Vocabulary flashcards covering all 47 questions from the RAD 161 examination on x-ray tube construction, physics, operational dynamics, and automatic exposure controls.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Heat
The thermodynamic form of energy into which the vast majority (greater than 99%) of the electron stream kinetic energy is converted upon striking the target, while less than 1% produces x-radiation.
Anode
The positively charged electrode within the x-ray tube that serves as the target for the accelerated electron stream.
Cathode End
The terminal of the x-ray tube that should be oriented toward the thicker anatomy of the patient to take advantage of higher beam intensity caused by the anode heel effect.
Ball Bearings, Anode Neck, and Rotor
The specific assemblies of the rotating anode induction motor that operate within the vacuum envelope of the x-ray tube.
Compress the Electron Stream
The primary function of the metal collars (focusing cup) surrounding the cathode filaments in modern diagnostic x-ray tubes.
Actual Focal Spot Larger than Effective Focal Spot
The geometric principle described by the line focus principle concerning target bevel angulation and projected focal spot size.
Electromagnetic Induction Motor
The type of electric motor utilized to rotate the anode target without physical mechanical connections passing through the glass envelope.
Potential Difference
The electrical charge difference established between the cathode and anode during an exposure that accelerates electrons across the tube gap.
Heating of the Anode
The leading physical cause of x-ray tube damage and failure in diagnostic imaging.
Changing to a Larger Filament
The electrical and structural modification that takes place at the cathode when a radiographer switches from a small to a large focal spot selection.
Tungsten
The primary metallic element used in the manufacturing of rotating anode targets, stationary anode targets, and x-ray tube filaments.
5∘ to 17∘ Target Slant
The standard angular range of target bevel slants engineered into diagnostic x-ray tubes.
Anode (Electron Acceleration Target)
The positive terminal structure toward which electrons are accelerated upon application of kilovoltage across the tube gap.
Cathode
The negative electrode assembly of the x-ray tube that houses the filament wires and the focusing cup.
Actual Focal Spot
The physical area on the anode target surface that is directly struck by the accelerated electron stream.
100mR per Hour at 1Meter
The maximum permissible leakage radiation rate escaping the protective tube housing when operated under maximal conditions.
Heat Units
The thermal loading measurement calculated using the formula mA×kVp×time×rectification constant.
Thermionic Emission
The liberation of electrons from the surface of a heated filament wire in an x-ray tube.
Heat Conductivity, High Atomic Number, and High Melting Point
The three essential physical properties required of an effective x-ray tube target material.
Short SID and Larger Film Sizes
The technical and geometric conditions under which the anode heel effect becomes most noticeable across an image.
Space Charge
The localized cloud of liberated electrons surrounding the filament immediately following thermionic emission.
Dual Focus Tubes
Diagnostic x-ray tubes engineered with two filament wires to accommodate both small and large focal spot selections.
Less Than 1%
The fraction of electron kinetic energy converted into diagnostic x-rays upon striking the anode target.
Filaments, Focusing Cup, and Wiring
The structural components that collectively make up the cathode assembly of an x-ray tube.
Thermionic Emission (Non-Anode Function)
The electron liberation process associated exclusively with the cathode filament that is not a functional role of the anode.
Low mA Stations, Avoiding Rotor Holding, and Warm Anode
The operational precautions utilized to extend tube life: using low mA stations when possible, not holding the rotor switch unnecessarily, and avoiding exposures on a cold anode.
Capacitor
An electronic circuit device designed to store electrical charge.
Dissipate Heat (Tube Oil Bath)
The primary purpose of immersing the x-ray tube insert in specialized dielectric oil within the protective housing.
Applied Kilovoltage
The electrical parameter that primarily governs the force and kinetic velocity of the electron stream propelled across the tube gap.
10,000rpm Anode Rotation
The rotational speed achieved by specialized high-speed rotating anodes in diagnostic x-ray tubes.
Amperage to the Filament
The electrical parameter that primarily controls filament temperature and the amount of space charge liberated at the cathode.
Cathode (Negative Terminal)
The negatively charged terminal and electrode system within the x-ray tube.
Thermionic Emission (Non-Anode Association)
The operational phenomenon associated with the cathode rather than the anode (unlike the line focus principle, heel effect, and focal track).
Rotating Anode Advantages (Heat Dissipation, Target Area, Reduced Roughening)
The three operational benefits of a rotating anode over a stationary anode: increased heat dissipation, expanded target area, and reduced focal track roughening.
Superiorly (Cathode Orientation for AP T-Spine)
The direction the cathode end must be oriented during an AP thoracic spine examination to take advantage of the anode heel effect.
Crookes Tube
The partially evacuated, cold-cathode glass tube Wilhelm Roentgen was operating when he discovered x-rays in 1895.
Smaller than the Actual Focal Spot
The size relationship of the effective focal spot projected toward the patient and image receptor relative to the actual focal spot on the target.
Behind the Grid and in Front of the IR
The exact positioning of the automatic exposure control (AEC) ionization chamber within the Bucky assembly relative to the grid and image receptor.
Maintain a Vacuum
The primary operational purpose of the Pyrex glass envelope housing the internal cathode and anode assemblies.
Ionization Chamber and Phototimer
The two primary categories of radiation detection systems utilized in automatic exposure control (AEC).
Accurate Patient Positioning and Detector Selection
The two critical technical responsibilities required of a radiographer when operating an automatic exposure control (AEC) system.
Stator
The stationary electromagnetic component of an induction motor situated outside the glass envelope that induces rotation in the internal rotor.
Consistent, Accurately Exposed Radiographs
The principal diagnostic benefit achieved by utilizing an automatic exposure control (AEC) system.
AEC Off-Centering Consequences (Improper Exposure, Rapid Capacitor Charging, Short Exposure Time)
The technical failures that occur when anatomical structures fail to cover the active AEC detector cell: the capacitor charges prematurely fast, exposure time terminates too quickly, and the radiograph is underexposed.
Automatically Terminate the Exposure at 600mAs
The federal safety mandate requiring that an automatic exposure control backup timer shut off the exposure to protect the patient and equipment.
Pneumonia and Ascites
Pathological conditions that increase anatomical density, classified as additive diseases that are difficult for x-rays to penetrate.
Emphysema and Atrophy
Pathological conditions that decrease tissue density or cause tissue loss, classified as destructive diseases that are easy for x-rays to penetrate.