1
New cards

picture archiving and communication system

What does PACS stand for?

2
New cards

PACS

what stores, retrieves, distributes, and manages medical images digitally

3
New cards

digital imaging and communications in medicine

what does DICOM stand for?

4
New cards

DICOM

What is the standard format that allows medical imaging equipment to store, transmit, and share images and related information

5
New cards

images can be accessed quickly from multiple locations; requires less physical storage space and reduces the risk of lost or damaged films

What are 2 advantages of PACS over physical image archives

6
New cards

amplification - compensation - compression - sampling - encoding

What is the 5-step process used to translate analog information into digital?

7
New cards

amplification

strengthens the received electrical signal

8
New cards

compensation

adjusts signal strength to account for attenuation (TGC)

9
New cards

compression

reduces the wide range of signal amplitudes

10
New cards

sampling

measures the analog signal at specific time intervals

11
New cards

encoding

assigns numerical values to the samples and converts them into binary digital information

12
New cards

dynamic range

ratio between the strongest and weakest signals a system can process

13
New cards

decibels (dB)

what is dynamic range measured in?

14
New cards

low dynamic range

fewer shades of gray, higher contrast, more black and white appearance, narrow

15
New cards

high dynamic range

more shades of gray, lower contrast, more gradual gray transitions, wide

16
New cards

echo signals have an extremely wide range of amplitudes; reduces this range so the system can display the signals within its limited grayscale capabilities

why is compression needed as echoes pass through the ultrasound system?

17
New cards

75 + 35 = 110 dB

A signal has a dynamic range of 75 dB after 35 dB of compression. What was its original dynamic range?

18
New cards

20log10 (1000/1) = 60 dB

Strongest and weakest voltage amplitudes are 1,000 V and 1 V. Calculate the dynamic range.

19
New cards

below - signals are rejected and displayed as black; above - signals are clipped and cannot be displayed any brighter than the max level

what happens to signals below the threshold and above the saturation level?

20
New cards

spatial compounding

images acquired from different angles are combined

21
New cards

reduces speckle and artifacts

main benefit of spatial compounding

22
New cards

frequency compounding

images acquired using different frequencies are combined

23
New cards

reduces speckle and improves image quality

main benefit of frequency compounding

24
New cards

temporal compounding

multiple frames acquired at different times are averaged

25
New cards

reduces noise

main benefit of temporal compounding

26
New cards

coded excitation

uses a longer coded pulse during transmission that is decoded after reception

27
New cards

transmitter and receiver

where is coded excitation performed?

28
New cards

improves signal-to-noise ratio; improves penetration while maintaining good axial rsolution through pulse compression

benefits of coded excitation?

29
New cards

panoramic imaging

combines multiple images as the transducer moves across a large area to create one extended image

30
New cards

fusion imaging

combines ultrasound with another imaging modality, such as CT or MRI, to display corresponding anatomy together

31
New cards

when ultrasound waves travel through tissue and become distorted due to nonlinear propagation

how are tissue harmonics generated?

32
New cards

2.5 × 2 = 5 MHz

What second-harmonic frequency is associated with a 2.5 MHz fundamental?

33
New cards

oscillate when exposed to ultrasound

how do microbubbles generate contrast harmonics?

34
New cards

micorbubbles

oscillations are nonlinear, producing harmonic frequencies that improve blood pool visualization

35
New cards

shell composition/thickness; gas compostion

characteristics of microbubbles that affect stability and longevity

36
New cards

strain elastography

measures tissue deformation when compressed; primarily qualitative; shows relative tissue stiffness; stiffer tissue deforms less

37
New cards

shear wave elastography

measures speed of shear waves traveling through tissue; quantatative; provides stiffness measurements; stiffer tissue produces faster shear waves

38
New cards

quality assurance

ensures us equipment produces accurate, consistent, high-quality images; helps identify equipment problems before they affect patient examinations

39
New cards

image quality testing using phantoms; equipment and transducer inspections

name 2 components of a QA program

40
New cards

transducer

Which system component is especially likely to degrade?

41
New cards

helps identify gradual changes in equipment performance and detect problems early

Why compare QA results over time?

42
New cards

parallel to the us beam (one behind another)

which pin arrangement tests axial resolution?

43
New cards

perpendicular to the beam (side by side)

which pin arrangement tests lateral resolution?

44
New cards

elevational resolution

what does a slice thickness phantom assess?

45
New cards

causes partial volume artifacts; a cyst may appear to contain internal echoes even though it is actually anechoic

How can excessive slice thickness affect a cyst target?

46
New cards

dead zone

area immediately beneath the transducer where structures cannot be clearly visualized due to the transducer’s initial pulse and ring-down

47
New cards

standoff pad

adds distance between the transducer and superficial structure, moving the structure beyond the dead zone

48
New cards

viscous loss

energy lost as blood flows because of its internal resistance to movement

49
New cards

frictional loss

energy lost from blood interacting with vessel walls

50
New cards

narrowest portion of the stenosis

at a stenosis, where is velocity greatest?

51
New cards

pressure decreases

at a stenosis, how does pressure change there under Bernoulli conditions?

52
New cards

increase the pressure gradient; decrease resistance

Name two ways to increase flow

53
New cards

doppler angle is 90 degrees so cosine is 0

Why can a patent vessel show little or no color when the beam is perpendicular to flow?

54
New cards

nyquist limit

maximum doppler shift that can be accurately measured before aliasing occurs

55
New cards

PRF/2

Nyquist = ?

56
New cards

6000/2 = 3000 Hz; yes, since 4000 exceeds 3000

At a PRF of 6,000 Hz, would a 4,000 Hz Doppler shift alias?

57
New cards

increase PRF/scale and shift the baseline

Give two adjustments that can reduce aliasing.

58
New cards

improves color sensitivity and accuracy by collecting more pulses per scan line

Name one benefit of increasing the color Doppler

59
New cards

decreases frame rate because more time is needed to acquire data

Name one tradeoff of increasing the color Doppler

60
New cards

autocorrelation

What processing method is used for color doppler?

61
New cards

fast fourier transform (FFT)

what processing method is used for spectral doppler?

62
New cards

wall filter remove low-frequency doppler signals caused by tissue motion and vessel wall movement

Slow venous flow disappears after the wall filter is raised. Why?

63
New cards

increase background noise; make the spectral window appear filled in; can make the waveform look thicker and less defined

How would excessive spectral gain affect the display?