radiographic image quality
the exactness of representation of the patient’s
anatomy on the image.
FILM FACTORS
GEOMETRIC FACTORS
SUBJECT FACTORS
three major interrelated categories of radiographic quality:
radiographic quality
anatomic structure being examinged is imaged on the radiograph
resolution
ability to image two separate objects and visually distinguish one from the other.
spatial resolution
refers to the ability to image small object that have high subject contrast
cpnventional radiography
has excellent spatial resolution
contrast resolution
to distinguish anatomic structures of similar subject contrast
computed tomography
has excellent contrast resolution
detail or recorded detail
sometimes used instead of spatial and contrast resolution
refere to the degree of sharpness of structursl lines.
visibility of detail
The ability to clearly see and distinguish anatomical structures on a radiographic image.
noise
is the random fluctuation in the OD of the image.
film graininess
refers to the distribution in size and space of the silver halide grains in the emulsion.
inherent in the image receptor
structure mottle
refers to the phosphor of the radiographic intensifying screen
inherent in the image receptor
caused by physical structures or imperfections
quantum mottle
refers to the random in which x-rays interact with the image receptor
- is radiographic noise caused by too few X-ray
photons reaching the image receptor, resulting in a
grainy or mottled appearance.
film factors
are the factors that affect the quality
of a radiographic film image, particularly its density
(darkness) and contrast
sensitometry
the study of the relationship between the intensity of exposure of the film and the blackness after processing.
- the study of how film emulsion responds to
radiographic exposure.
characteristic curve
the primary use is to describe the contrast characteristics of the film throughout a wide exposure range
H&D curve hunter and driefield
sensitometric curve
D-log E curve
other names for characteristic curve
horizontal
log relative exposure
horizontal or vertical?
vertical
optical density
horizpntal or vertical?
toe, shoulder, straight line portion
three portions of the curve:
toe
The lower, gently rising portion. At low exposure, the
film produces only a small increase in density. Details
may be difficult to distinguish
straight line portion
The steep, nearly linear portion. A change in
exposure produces a relatively predictable change in
density. This is where film contrast is commonly
evaluated
shoulder
The upper portion where the curve flattens.
Increasing exposure produces progressively smaller
increases in density, approaching the film's
maximum useful density.
base plus fog
region to the left of toe
straight line portion
region between toe and shoulder
d max
ragion to right of shoulder
expose and process the film
measuring the densities produced
plotting a graph
how to obtain characteristic curve:
time scale sensitometry
kVp, mAs, focal spot
setting, ffd used are constant but exposure time will
vary
intensity scale sensitometry
using step wedge
densitometry
used to measure
the densities produced on each part of the film.
plotting a graph
a graph is plotted with recoded optical density
film speed / sensitivity
contrast
latitude
information provided by characteristic curve: (3)
below 0.2
optical densities of Refion left of toe (base plus fog)
base density and fog density
optical density on the film is due to inherent density its called ___ and ___.
base density
due to absorption of light transmitted through the film base
fog density
produced by the development of silver halide grain due to unintentional exposure.
storage fog
due to poor storage condition (high temp, high humidity fumes)
safe light fog
due to exposure to dark room
- safe light illumination
chemical fog
due to:
1. excessive development
2. high development temperature
3. chemical contamination
4. hyper active developer
0.2 and 2.5
optical densities of region between toe and shoulder:
subject contrast
It is the difference in x-ray intensities transmitted through different parts of patient.
it depends on:
patient thickness
tissue density
photon energy
effective atomic number
contrast media
scatter radiation
film contrast
Inherent in the film and influenced by processing
it depends on charqcteristic curve of film
film density
processing
film gamma
max slope of the characteristic curve
it tells us how much the film density changes when the film demsity changes when exposure changes
average gradient
is the average slope of the characteristic curve between two specified density
points, representing the overall contrast of the film.
film latitude
refers to the range of exposure that produces diagnostically useful densities on a radiographic film.
exposure error
is the difference between the exposure used and the exposure needed to produce
the desired radiographic image density.
over exposure and under exposure
two types of exposure: