What are the three main components of the entire x-ray imaging system?
The tube, the operating console, and the high voltage generator
What is the main function of the x-ray imaging system?
To provide a controlled flow of electrons to produce high-quality x-rays with minimal radiation exposure
What energy levels are regular diagnostic x-ray imaging systems identified by?
Approximately 25-150 kVp and 100-1200 mA
What are the three main circuits of the x-ray imaging system?
Primary circuit, secondary circuit, and filament circuit
What components are found in the primary circuit?
Main power switch, circuit breakers, autotransformer, timer circuit, and primary side of the step-up transformer
What components are found in the secondary circuit?
Secondary side of the step-up transformer, mA meter, rectifier bank, and x-ray tube (except filaments)
What components are found in the filament circuit?
Rheostat, step-down transformer, and the filaments
Where is the primary or low-voltage side located?
At the operating console
What is the purpose of the filament circuit?
To provide electrons for the x-ray tube current
What does the operating console control?
Voltage applied to the x-ray tube, current through the tube, and exposure time (kVp, mA, and exposure time)
What voltage does the x-ray imaging system operate on?
220 volts
What is the function of the line meter?
Measures how much voltage is coming into the circuit
What is the function of the line compensator/selector?
Adjusts voltage to ensure a precise 220 V is sent to the autotransformer
What does the autotransformer do?
Operates on self-induction with a single winding; supplies a precise voltage to the filament and high-voltage circuits
What is kVp adjustment responsible for?
Determines the quality or penetrability of the x-ray beam
How do the kVp major and minor selectors function?
kVp major selector adjusts in tens; kVp minor selector adjusts in ones
What does the kVp meter do?
Pre-reads the kilovoltage that will be applied
What determines the number of electrons crossing from cathode to anode?
The temperature of the filament, which controls mA (tube current)
What controls the filament temperature?
The filament current (amperage)
Where is the exposure timer circuit located?
On the primary side of the high-voltage transformer
What are the four types of exposure timers?
Synchronous, electronic, mAs timer, and automatic exposure control (AEC)
What is a synchronous timer?
Operates with a synchronous motor at 60 revolutions per second; shortest exposure time is 1/60 second
What is an electronic timer?
Most sophisticated, accurate, and complex timer used in most modern equipment
What is an mAs timer?
Monitors the product of mA and time and terminates exposure once the desired mAs is reached
What is an AEC timer?
Automatic exposure control; terminates exposure when the correct amount of radiation is received, typically set at 1.5 times the expected exposure
What is the difference between photodiode and ion chamber AEC detectors?
Photodiode detectors are behind the image receptor (older method); ion chambers are in front of the receptor (most common)
What are the three primary parts of the high-voltage generator?
High-voltage transformer (step-up), filament transformer (step-down), and rectifiers
What is the function of the step-up transformer?
Converts voltage to kilovoltage for x-ray production
What is the function of the step-down transformer?
Reduces current from amps to milliamps for filament heating
Why is AC current required in transformers?
To induce current through electromagnetic induction
What is rectification?
The process of converting alternating current (AC) to direct current (DC)
What device performs rectification?
Diodes
What is the function of diodes in the x-ray circuit?
Allow current to flow in one direction only, ensuring electron flow from cathode to anode
What happens with unrectified power?
Negative half of the AC cycle is wasted, resulting in poor image quality and inefficient x-ray production
What is the main purpose of rectification?
To provide positive power flow, reduce patient exposure time, and protect the x-ray tube by ensuring current flows cathode to anode only
How many pulses per second are produced in a single-phase full-wave rectified unit in the U.S.?
120 pulses per second
How many pulses per second are produced in a three-phase full-wave rectified unit?
360 pulses per second
What is voltage ripple?
The percentage fluctuation in voltage from zero to peak during x-ray production
What is the ripple for different power systems (Single-phase half or full wave)?
100% ripple
What is the ripple for different power systems (Three-phase, 6-pulse)?
14% ripple
What is the ripple for different power systems (Three-phase, 12-pulse) and ?
4% ripple
What is the ripple for different power systems (High-frequency)?
What is the advantage of high-frequency generators?
Less than 1% voltage ripple, more consistent output, reduced patient dose, and better image quality
What is the disadvantage of high-frequency generators?
They are more expensive
What is a capacitor discharge generator?
Uses stored charge from capacitors; voltage falls approximately 1 kV/mAs during exposure
Where are high frequency generators commonly used?
In mobile radiography
What is a falling-load generator?
A generator where mA drops during exposure to prevent overheating; used in high-capacity tubes like CT or IR
Which generator type provides the shortest exposure time for a given mAs?
Falling-load generator
How does kVp behave during capacitor discharge exposures?
It decreases throughout the exposure
How does kVp behave during falling-load exposures?
It remains constant throughout the exposure
What type of power supply gives the most consistent and high-quality x-ray beam?
Three-phase or high-frequency generators
What power rating formula is used for three-phase and high-frequency generators?
Power (kW) = (mA × kVp) / 1000
What power rating formula is used for single-phase generators?
Power (kW) = (0.7 × mA × kVp) / 1000
What determines focal spot size?
The mA selector
Where is the mA meter connected?
At the zero tap point on the high-voltage side (no voltage present)
What corrects for variations in power distribution to the x-ray machine?
The line voltage compensator
A diode allows electrons to flow in which direction?
From the cathode to the anode
A full-wave rectified three-phase unit provides how many x-ray pulses per second?
360 pulses per second
A single-phase waveform has what percent voltage ripple?
100%
What is the relationship between generator type and patient dose?
Better (higher phase) generators reduce patient dose and produce better image quality
Why is the step-up transformer named as such?
Because it has more windings on the secondary side than on the primary side
What are the key sections of the x-ray circuit?
Control console (low voltage), high voltage x-ray tube circuit, and filament circuit
What is the main voltage supplied to the x-ray imaging system from the power line?
220 V alternating current (AC)
What is the typical line frequency in the U.S. that determines pulse rate?
60 Hz