Rationale behind DSCT

to create a CT system capable with speeds sufficient to capture data in cardiac CTAs that froze cardiac motion, (to reduce motion artifacts)

2
New cards

DSCT Flash Mode main functional element

uses both tubes plus detector arragnements to acquire 2 sets of data simulataneously without overlap, no overlap is ensured by increasing the table speed, thus maximum pitch = 3.2 (3rd gen max pitch = 1.5)

3
New cards

Applications of DSCT Flash mode

when acquisition speed is essential, eg. polytrauma CT for unstable pts, pediatric CT

4
New cards

DSCT XL Mode main functional element

to accomodate very large pts that require longer scan times to fully penetrate the anatomy, however this affects single breath hold scans, and contrast enhancement phases. if the scan time was kept to the norm the resulting image would be noisy. THUS, in this mode both tubes operate at the same technical factors with a reduced pitch that allows more data samples to be collected with increase pt dose

5
New cards

DSCT Dual Energy CT main functional element

one tube with a higher and the other with a lower kVp to analyze the attenuation variation of predetermined values to produce colour mapping and volume measurements. The Tin filter can also reduce pt dose and increase spectral differentiation. however the main disadvantage is the smaller SFOV of Tube B. thus the tech must ensure that the anatomy of interest is within the SFOV of Tube B

6
New cards

DSCT as 3rd Generation

to only use Tube A, this variation permits a gradual transition for departments to develop new DSCT protocols and give time for techs to learn the new equipment