lines, averages the amplitude of echoes of nearby scan-lines

29
New cards

array sequencing

turning on of groups of elements in succession

30
New cards

mechanical transducer

only 1 active element

sector shaped image

mechanical steering and focusing

damaged PZT = loss of image

31
New cards

mechanical annular array

concentric element rings

variable focus by turning on/off the rings

produces sector shaped images

damaged PZT = horizontal dropout

mechanically steered, electronically focused

32
New cards

linear sequential array

large, rectangular footprint

120-250 elements in line

uses sequencing to create scan lines

manual steering

PZT damage = dropout vertical

33
New cards

linear sequential to wide scan

produces a larger fov

sequencing for middle part

phasing for wing production

34
New cards

linear phased array

used for 2D and Doppler

rectangular image

200-300 elements

steered and focused electronically by phasing

fixed elevation focus and steer

35
New cards

curved linear arary

2D and Doppler

curved image by sequencing

electronic focusing by phasing

fixed elevation focus

mechanical steering

36
New cards

vector array

combo of linear sequential and linear phased

smaller footprint

trapezoidal pie shape

electronic steering and focusing

37
New cards

sector phased array

2D, M-mode, Doppler

sector shaped (pie), small footprint

steered electronically

variable transmit focus via phasing

damaged crystal = inconsistent/ erratic beam steering and focusing

38
New cards

3D imaging

created from 2D array

checkerboard array (2-9k elements)

info gathered from 3 planes

39
New cards

volume elements

aka voxels