X-ray Tube
What Is Tube Housing?
Material: Lead Lined
Description: Protective Device Prevents leaking
X-ray Tube
What is Glass Envelope?
Material: Pyrex
Description: Located in Tube Housing, Window of Glass envelope which is a small area made in thinner so X-ray Photons can pass easily
X-ray Tube parts
what are the two Electrons?
Positive Anode
Negative Cathode
Cathode Negative Parts
A filament is a source of electron what is it made off?
small coils of tungsten wire about 1 to 2 cm in length
Cathode:
2 Types of Filament
Small filament or small focus directs electrons to a smaller area on the target Large filament or large focus directs electrons to a larger area on the target
Cathode:
What is Thermionic Emission?
occurs when the negatively charged electrons break away from the hot filament to form a space charge
Cathode:
What is Space Charge?
a cloud of electrons around the filament which becomes the source of electrons in xray production
Cathode:
What is Space Charge Effect?
phenomena where further thermionic emission is halted by the space charge itself
Cathode:
what is Focusing Cup?
encases the two filaments and focuses the space charge on a target area on the anode
What are the Two Types of Anode?
Rotating (Mostly), Stationary (Outdated)
Anode:
What is a Rotor?
Material: shaft-like structure made of copper
Description: rotation device that turns the stem and target area
Anode:
What is a Stem:
Material: Copper
Description: attaches rotating target area to rotor
Anode:
What is a Rotating Target?
circular, disk-like structure with a beveled edge o receives electrons as they move from cathode to anode
Anode:
Rotating Target What is the Material made off?
tungsten
Anode:
Angle of Rotating Target and its no. of Rotation?
8 to 20°/ 9 to 15°
3000 to 10,000 rpm
Anode:
Focal Spot?
area of focal track that is struck by electrons
Anode:
Focal Track?
runs the circumference of the anode disk
Actual Focal Spot?
area being struck by electrons
Effective Focal Spot?
area directly below the target projected towards the window
Stationary anode?
Usually in Dental X-ray Units made out of Copper and Tungsten Target
X-ray Productions:
On and Off Switch
to connect the electrical power to the x-ray generator
X-ray Production:
Exposure Timer
determine the length of time that x-rays will be produced
X-ray Production:
Mechanical Timer
similar to spring-set dial, shortest setting is 1/10 s
X-ray Production:
Synchronous Timer
motor driven, utilizes the 60-Hz AC, has intervals related to 1/60 s, shortest setting is 1/120 s
X-ray Production:
Electronic Timer
most common, accurate, and reliable, designed with complex circuitry, shortest setting is less than 120 s
Evaluation of Timers:
Spinning Top Test
1-ø, full-wave rectified x-ray generators
Evaluation of Timer: Spinning Top Test
1 Dot per
1/120 sec exposure
Evaluation of Timer: Spinning Top Test
2 Dot per
1/60 sec exposure
Evaluation of Timer: Spinning Top Test
4 Dots per
1/30 sec exposure
Evaluation of Timer: Spinning Top Test
12 Dots per
1/10 sec exposure
Evaluation of Timer:
Calibrated oscilloscope
3-ø x-ray generator
compared to the selected time to verify accuracy
Evaluation of Timer:
Pulse-counter device
1-ø and 3-ø x-ray generator
placed in the primary x-ray beam, where pulses are counted and evaluated
Milliampere selector
counts the amount of electrical current flowing through the cathode filament
Milliampere selector Ranges
25 to 2000 mA
low current
small focal spot
high current
large focal spot
mAs Formula:
milliampere × time (seconds)
Kilovoltage selection
determine the specific tap on the autotransformer
kV major
changes of ~10 kV per turn
kV minor
changes of 1 to 2 kV per turn
“prereading” volt meter
position of meter indicates the amount of PD or voltage that will move across the x-ray tube during exposure
X-ray Production
Rotor switch
initiates the rotation of the anode in preparation for the exposure o maximum anode rotation is accomplished in ~1 sec
X-Ray Production:
Exposure switch
activated when the maximum RPM of the anode is reached
results in the production of x-rays
X-ray Production:
Accessory Controls Include:
Potter-Bucky diaphragm selection
vertical or horizontal devices
line voltage adjustor
X-ray Production:
Automated Exposure Controls (AEC)
termination occurs when desired amount of exposure reaches the film
X-ray Production:
PhotoTimer
terminates exposure when a proper amount of light reaches the photomultiplier tube
Location: Behind Cassette
X-ray Production:
Ionization Chamber
terminates exposure once a preset charge is accumulated and detected
Location: B/W cassette Holder and Pt
X-ray Production:
Back-up timer selection
a manual timer switch (ex: 1 sec) that will terminate exposure if the AEC failed, thus protecting the patient and the x-ray tube from continuous exposure
X-ray production:
Machine Controls mAS while kVP is?
AEC preselected Fixed KVP
X-ray Production:
Overexposure or underexposure in AECs
caused by improper chamber selection and/or patient positioning errors
Proper Care of Radiographic Equipment:
Warm Up Procedure: Anode Target Area
preheated each morning before use
X-ray Warm Up:
Exposures are Made using?
Low mA (50-100 mA), large focal spot, low to moderate kVp (65 to 70 kV), long exposure time of 2 to 3 seconds
Switch precaution
prolonged rotation with the presence of a “hot” filament will shorten the life of the anode and the cathode filament
Characteristic of X-Radiation:
Image Forming X-rays also known as Remnant Beam
those that exit the patient and interact with the image receptor
Characteristic of X-ray Radiation:
Exit Beam
the x-rays that remain as the useful beam exits the patient
X-ray Film:
Biggest Size
35 × 43 cm. or 14 × 17 in.
X-ray Film:
Smallest Size
20 × 25 cm. or 8 × 10 in.
X-ray Film:
Thickness
150 to 300 μm
Film Composition:
Base Purposes
support that allows the film to maintain its shape, prevent bending and provide support for the emulsion
Skeletal Framework
Film Composition:
Base Characteristics:
1. lucency – must be nearly transparent to white light
2. dimensional stability – maintain size and shape to prevent distortion of image
3. superior physical stability – flexible and sturdy to prevent damage when handling and processing
blue tint what its purpose
decrease eye strain, decrease parallax effect, improve contrast, contributes to 0.15 OD
X-ray Film:
Base is made up of what?
polyester – a polymer matrix
• ethylene glycol
• dimethyl terephthalate
X-ray Film:
Adhesive Layer
substratum layer
coated on the base before adding the emulsion simply adheres the base to emulsion
provides maximum contact b/w the base and emulsion
X-ray Film:
Emulsion Layer
radiation sensitive and light sensitive layer of the film coated on both sides of the base (duplitized)
X-ray Film:
the spectral response of an emulsion is
absorption of visible light emitted by the phosphors
X-ray Film:
To increase Contrast Emulsion you need
small, uniform grain size
X-ray Film:
Emulsion: made up of
Silver Halided Crystal
X-ray Film:
Silver Halide Crystal is Suspended in?
Gelatin
X-ray Film:
Emulsion: Gelatin is made out of what
a colloid extracted from cattle hooves and hide, then rendered into a very pure, uncontaminated state
X-ray Film:
Gelatin during
processing so that chemicals can penetrate the silver halide crystals
X-ray Film:
Gelatin Rehardens when
the invisible image becomes visible
X-ray Film:
Silver Halided Crystals Size
1.0 to 1.5 μm
X-ray Films:
Silver Halide Crystals Functions
energy detectors
X-ray Film:
Sensitivity specks
imperfections in the crystal structure
X-ray Film:
Silver Halide Crystal is made out of?
silver bromide – ~98%
silver iodide – ~2%
X-ray Film:
SuperCoat Layer
topcoat or overcoat
a durable protective layer that is intended to prevent damage to the sensitive emulsion layer underneath it
Characteristic of X-ray Films:
Film Speed
degree to which the emulsion is sensitive to x-rays or to light
As Film Speed increases
Increases Sensitivity, Decreases Exposure
2 primary factors that affect the speed point
the number of silver halide crystals
the size of silver halide crystals
Film Speed: Fast Number
200, 300
Film Speed: Average Number
100
Film Speed: Slow Number
>50
IR Speed Formula
1 divided by exposure
Density Maintenance Formula
old speed/new speed = new mAs/old mAs
Film Latitude
margin of errors in technical factors that the film allows that will still produce an acceptable range of densities
As Film latitude Increases
Decrease Contrast
Narrow Film Latitude
High Contrast Emulsion, Short GRAY Scale, High Contrast, High wavelength, Low Kvp, Low Penetrability
Wide Film Latitude:
Low Contrast Emulsion, Long Gray Scale, high KVP, High Penetrability, Low Wavelength
Types of X-ray Film:
Screen Type Film
the most widely used radiographic film, general choice for diagnostic radiology
one or two intensifying screens
parallax effect
crossover, crosstalk or print-through, occurs when light forms an angle affecting a spot in the emulsion and travels through the other side striking a spot not exactly corresponding to the opposite layer, producing an effect similar to penumbra
Ways to Reduce Parallax Effect
1. tinting the film with blue dye
2. using a monochromatic film – blue light cannot crossover blue dye, orthochromatic films will always have parallax as green light can crossover the blue dye
3. adding an anti-crossover layer
4. using a thinner base
Advantages of Screen Type Film
↓ emulsion thickness = ↓ developing time
↓ patient dose
disadvantages of screen-type film
↑ parallax effect = ↓ decreased resolution
↑ quantum mottle
Types of X-ray Film:
Direct Exposure Film (Non Screen Film)
designed for use without the need of an intensifying screen
Direct Exposure Film is commonly used
low-risk or low-exposure examinations of thin areas of the body
intraoral dental radiography
Advantages of Direct Exposure Film
↓ parallax effect/crossover/blur
↓ quantum mottle = ↑ image quality
Disadvantages of Direct Exposure Film
↑ patient dose ↑ emulsion thickness = ↑ development time
Type of X-ray Film:
Single Emulsion Film
has emulsion coated on one side of the base only
no parallax effect
Type of X-ray Film: Single Emulsion Film:
Anti Halation backing
prevents halation of light as the screens fluoresce,
↑ sharpness of image o requires ↑ exposure
Types of film according to spectral sensitivity
Monochromatic film is Sensitive to what Color?
blue sensitive, sensitive to wavelength