How does SAD differ from SSD for measuring attenuation?
Removes SSD dependence
Uses TAR, BSF, TPR, and TMR
PDD measures ____ measurements only and is ____ dependent
Central axis SSD
As a radiation beam passes through matter it undergoes a _____ reduction of the number of photons per area
Photons may be ___ or ____
Gradual
Absorbed or scattered
___ is the first isocentric non-SSD dependent measure of attenuation
TAR (tissue to air ratio)
TAR =
Absorbed dose in tissue at a depth / absorbed dose in air
TAR depends on ___, ___, and ___ but is independent of ____ because both methods are measured at the same distance from the source
energy, depth, field size
SSD (distance)
TAR is used for SAD treatments for energires < ___MV
4MV (aka Cobalt)
Back scatter factor = ___ at dmax
TAR
BSF =
also called peak scatter factor
dose in tissue at dmax / dose in air
Where is BSF measured for low energies vs megavoltage?
KV = at the surface
Megavoltage = at dmax
As E increases, the BSF will?
Decrease due to the scatter being more forward directed
As field size increases, the BSF will?
Increase because more scatter contributes to dose at central axis
Why do we use BSF?
Because it is affected by scatter from the patient, so we apply it whenever the physical field size on the patient has changed
MU's =
TD/ DR x TAR x BSF
When should we use BSF?
Accounts for scatter change at a depth of dmax so useful for orthovoltage when dose is prescribed at dmax
When shielding is added (change in physical field size)
If a cutout is added, how does the dose rate change?
dose rate x BSF (cutout)/BSF (cone)
Add in cutout BSF, take out cone BSF
How does PDD relate to TAR?
PDDd =
TARd x 1/BSF x ISL x 100%

SAR
the ratio of the scattered dose at a given point in a medium to the dose in air at the same point in space
Scatter air ratio is derived from the ____ by subtracting the ___ component from the total dose
Tissue air ratio
primary beam

____ is a variation of SAR but is used for calculating the scattered dose in a medium for energies over 3MV
SMR
TAR, BSF, and SAR limitations
TAR not useful for MU/Time calcs unless dose rate is also given in air
Limited to beam energies < 4MV
Scatter generally increases the PDD
True
TAR decreases with an increase in depth
True
At shallow depths, TAR is greater for low energy beams because of the increased scatter
True
TAR can be measured or derived from the PDD data
True
The isocentric treatment technique ensures
Reproducibility of treatment setup
Measured surface dose is
Dose at dmax for the specific field