What year were x-rays discovered
1895
What was the first kind of low-energy equipment made
Superficial
Look over slide 3 (but don’t memorize in detail)
What is beam quality
Energy of the beam which has a specific d-max
What is d-max
Depth of maximum equilibrium
Where 100% of the dose is
D-max values for various energies
Superficial = 0 cm
Orthovoltage = 0 cm
Cesium-137 = 0.1 cm
Cobalt-60 = 0.5 cm
6 MV = 1.5 cm
10 MV = 2.5 cm
15 MV = 3 cm
18 MV = 3.5 cm
What does an isodose chart show
Represents levels of absorbed dose at different depths
Composite of the actual % depth dose distribution as it enter phantom
What do isodose charts typically always have and why
Flattening filters to have more even doses
What is teletherapy
Treatment from a long distance (from source to patient)
What is external beam radiation therapy
Radiation is delivered to patient from external source or machine
Known as teletherapy
May use low or high energy beam
What is HVL
Filtration known as a half-value layer
These filters absorb damaging low energy x-rays that scatter
Various materials used as a HVL depending on energy produced
What energy does contact therapy operate at
40-50 kV, tube current of 2 mA
What is the SSD of contact therapy
Very short
How is the beam applied in contact therapy
Via applicators
What filter is used for contact therapy
0.5-1 mm aluminum
View contact therapy machine on slide 14
What energy do superficial treatments use
50 - 150 kV
What kind of fall off does superficial treatment units have
Dose to surface of skin with rapid fall off
What kind of HVL does a superficial treatment unit use?
1-6 mm aluminum filter
What are attached to superficial treatment units and what do they act as
Cones are attached to machine as the collimator
What is the SSD for superficial treatment units
15-20 cm
What guides some superficial treatment machines
Ultrasound
Review what the superficial unit looks like on slide 16 and 17
What energy does orthovoltage therapy use
Capable of 200 - 500 kV but usually operates at 200 - 300 kV with 10-20 mA
Orthovoltage therapy d-max
D-max is on skin with 90% of dose at 2 cm
(Deeper than superficial machines)
What filter / HVL does orthovoltage therapy use
1-4 mm copper filtration (Thoreaus 3 filter)
ACT up (Aluminum, Tin, Copper) → order of materials is from patient to source
SSD for orthovoltage therapy
50-70 cm SSD with cones or movable diaphragm
Review what the orthovoltage XSTRAHL looks like on slide 19
Disadvantages of low energy machines
Low penetration since low-energy photons cause high mass attenuation coefficients
High skin dose
Output is low causing longer treatment times
Large penumbra
Difficult to aim beans accurately for bilateral or orthogonal beams (even tho its usually single field)
What is supervoltage therapy usually classified as
A higher voltage machine
Energy range of supervoltage therapy
500 - 1000 kV/1MV
How does a Van de Graaff generator work
High negative potential is applied across the x-ray tube
When electrons strike target, it makes x-rays
How is energy determined in a Van de Graaff generator
Energy is determined by voltage on the dome
Energy range of a Van de Graaff generator
1-5 MV
Is a Van de Graaff generator isocentric or non-isocentric
non-isocentric
What is a betatron
Old machine that produced photon and electron beams
First electron beam
Not used anymore