OAE brief definition
We can physiologically measure the sounds the ear makes using a small microphone
generated by outer hair cells
low in amplitude, inaudible to us
How do the OHCs generate sound?
Change in length in response to stimulation
Push and pull on the basilar membrane → creates the sound
Purpose of OHC lengthening
amplify BM response for low level sounds; more signal reaches inner hair cells
How we can hear very soft and very loud sounds
By-product of OHC mobility
Classes of OAEs
Spontaneous OAEs: continuous tonal signals that occur without any stimulation
~50% of normal ears have these
If present, can have several SOAEs at multiple frequencies (not used clinically)
Evoked OAEs: produced by the ear in response to an acoustic stimulus. Used clinically - Two types:
Transient Evoked OAEs
Distortion Product OAEs
TEOAEs
Emission is evoked by a short, brief stimulus
Clicks (brief broad-band stimuli)
Tone pips (brief pure tones)
BROAD response from BM as a result
Ear’s response: broadband sound (a sound containing energy over a broad frequency range)
entire BM is stimulated with such a short sound
Annotated TEOAE chart
just try your best girl <3

DPOAEs
Stimuli is Primaries: two steady-state pure tones presented simultaneously
Primary frequencies: f1 and f2
Primary levels: L1 and L2
Ear’s response:
Several pure tones that are all mathematically related to the primaries
We measure the emission at the location corresponding to (2f1-f2)
Relationship between f1 and f
f2 > f1
Ratio is often 1.22 (1.22 x f1 = f2)
F2 selected matches up well with audiometric frequencies
Relationship between L1 and L2
L2 < L1
L1 = 65 dB SPL and L2 = 55 dB SPL
Measuring OAEs
Probe in ear
Hermetic seal
Probe: 1 or 2 loudspeakers and microphone
Quiet patient
1-5 minutes/ear, depending on the test and how much noise the patient makes
DP-Gram
plot of DPOAE amplitudes as a function of f2
each point reflects the “health” of the OHCs at the f2
To get a normal OAE, what conditions must be met?
Cochlea has to general the emission
Can’t have more than a mild SNHL (thresholds < 40 db HL)
Conductive mechanism must be normal
Stimulus must get to cochlea
Emission has to travel to microphone in ear canal
Auditory Evoked Potentials
Records electrical activity
Via electrodes on scalp or inside neural tissue
Cochlea converts energy from mechanical to electromechanical
electrochemical impulses carried by APs from auditory nerve to brainstem, then cortex
Electroencephalography (EEG)
Records electrical activity generated by brain function as evoked potentialsE
evoked potentials
when brain activity is evoked by a specific response
Evoked potential principles
EP is Small in amplitude relative to ongoing neural activity
Signal averaging: done to separate the neural activity in response to the evoking stimulus from general ongoing neural activity
Earliest response waves are generated by the cochlea (via ECoG)
Latest response waves are generated by the auditory cortex (ERPs)
Auditory Brainstem Response (ABR): response with the widest clinical application
Signal Averaging
Ongoing EEG activity is random
Positive and negative values
Will average out to 0
Evoked neural activity is positive OR negative
NOT RANDOM
Averaging ~2000 responses
So that we can separate evoked stimulus
Auditory Brainstem Response (ABR)
Measures the synchronous electrical impulses of the auditory pathway in response to an auditory stimulus
PHYSIOLOGICAL response, NOT HEARING
Waves
Early waves by auditory nerve
Late waves by brainstem
ABR response: first 10ms after stimulus
“Normal ABR”: implies normal cochlear, auditory nerve, and brainstem function
Measuring ABR
Between 3-5 electrodes
mastoids, forehead, vertex, nape of neck
Must avoid electrical noise
Electrically shielded room, insert earphones
Avoid muscle artifact
Patient should be relaxed and quiet
ABR test stimuli
clicks or tone bursts
Brief in duration to elicit a synchronous response from a fairly large neural population
Brief stimulus = broad spectrum
ABR presentation rate
33.1 clicks/sec for determining threshold
Threshold ABR used for babies or unreliable patients
11.1 clicks/sec for diagnostic ABR
ABR presentation level
Depends on purpose of test
Diagnostic ABR: tumor detection, type of hearing loss uses HIGH level
Threshold ABR: goes from HIGH to LOW until we don’t see a response
ABR response
series of waves labeled with roman numerals I-V
structure for ABR wave I
distal portion of CN VIII (vestibulocochlear)
structure for ABR wave II
proximal portion of CN VII (facial)
structure for ABR wave III
cochlear nucleus
structure for ABR wave IV
SOC
structure for ABR wave V
complex, but mostly from termination of lateral lemniscus in the inferior colliculus
ABR Analysis: Latency
where peaks occur on the x axis (msec)
2 measures are made and compared to age norms:
Absolute latencies
Interpeak latencies
Latencies INCREASE as the stimulus intensity DECREASES
Latencies INCREASE as the stimulus rate INCREASES
ABR analysis: amplitude
Measured in microvolts
From peak to the following trough
Amplitude DECREASES as stimulus intensity DECREASES
Extremely variable, not used diagnostically
Primarily for identifying wave V
Uses of ABR
Audiologic:
Assess threshold in difficult to test patient
Neurological purpose
Assess health of auditory nerve and brainstem
Done if suspecting a retrocochlear pathology
MRI vs ABR
Intraoperative monitory
Used during surgeries under which the auditory nerve (or cochlea) is at risk of being damaged