most detection instruments use electrons from ionization events and electronics to
convert energy into readable information
all detectors can be affected by
source geometry, so detector and source will change amount of interaction; typically simplified into 4pi and 2pi geometries
three types of gas detectors
ionization chamber
proportional counter
geiger-mueller counter
gas detectors are called such because
signal is produced when charged particles ionize the gas in the detection chamber
if gas is neonā¦
positive member of each ion pair is a neon ion while negative member is a freed electron
if detector gas is airā¦
positive is a positively charged nitrogen molecule and negative member of the ion pair is a freed electron
in the absence of an electric field, ion pairs created in gas of the detectorā¦
recombine
if ionization occurs bw 2 oppositely charged electrodes and the electric field is sufficiently strongā¦
positive and negative members of the ion pairs will separate and be collected at the electrodes
anode
positively charged electrode
cathode
negatively charged electrode
parallel plate electrodes
only used for ion chambers
coaxial electrodes
more common due to geometry achieving much higher electric fields for the same applied voltage
average energy lost by a charged particle per ion pair
about 34 eV
a typical electron/beta particle produces ___ ion pairs per cm of travel
100 ion pairs
bc of gas detector size, beta particles transferā¦
small portion of energy to detector gas
a typical alpha particle produces ___ ion pairs per cm
40000 ion pairs
bc of gas detector size, itās possible for some alpha particles to
expend all energy in detector gas
photons might interact directly w detector gas but are more likely to
interact w detector wall (cathode)
when photon interacts w cathode, an electron isā¦
knocked off of detector wall; this traverses the gas inside the chamber and might have sufficient energy to cause further ionization
magnitude of detector signal depends on
number of ion pairs produced in gas by incident charged particle
largest to smallest signal produced (generally) is
alpha (largest)
beta
photons
modes of operation detector gas
current, pulse mode
ion chambers usually operate in which mode
current; bc pulses are too small to be easily distinguished from electronic noise
proportional counters and geiger meuller detectors usually operateā¦
in pulse mode
current mode of operation
system measures current from chamber rather than counting pulses; calibration relates the current to a quantity such as exposure rate
current (I) is measured in units of..
amperes (A)
I = Q/t
collected charge (Q) is measured in
coulombs (C)
Q = n e
the current from a radiation detector is very small, often
on the order of picoamperes (pA)
pulse mode of operation characteristics
voltage (V)
as long as detector gas is not ionized, potential difference bw anode and cathode isā¦
constant
during movement of electrons to positive ions to electrode, the potential difference..
decreases
after some time, voltage stops decreasing andā¦
returns to baseline voltage
decrease and recovery = pulse
maximum change in voltage is
pulse height (amplitude)
when a detector operates in pulse mode, we mightā¦
count pulses and/or measure amplitude
for pulse height to carry useful info abt energy deposited in detector gasā¦
must be directly related to collected charge
a proper choice of the time constant is
necessary for pulse heigh to be related to collected charge
pulse width indicates
the charge collection time
the potential difference (voltage) reflects the
instantaneous charge in the circuit rather than the total collected charge
ĻĀ« charge collection time: pulse height does NOT reflect
energy deposited in detector gas; in other words it carries no useful information
large resistance to current and large capacitance of electronics meansā¦
current is āheld upā, causes potential difference (volts) to build up to potential max voltage
(Ļ) Ā» charge collection time: max pulse height reflects
energy deposited in detector gas; output pulse from detector is negative (voltage decreases)
six region curve
plot pulse height as a function of high voltage applied to a gas detector
gas amplification
number of ion pairs created by voltage from each primary ion pair created
all gas filled detectors operate on same principle of
ionized e- creating an electric current, but voltage applied across gas chamber changes how they react
six sections of six region curve
recombination
ionization
proportional
limited proportionality
geiger-mueller
continuous discharge
recombination region
at lowest voltages, strength of electric field is sufficiently low that many primary ion pairs created simply recombine
is no electrons reach anode
there is no pulse
ionization region
electric field just strong enough to prevent recombination of ion pairs produced in gas
ionization region: resulting current is referred to as
saturation current
why are alpha pulses correspondingly larger
alpha produces considerably more ion pairs than a beta
ion chambers used to measure
relatively high exposure rates, like over 1 mR/hr
ion chambers read accurately no matter
energy which they are calibrated at, bc of flat energy response
ion chambers measure dose accurately when it has
chamber material w tissue equivalent to stimulate rad interaction w tissue
dose calibrators
ion chambers that convert charge collected to activity for a selected isotope
amount of charge collected is different for each isotope per amount of activity, soā¦
ion chambers have multiplication factors for each isotope they use