1/74
All information taken from LO1 LS3 PP / Slide Deck
sterilization
kills all microorganisms and must be able to test for quality assurance
disinfection
kills some but not all microorganisms and cannot be tested for quality assurance
categories of re-usable client care items
critical items, semi-critical items, and non-critical items
critical items penetrate
soft tissue or bone
critical items must be
sterilized
what are examples of critical items
extraction forceps, explorer, and probe
semi-critical items touch
mucous membranes or non-intact skin
semi-critical items must be
sterilized
what is an example of a semi-critical item
mouth mirror
non-critical items touch
intact skin only
what do you do to non-critical items if contaminated during dental appointment
clean and disinfect
what is an example of a non-critical item
blood pressure cuff
for processing of critical and semi-critical items, what procedures do you follow for re-useable client care items
clean, package, sterilize, and store
why do we clean re-useable client care items prior to sterilization
to remove bioburden from surface of item
what are the 3 methods of cleaning prior to sterilization
automated instrument washers, ultrasonic cleaning, and hand scrubbing
what are the 3 steps of cleaning prior to sterilization
wash, rinse, and dry
after the 3 steps of cleaning, what must you do
inspect instruments to ensure bioburden has been removed
what is an automated instrument washer
a machine that washes, rinses, and dries instruments
automated instrument washer is the
safest and most efficient method
why is the automated instrument washer the safest and most efficient method
because you do not need to handle sharp contaminated instruments
what is a brand of an automated instrument washer
Hydrim
what is an ultrasonic cleaner
a machine that removes bioburden from surface by using ultrasound to agitate the liquid
ultrasonic cleaner is
less safe and not as efficient
what do you do after removing an instrument from the ultrasonic cleaner
must rinse and dry
how do you hand scrub
use long handle brush under running water and scrub one instrument at a time
hand scrubbing is
least safe and inefficient
why is hand scrubbing the least safe and inefficient
because DA must handle sharp contaminated instruments
what do you must do after hand scrubbing an instrument
must rinse and dry and be careful
guidelines for pre-cleaning instruments
instrument must be cleaned before
they are sterilized
removes bioburden to allow
sterilization to occur
wear utility gloves, safety glasses and
mask
instruments are still contaminated
after they have been cleaned!!
must visually inspect to ensure
instruments are clean
be careful
why do we use enzymatic spray or holding solution
because dried blood and debris is very difficult to clean off
instruments
when do we use enzymatic spray or holding solution
to prevent bioburden from drying on instruments if cannot clean instruments immediately
what is an instrument cassette
metal tray with a lid
why do we use instrument cassettes
minimize handling of sharp instruments and keep instruments organized
instruments are packaged before
sterilization
why do we package instruments for sterilization
it allows instruments to remain sterile for storage
what is key to remember about packaging instruments for sterilization
do not overfill the package
what are the different kinds of sterilization packaging
disposable pouches, disposable wraps, and reusable wraps
in sterilization packaging, each product requires
specific sterilization parameters of time, temp, and pressure
what machine uses stream under pressure
autoclave
what are the 3 parameters of an autoclave (steam under pressure machine)
temperature, pressure, and time
what are the three different cycles of an autoclave
gravity displacement, pre-vacuum, and steam flush pressure pulse (SFPP)
what machine is used for flash sterilization
statim
statim is used for what kind of sterilization
rapid sterilization
when you use rapid sterilization, that means
instruments are not packaged and therefore cannot be stored
after putting instruments through rapid sterilization
instruments must be used immediately
when should you use rapid/flash sterilization
for emergency situations only
types of sterilization
steam under pressure, flash/rapid sterilization, dry heat sterilization, chemical vapor sterilization, and ethylene oxide sterilization
what type of sterilization is not acceptable in SK
chemical vapor sterilization
what kind of sterilization monitors are there for quality assurance
physical/mechanical monitoring, chemical monitoring, and biological indicators
protocols and procedures used to verify (test) physical/mechanical monitoring
read gauges or displays of each cycle to verify temp, time, and pressure parameters are met
physical/mechanical monitoring machine may display
an “error” messages
what does a correct and incorrect reading mean on a physical/mechanical monitor
correct readings do not guarantee sterilization but incorrect readings likely mean not sterile
what does chemical monitoring use
chemical process indicators
what happens to chemical process indicators
change color when exposed to sterilization process
chemical process indicators have how many classes
5
what is a class 1 process indicator (chemical monitor)
external indicator
where do you use a class 1 process indicator
must be used on outside of every package
what happens to a class 1 process indicator
changes color when package has been through sterilization cycle
what does a class 1 process indicator verify
that the package has been through the sterilization cycle
and prevent confusion of packages which have and have not been processed
what is a class 2 chemical process indicator called
bowie-dick indicator
what is a class 2 bowie-dick indicator used for
pre-vacuum steam sterilizers to verify air is being removed during cycle and this test must be performed every day before the first load is processed
what does a class 3 temperature specific indicator do
change color to verify minimum temperature is reached within the sterilization chamber
what chemical process indicator is not used in SK Polytech Dental Clinic
class 3 temperature specific indicator
class 4 multi-parameter indicators must be used
inside every package
what does a class 4 multi-parameter indicators do
change color to verify 2 or more sterilization parameters have entered the package
what does the multi-parameter mean in a class 4 indicator
it monitors 2 or more parameters
what does a class 5 integrating indicator do
change color to verify all 3 sterilization parameters have been met
where do you place a class 5 integrating indicator
inside a process challenge device
when do you use a class 5 integrating indicator
must be used in each load processed through the sterilizer
what is a PCD
process challenge device
what is a process challenge device (PCD)
a sterilization pouch or wrapped cassette that is used to verify the performance of a sterilizer
what does a process challenge device duplicate
the conditions of an instrument being processed
class 5 integrating indicators and biological indicators are placed
inside the PCD then processed
what is the highest standard test
biological indicators (BI)
what are biological indicators used for
to verify sterilizer is killing all microorganisms