What is the definition of an artifact in CT?
A distortion or error in an image unrelated to the subject being studied
How else is a CT artifact defined based on attenuation?
A discrepancy between reconstructed CT numbers and true attenuation coefficients
What is another definition of a CT artifact related to detectors?
Anything causing incorrect measurement of transmission readings by detectors
How else can an artifact be defined in CT images?
Anything appearing on the image that is not present on the object scanned
What are the three main categories of CT artifacts?
Physics-based, patient-based, and equipment-induced
What causes physics-based CT artifacts?
The physical processes associated with data acquisition
What causes patient-based CT artifacts?
Patient motion, positioning, or anatomy
What causes equipment-induced CT artifacts?
Scanner hardware or technical failures
What is beam hardening in CT?
An increase in the mean energy of the x-ray beam as it passes through the patient
Why does beam hardening occur?
Lower energy photons are absorbed, leaving higher energy photons
What is the typical appearance of beam hardening artifact?
Dark bands or streaks with the periphery appearing lighter than the center
What is the cupping artifact?
A beam hardening artifact where the periphery appears lighter than the center
What is partial volume averaging?
Mathematical averaging of voxel tissue attenuations when more than one tissue is in a voxel
What causes partial volume artifact?
A dense object lying at the edge of the FOV with too few views for accurate description
How can partial volume artifacts be reduced?
By using thinner slices
What is an edge gradient artifact?
Artifact caused by sampling attenuation profiles around complex structures with sharp density differences
What does an edge gradient artifact look like?
Streaks and shading that can be light or dark
How can edge gradient artifacts be reduced?
By using thinner slices
What is voluntary patient motion in CT?
Motion controlled by the patient such as swallowing or breathing
What is involuntary patient motion in CT?
Motion the patient cannot control such as peristalsis or cardiac motion
What is the appearance of patient motion artifacts?
Shading, ghosting (shadow), streaking, or blurring
What are metallic artifacts in CT?
Star-shaped streaks caused by metal blocking x-ray transmission
Why do metallic artifacts appear white on CT?
Because no transmission reaches the detector
How can metallic artifacts be reduced?
Remove external metal, use metal artifact reduction software, adjust scan range or gantry tilt, increase kVp
What is an out-of-field artifact?
Occurs when part of the patient lies outside the scan FOV
Why do out-of-field artifacts occur?
Anatomy outside the FOV contributes to attenuation but is ignored in reconstruction
What is the appearance of out-of-field artifacts?
White shading or streaks
How can out-of-field artifacts be reduced?
Ensure all anatomy is inside the FOV, properly center the patient, and use correct scan FOV
What is a ring artifact?
An artifact that appears as concentric rings on the image
What causes ring artifacts?
A bad detector
Which CT scanner generation commonly produced ring artifacts?
Third-generation scanners
How can ring artifacts be reduced?
Recalibration or replacement of the bad detector
What is tube arcing?
An undesired surge of electrical current within the x-ray tube
What causes tube arcing?
Vacuum compromise in the x-ray tube
What is the appearance of tube arcing artifact?
Streaks and noisy or non-diagnostic images
How is tube arcing artifact reduced?
Routine equipment maintenance
What is a cone beam artifact?
An artifact caused by the geometry of cone-shaped x-ray beams in multidetector CT
What is the appearance of cone beam artifact?
Irregular borders and shading around object edges
Why is cone beam artifact considered problematic?
It does not represent the true shape of the object
How can cone beam artifacts be reduced?
Using cone beam reconstruction algorithms and using only center detector rows
What is the main goal of CT artifact recognition?
To correctly identify image distortions and avoid misdiagnosis
Why is CT artifact reduction important?
To improve image accuracy and diagnostic quality
Which artifact type is caused by physical x-ray interactions?
Physics-based artifacts
Which artifact type is caused by patient movement?
Patient-based artifacts
Which artifact type is caused by scanner malfunction?
Equipment-induced artifacts
Which artifacts are reduced by thinner slice thickness?
Partial volume and edge gradient artifacts
Which artifact produces star-shaped streaks?
Metallic artifact
Which artifact produces concentric rings?
Ring artifact
Which artifact produces irregular object borders?
Cone beam artifact
Which artifact produces dark streaks due to high-density material?
Beam hardening artifact
Which artifact produces ghosting and blur?
Patient motion artifact
What is the best general method to prevent equipment-induced artifacts?
Routine quality control and maintenance
Why is proper patient centering critical in CT?
To prevent out-of-field artifacts and beam hardening
Why should metal be removed before CT scanning?
To prevent severe streak artifacts and image degradation