picture archiving and communication system
What does PACS stand for?
PACS
what stores, retrieves, distributes, and manages medical images digitally
digital imaging and communications in medicine
what does DICOM stand for?
DICOM
What is the standard format that allows medical imaging equipment to store, transmit, and share images and related information
images can be accessed quickly from multiple locations; requires less physical storage space and reduces the risk of lost or damaged films
What are 2 advantages of PACS over physical image archives
amplification - compensation - compression - sampling - encoding
What is the 5-step process used to translate analog information into digital?
amplification
strengthens the received electrical signal
compensation
adjusts signal strength to account for attenuation (TGC)
compression
reduces the wide range of signal amplitudes
sampling
measures the analog signal at specific time intervals
encoding
assigns numerical values to the samples and converts them into binary digital information
dynamic range
ratio between the strongest and weakest signals a system can process
decibels (dB)
what is dynamic range measured in?
low dynamic range
fewer shades of gray, higher contrast, more black and white appearance, narrow
high dynamic range
more shades of gray, lower contrast, more gradual gray transitions, wide
echo signals have an extremely wide range of amplitudes; reduces this range so the system can display the signals within its limited grayscale capabilities
why is compression needed as echoes pass through the ultrasound system?
75 + 35 = 110 dB
A signal has a dynamic range of 75 dB after 35 dB of compression. What was its original dynamic range?
20log10 (1000/1) = 60 dB
Strongest and weakest voltage amplitudes are 1,000 V and 1 V. Calculate the dynamic range.
below - signals are rejected and displayed as black; above - signals are clipped and cannot be displayed any brighter than the max level
what happens to signals below the threshold and above the saturation level?
spatial compounding
images acquired from different angles are combined
reduces speckle and artifacts
main benefit of spatial compounding
frequency compounding
images acquired using different frequencies are combined
reduces speckle and improves image quality
main benefit of frequency compounding
temporal compounding
multiple frames acquired at different times are averaged
reduces noise
main benefit of temporal compounding
coded excitation
uses a longer coded pulse during transmission that is decoded after reception
transmitter and receiver
where is coded excitation performed?
improves signal-to-noise ratio; improves penetration while maintaining good axial rsolution through pulse compression
benefits of coded excitation?
panoramic imaging
combines multiple images as the transducer moves across a large area to create one extended image
fusion imaging
combines ultrasound with another imaging modality, such as CT or MRI, to display corresponding anatomy together
when ultrasound waves travel through tissue and become distorted due to nonlinear propagation
how are tissue harmonics generated?
2.5 × 2 = 5 MHz
What second-harmonic frequency is associated with a 2.5 MHz fundamental?
oscillate when exposed to ultrasound
how do microbubbles generate contrast harmonics?
micorbubbles
oscillations are nonlinear, producing harmonic frequencies that improve blood pool visualization
shell composition/thickness; gas compostion
characteristics of microbubbles that affect stability and longevity
strain elastography
measures tissue deformation when compressed; primarily qualitative; shows relative tissue stiffness; stiffer tissue deforms less
shear wave elastography
measures speed of shear waves traveling through tissue; quantatative; provides stiffness measurements; stiffer tissue produces faster shear waves
quality assurance
ensures us equipment produces accurate, consistent, high-quality images; helps identify equipment problems before they affect patient examinations
image quality testing using phantoms; equipment and transducer inspections
name 2 components of a QA program
transducer
Which system component is especially likely to degrade?
helps identify gradual changes in equipment performance and detect problems early
Why compare QA results over time?
parallel to the us beam (one behind another)
which pin arrangement tests axial resolution?
perpendicular to the beam (side by side)
which pin arrangement tests lateral resolution?
elevational resolution
what does a slice thickness phantom assess?
causes partial volume artifacts; a cyst may appear to contain internal echoes even though it is actually anechoic
How can excessive slice thickness affect a cyst target?
dead zone
area immediately beneath the transducer where structures cannot be clearly visualized due to the transducer’s initial pulse and ring-down
standoff pad
adds distance between the transducer and superficial structure, moving the structure beyond the dead zone
viscous loss
energy lost as blood flows because of its internal resistance to movement
frictional loss
energy lost from blood interacting with vessel walls
narrowest portion of the stenosis
at a stenosis, where is velocity greatest?
pressure decreases
at a stenosis, how does pressure change there under Bernoulli conditions?
increase the pressure gradient; decrease resistance
Name two ways to increase flow
doppler angle is 90 degrees so cosine is 0
Why can a patent vessel show little or no color when the beam is perpendicular to flow?
nyquist limit
maximum doppler shift that can be accurately measured before aliasing occurs
PRF/2
Nyquist = ?
6000/2 = 3000 Hz; yes, since 4000 exceeds 3000
At a PRF of 6,000 Hz, would a 4,000 Hz Doppler shift alias?
increase PRF/scale and shift the baseline
Give two adjustments that can reduce aliasing.
improves color sensitivity and accuracy by collecting more pulses per scan line
Name one benefit of increasing the color Doppler
decreases frame rate because more time is needed to acquire data
Name one tradeoff of increasing the color Doppler
autocorrelation
What processing method is used for color doppler?
fast fourier transform (FFT)
what processing method is used for spectral doppler?
wall filter remove low-frequency doppler signals caused by tissue motion and vessel wall movement
Slow venous flow disappears after the wall filter is raised. Why?
increase background noise; make the spectral window appear filled in; can make the waveform look thicker and less defined
How would excessive spectral gain affect the display?