The Physics and Treatment Parameters of Therapeutic Ultrasound

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/41

encourage image

There's no tags or description

Looks like no tags are added yet.

42 Terms

1
New cards

characteristics of the ultrasound wave (4)

  • audible sound ranges b/w 16-20,000 Hz

  • sound energy > 20,000 = ultrasound

  • 1 Hz = 1 cycle/second

  • number of collisions undergoes in 1 sec defines the freq of a sound wave and is measured in Hz


2
New cards

characteristics of the ultrasound wave components (3)

  • frequency

  • mode

  • intensity


3
New cards

frequency

1 MHz or 3 MHz

  • 3 MHz used when target tissue is within 1-2cm from body surface (aka superficial)

  • 1 MHz used when target tissue is > 2cm beneath the skin (aka deep)


4
New cards

3 MHz may be used to treat

plantar fasciitis, achilles tendinitis, tennis elbow

5
New cards

1 MHz may be used to treat

soleus, piriformis, hip adductors, or pts with high % of body fat

6
New cards

duty cycle/mode can either me

continuous = thermal effect

pulsed - non-thermal effect

7
New cards

continuous duty cycle/mode (3)

  • thermal effect

  • sound energy is constant

  • u/s energy produced 100% of the time


8
New cards

continuous duty cycle is used

later in the healing process

9
New cards

pulsed duty cycle/mode (3)

  • non-thermal effect

  • energy is periodically interrupted

  • no u/s energy is produced during the off period


10
New cards

pulsed duty cycle is used

during acute inflammatory phase

11
New cards

treatment parameters: power

energy is produced by an u/s generator and is measured in watts (W)

12
New cards

power (2)

  • function of both pulse width and pulse frequency

  • represents the amount of energy being produced by the transducer


13
New cards

treatment parameters: intensity

rate at which energy is being delivered per unit area (W/cm2)

14
New cards

intensity (2)

  • strength of sound waves at a given location within tissue being treated

  • also referred to as SAI (spatial average intensity)


15
New cards

spatial average intensity (SAI) def

intensity of the u/s beam averaged over the area of the soundhead (central third)

16
New cards

most therapeutic u/s generators are limited to a max output of

3.0 W/cm2 (WHO guidelines)

17
New cards

intensities greater than 10 W/cm2 are used to

destroy tissues (surgically)

18
New cards

intensities less than 0.1 W/cm2 used for

diagnostic purposes

19
New cards

SAI: goal

to select the lowest intensity to achieve the desired therapeutic effect

20
New cards

effective radiating area (ERA)

actual cross sectional area of the beam as it exits the metal end plate

  • always smaller than the sound head because the crystal is smaller than the sound head


21
New cards

ERA facts (4)

  • ideally crystal and sound head should be close to the same size

  • measured 5mm away from sound head

  • most energy is concentrated at the center of the head

  • large ERAs produce a more collimated beam vs smaller ERAs


22
New cards

ERA is expressed in

square cm

23
New cards

large ERA

collimated beam

24
New cards

smaller ERA

more divergent beam

25
New cards

spatial average intensity (SAI) def

amount of ultrasound energy passing through the ultrasound head, expressed in watts per square centimeter (W/cm2)

26
New cards

SAI formula

SAI = total watts (W) / effective radiating area (cm2)

27
New cards

beam nonuniformity ratio (BNR) def

indicator of the variability of intensity within the beam

28
New cards

BNR ratio

ratio of the peak intensity : SAI

29
New cards

BNR cont (4)

  • BNR should fall between 2 and 5

  • the lower the BNR the more uniform the output

  • high intensity areas of the beam are called “hot spots”

  • pt may complain of severe pain or a dull ache


30
New cards

depth of penetration

the higher the freq, the less the sound will diverge or spread out

  • depth of penetration is inversely related to output freq


31
New cards

collimated def

beam of parallel rays/waves forming a column of energy

32
New cards

due to the depth of penetration heat produced by 1MHz

lasts longer than 3 MHz

33
New cards

in bone u/s travels in

longitudinal and transverse waves

34
New cards

in other biological tissues u/s travels in

ONLY longitudinal waves

35
New cards

acoustic energy can be (4)

  • transmitted

  • absorbed

  • reflected

  • refracted


36
New cards

acoustic impedance def

a material’s ability to transmit sound

37
New cards

reflection and refraction

occur when energy is transmitted b/w materials with different impedances

38
New cards

standing wave def

a single frequency wave caused by the collision of 2 waves of equal frequency and speed travelling in opposite directions

  • inc risk of tissue damage


39
New cards

refraction def

the bending of waves as a result of a change in the speed of a wave as it enters a medium of a different density

  • when energy leaves a dense layer and enters a less dense one the speed inc

  • tends to occur as it passes through joint spaces


40
New cards

factors influencing energy absorption (2)

  • tissue (all or part of the energy may be absorbed)

  • each time some of energy is absorbed/refracted/reflected less energy is available to transmit into deeper tissues


41
New cards

attenuation def

measure of the dec in sound energy as the sound wave travels

42
New cards

factors influencing energy absorption from most → least collagen

1) tendon

2) ligaments

3) cartilage

4) muscle

5) bone