How does SAD differ from SSD for measuring attenuation?

Removes SSD dependence

Uses TAR, BSF, TPR, and TMR

2
New cards

PDD measures ____ measurements only and is ____ dependent

Central axis SSD

3
New cards

As a radiation beam passes through matter it undergoes a _____ reduction of the number of photons per area

Photons may be ___ or ____

Gradual

Absorbed or scattered

4
New cards

___ is the first isocentric non-SSD dependent measure of attenuation

TAR (tissue to air ratio)

5
New cards

TAR =

Absorbed dose in tissue at a depth / absorbed dose in air

6
New cards

TAR depends on ___, ___, and ___ but is independent of ____ because both methods are measured at the same distance from the source

energy, depth, field size

SSD (distance)

7
New cards

TAR is used for SAD treatments for energires < ___MV

4MV (aka Cobalt)

8
New cards

Back scatter factor = ___ at dmax

TAR

9
New cards

BSF =

also called peak scatter factor

dose in tissue at dmax / dose in air

10
New cards

Where is BSF measured for low energies vs megavoltage?

KV = at the surface

Megavoltage = at dmax

11
New cards

As E increases, the BSF will?

Decrease due to the scatter being more forward directed

12
New cards

As field size increases, the BSF will?

Increase because more scatter contributes to dose at central axis

13
New cards

Why do we use BSF?

Because it is affected by scatter from the patient, so we apply it whenever the physical field size on the patient has changed

14
New cards

MU's =

TD/ DR x TAR x BSF

15
New cards

When should we use BSF?

Accounts for scatter change at a depth of dmax so useful for orthovoltage when dose is prescribed at dmax

When shielding is added (change in physical field size)

16
New cards

If a cutout is added, how does the dose rate change?

dose rate x BSF (cutout)/BSF (cone)

Add in cutout BSF, take out cone BSF

17
New cards

How does PDD relate to TAR?

PDDd =

TARd x 1/BSF x ISL x 100%

<p>TARd x 1/BSF x ISL x 100%</p>
18
New cards

SAR

the ratio of the scattered dose at a given point in a medium to the dose in air at the same point in space

19
New cards

Scatter air ratio is derived from the ____ by subtracting the ___ component from the total dose

Tissue air ratio

primary beam

<p>Tissue air ratio</p><p>primary beam</p>
20
New cards

____ is a variation of SAR but is used for calculating the scattered dose in a medium for energies over 3MV

SMR

21
New cards

TAR, BSF, and SAR limitations

TAR not useful for MU/Time calcs unless dose rate is also given in air

Limited to beam energies < 4MV

22
New cards

Scatter generally increases the PDD

True

23
New cards

TAR decreases with an increase in depth

True

24
New cards

At shallow depths, TAR is greater for low energy beams because of the increased scatter

True

25
New cards

TAR can be measured or derived from the PDD data

True

26
New cards

The isocentric treatment technique ensures

Reproducibility of treatment setup

27
New cards

Measured surface dose is

Dose at dmax for the specific field