Velocity/pressure relationship
at any area of significant narrowing in the flow stream
(stenotic valve, VSD, regurgitant valve orifice…) – flow velocities increases in relation to the degree of narrowing.
PROXIMAL TO A STENOSIS
Pressures are Increased and Velocities are Decreased
WITHIN A STENOSIS
Pressures are Increased and Velocities are Decreased
DISTAL TO A STENOSIS
Pressures are Increased and Velocities are Decreased
Poiseuille’s Law
The greater the pressure gradient – the greater the volume of flow (velocity).

VTI – velocity time integral
(how much volume went through in a specific amount of time) – mean velocities.
VTI or TVI works better in patients with what?
Poor LV function and when moderate to severe AI is present than peak velocities.
A valve opens when the pressure in the delivering chamber exceeds the pressure in the chamber?
Receiving Chamber
A valve closes when the pressure in the receiving chamber exceeds the pressure in the chamber?
Delivering chamber
*When a normal valve is open, there is very little pressure difference on
either side of the valve (pressure gradient P1-P2).

PW Doppler: PV with PI
Normal systolic PV flow travels
away from the transducer so the waveform appears below the baseline.

Mild PI travels
Toward the transducer so the waveform appears above the baseline during systole.

Normal Diastolic TV flow travels
Toward the transducer so the waveform appears above the baseline
TR travels
Away from the transducer so the waveform appears below the baseline during systole


Is the Conventional 4C not parallel to the mitral flow?
Yes

Is the Nonconventional 4C parallel to mitral flow?
Yes

Normal diastolic MV flow travels
towards the transducer so it appears above the baseline.

MR travels
away from the transducer so the waveform appears below the baseline during systole.

Normal systolic LVOT flow travels
Away from the transducer so the waveform appears below the baseline

Normal systolic AOV flow travels
Away from the transducer so it appears below the baseline.

Abnormal AI travels
Towards the transducer so it appears above the baseline during Diastole.
Modified Bernoulli’s formula –
is used to measure pressure gradient between two chambers.
Pressure gradient (∆P)
= 4v²
∆P
the instantaneous pressure gradient (mmHg)
V
the instantaneous velocity (m/s)
*PW or CW is used to measure PR?
PW
CW or PW to measure TR?
CW
PW or CW to measure MV flow at the leaflet tips?
PW
CW or PW to measure MR and Stenosis at the valve level?
CW
PW or CW to measure LVOT
PW
CW or PW to measure AOV flow and AR at the valve level?
CW
RVOT and pulmonary artery:
PW Doppler ( looks similar to AoV)
PLAX RVOT or PSAX
RVIT (tricuspid valve):
PSAX, PLAX or apical 4ch (CW Doppler)
And we now can use our new AP3CH RVIT view to Doppler as well.
What is used to find RVSP?
The TR jet
RVSP =
4(V²) + RAP
max TR velocity + 3mmHg
Plethora
lack of collapsing of IVC with breathing
IVC<2cm and collapses with inspiration
+10mmHg
IVC>2cm and collapses
50% with inspiration +15mmHg
IVC>2cm and does not collapse w/insp.
+20mmHg
+3 mmHg: Normal IVC measuring
<2.1cm and collapses with “sniff” (>50%)
+8mmHg: Intermediate
unable to sniff or IVC collapses <20%
+15mmHg:
Dilated >2.1cm and doesn’t collapse

Normal systolic RVOT travels
Away from the transducer so it appears blue

Trivial PI travels
Toward the transducer so it appears red and turbulent during diastole
Normal systolic PV flow travels
Away from the transducer so the waveform appears below the baseline.
Mild PI travels
Toward the transducer so the waveform appears above the baseline during Diastole

Pulmonary Hypertension
Flying W - Decreased A
Pulmonic Stenosis
Increased A
Stroke volume
is the amount of blood ejected with each heart beat.
Stroke volume by Doppler =
CSA X VTI (CSA = πr²)
*This method can be used to measure stroke volume, cardiac output, valvular regurgitation and shunts.