rad physics unit 6 first notes

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/87

encourage image

There's no tags or description

Looks like no tags are added yet.

88 Terms

1
New cards

analog systems

  • represents changing values as continuously variable physical quantities

    • similar to the brems emission from the anode


2
New cards

images recorded on radiographic fild are considers

analog

3
New cards

ditigal systems

  • divide the image into cells


4
New cards

each cell of information is a

  • pixel (picture element) that corresponds to the intensity of the x-ray beam that strikes a particulat area


5
New cards

dicrete values

  • represents the amount of radiation hitting each portion of the IR

    • similar to characteristic emission from the anode

    • includes CT, MR, US as well as CR and DR


6
New cards

in digital radiography the x rays form a store electronic image that is displayed as a

matrix of intensities

7
New cards

matrix

a series of boxes that gives form to the image

8
New cards

numeric info contained in each pixel is used to determine

the shade of gray or brightness level

9
New cards

numerical value assigned to the pixel is

directly proportional to the x-ray exposure to that area of the image receptor

10
New cards

what is pixel size measured in

  • microns (um)

  • 1/1000 of a millimeter


11
New cards

attenuation coefficient

  • the ratio or percentage of original x ray beam intensity absorbed by a particular tissue area in the patient


12
New cards

voxels

  • volume elements

  • data acquired from 3 dimensional volumes of tissue within the patient

  • determines the attenuation coefficient


13
New cards

each pixel in a radiographic image represents what within the patient

a voxel within the patient

14
New cards

degree of beam attenuation of the voxel determines the?

numeric value of the pixel

15
New cards

CR and DR are both what?

  • both computed

  • both make digital images


16
New cards

CR

  • Computed radiography

  • cassette based imagining

  • tech uses a cassette to transfer the digital image to the processor


17
New cards

DR

  • Direct radiography

  • cassette-less system

  • image is transmitted directly to the processor without thecnologist intervention


18
New cards

Digital radiography

  • a form of radiography that uses x-ray sensitive plates to capture data during the patient examination, immediately transferring it to a computer system without the use of an intermediate cassette

    • synonyms: DR or flat panel systems or cassetteless


19
New cards

two general categories of DR systems

  • direct conversion

  • indirect conversion


20
New cards

what is direct conversion or capture

  • x rays are captured and converted straaight to electrical signal without any steps in between

  • direct use photoconductor

  • one type - TFT


21
New cards

what is indirect conversion or capture

  • converts x rays to light and light to electrical signal

  • all indirect detectors use scintillator and a photodiode


22
New cards

pixel =

picture element in the final image

23
New cards

voxel =

a volume element within the patient

24
New cards

a del (detector element) =

  • a hardware device for detecting radiation

  • it is part of the image receptro

  • typical size is about 100 microns (1/10th millimeter)


25
New cards

3 components of a DEL

  1. semiconductor detector surface

  2. microscopic storage capacitor

  3. TFT (thin film transistor)


26
New cards

semiconductor detector surface

  • made of semiconductor material

  • collects the x rays or light and converts them to electrons


27
New cards

microscopic storage capacitor

  • allows the DEL to store electrical charge until it is read


28
New cards

TFT (thin film transistor)

  • acts as a switch (sometimes called a switching gate) that releases electrons when detector is “read”


29
New cards

semicondunctor detectors are built in?

layers

  • top, middle, and bottom


30
New cards

top layer of semiconductor detector

  • electrode


31
New cards

middle layer of semiconductor detector

  • the semiconductor detector element


32
New cards

bottom layer of semiconductor detector

another electrode that the TFTs connected to storage capacitors

33
New cards

the sensing process of semiconductor detectors

  1. x rays absoption causes ionization within the detector material

  2. voltage to the electrodes pull electrons and positive holes to opposite electrodes

  3. lower electrode conducts the current to the capacitor for storage


34
New cards

semiconductor detectors are sensitive to

  • x rays (if made of amorphous selenium, a-Se)

    • directly converts ionization from x rays into electronic signal

  • light (if made of amophous silicon, a-Si:H)

    • converts light to electronic signal (indirect capture)


35
New cards

fill factor

  • the ratio of the sensing area of a DEL to the area of the DEL itself


36
New cards

detectors with high fill factors (large sensing areas) provide?

better spatial and contrast resolution

37
New cards

lower fill factors require an

  • increase in technique, increasing patient dose


38
New cards

capatictor

  • stores electronic charge


39
New cards

what is capacitor made of

  • 2 conductive plates separated by dielectric insulating materials


40
New cards

how does a capactior work

  • when the charges created in the detector surface are sent to the capacitor the charges build up on one plate of the capacitor

  • the electric field attraction betwen the electrodes hold the electrons in place until the TFT releases them


41
New cards

Thin film transistors

  • in each DEL the TFT acts a a switching gate to release the built - up electrical charge when the del is “read out”

    • TFTs act like a switch with no moving parts


42
New cards

what happens under normal circumstances in TFT

  • the electrons trapped in the source (the capacitor) do not cross to the drain


43
New cards

what happens to the TFT when a voltage is applied to the gate lines

  • it creates a potential different between the source and the drain and electrons flow out of the capactior


44
New cards

active matrix array (AMA)

  • the grid DELs are arranged in


45
New cards

each del is connected to the learby dels by?

gate lines and data lines

46
New cards

address driver

  • controls the gate lines

  • which controls the order in which the dels are read out


47
New cards

when a voltage is applied to a gate line the TFT gates do what

  • open in sequence and dump their charge into a data line

  • this surge of charge flows to an amplifier, then to the analog to digital converter


48
New cards

direct detectors type

  • one type: flat panel TFT

  • converts x rays directly to electronic signal


49
New cards

what charge is applied to the top electrode

  • a negative charge


50
New cards

what charge is applied to the bottom electrode

  • a positive charge


51
New cards

what is between the electrodes

  • a photoconductor made of amophous selenium


52
New cards

what happens to the x rays exiting the patient

  • the x rays exit the patient and are absorbed by the amorphous selenium a-Se photo conductor

  • x ray absorption (both compton and photoelectrict) ionize the amorphous selenium, creating ion pairs


53
New cards

what happens to the ion pairs made

  • the ion pairs separate

  • the positive charges mitrate to the top (negative) electrode

  • negative charges migrate to the bottom (positive) electrode


54
New cards

what happens to the charges

  • positive charges return to the circuit

  • electrons are stored in the capactior


55
New cards

TFT releases electrons from the capactior

  • the address driver applies negative charge to the gate lines in a sequential pattern related to their position within the detector matrix

  • forcing stored electrons to exit that capacitor through the data lines


56
New cards

ADC

  • the charge flows through the analog to digital convertor (ADC)

  • the adc measures and numbers the charge and converts that information to binary code

  • ADC transmits the image data to the computer (CPU)

  • the computer processes the DEL values, assigns a shade of gray to each pixel and creates the image


57
New cards

Direct capture TFT system entire process:

  1. before exposure, an electric field is applied through the bias electrode across the surface of the amorphous selenium layer

  2. once exposure begins, x rays are absorbed by the amorphous selenium, creating ions (positive and negative electric charges). the ions created are in proportion to the x ray exposure received

  3. under the influence of the electric field applied in step 1, the electrons migrate toward positive electrode and are stored in the capacitors

  4. when the address driver applies voltage to the gate lines, the TFT switches release the charge from the capacitor

  5. the charge is then amplified and sent to the analog to digital convertor (ADC) which measures and numbers the ccharge and sends the signal to the computer workstation for processing and display


58
New cards

all indirect capture DR systems use

  • scintillator

  • photodetectors

  • charge emiiter

    • thin film transistor (TFT array)

    • charge couple device (CCD)

    • complementary metal oxide semiconductor (CMOS)


59
New cards

what is a scintillator

  • a device made of crystal phosphors that exhibit the property of luminescence


60
New cards

luminescent materials

  • absorb energy from particles or photons and re-emit the absorbed energy as light

  • 2 types: fluorescence and phosphorescence


61
New cards

fluorescence

glows as the interaction is occuring

62
New cards

phosphorescence

can emit light for some time after the interaction occurs

63
New cards

scintillator layers covers?

the photodetector and charge emitter

64
New cards

phosphors

  • crystalline powders that convert x rays into visible light and form the active component of the scintillator

  • phosphor crystals come in 2 shapes: turbid and columnar


65
New cards

Turbid phosphor

  • phosphor layer with a random distribution of phosphor crystals within the active layer

  • chemical make-up: gadolinium oxysulfide

  • emits light isotropically

    • reduces spatial resolution

  • cheaper to manufacture


66
New cards

columnar (needle) phosphor

  • crystals are manufactured as needles or columns

  • chemical make up: cesium iodide

  • needle crystals direct light uniformly downward to the photodetector

    • results in less light divergence and better spatial resolution

  • more expensive to manufacture


67
New cards

indirect detector types

  • flat panel TFT

  • charge-coupled device (CCD) or complementary metal oxide semiconductor (CMOS)

  • both need a scintillator to convert x rays to light


68
New cards

indirect capture TFT systems structure

  • a scintiallator plate

  • TFT array with each pixel containing a

    • contains an amorphous silicon photodetec or photodiode


69
New cards

indirect capture TFT systems function

  • a scintillator converts x ray photons into light

  • light interacts with a photoconductive material (typically amorphous silicon) to convert the light photons into electrons

  • the electrons migrate to thin film transistors and produce an electric signal


70
New cards

what charge is applied to top electrode

positive charge


71
New cards

what charge is applied to the bottom electrode

negative charge

72
New cards

what is between the electrodes

the photodiode is made of amorphous silicon

73
New cards

indirect capture TFT system entire process

  1. before exposure, an electricfield is applied through the bias elecrode across the surface of the amorphous silicon layer in the TFT array

  2. once exposure begins, x rays are absorbed by scintillator crystals. the electrons in the scintillator crystals become excited and lose energy as light (fluoresce) when the electrons return to their normal orbits. the light released is in proportion to the x ray exposure received

  3. the light photons are absorbed by the amorphous silicon photodetector. in amorphous silicon, light photons have sufficient energy to ionize the atom, creating ion pairs. the ions created are in proportion to the light exposure, which, in turn, is proportional to x ray exposure received

  4. under the influence of the electric field applied in step 1, the electrons migrate toward positive electrode and are stored in the capacitors

  5. when the address driver applies voltage to the gate lines, the TFT switches release the charge from the capacitor

  6. the charge is then amplified and sent to the analog to digital convertor which measures and numbers the charge and sends the signal to the computer workstation for processing and display


74
New cards

indirect capture CCD system structure

  • a scintillator place

  • light focusing optics

  • charged coupled device (CCD) or complementary metal oxide semiconductor (CMOS)


75
New cards

CCDs and CMOSs differ from TFTs in what

  • their representation of pixels

    • the DEL in a TFT array represents a single pixel

    • each CCD/CMOS chip contains millions of diodes that collect the imput for a single pixel


76
New cards

both CCD and CMOS are types of ?

  • integrated circuits - IC

    • light a computer circuit board


77
New cards

what is the IC in both

  • a silicon-based semiconductor that can asorb light photons

    • absorption of the light ionizes the atoms in the semiconductor


78
New cards

each IC panel is made of…?

millions of tiny diodes

79
New cards

electrons produced by ionizations in the semiconductor are collected by the ?

closest diode

80
New cards

charge produced is proportional to?

the amount of light incident on the photodetector

81
New cards

CCD pros

  • has higher fill factor = more sensitive = lower patient dose

  • wider dynamic range = more sensitive to very low light levels = better @ low dose imaging

  • lower amount of noise

  • faster response time that TFTs


82
New cards

CCD cons

  • all peripheral components like timers and ADCs must be attached external to the sensor itself

  • this makes the chip larger

  • more expensive than CMOSs


83
New cards

CMOS uses?

  • integrated circuits, which allow many perhipheral circuits to be integrated inside of a single chip

  • this allows the voltage conversino, charge amplification and the analog-to-digital conversion to take place inside each pixel


84
New cards

CMOS pros

  • faster processing speed thanCCDs

  • uses less power

  • cheaper

  • responds to lower light levels than CCds = even lower pt. dose


85
New cards

CMOS cons

  • has lower fill factor = less sensitve = higher pt. dose

  • more noise in the image

  • separate amplifiers for each pixel leads to variations and non-unifomity across the image


86
New cards

how do scintillator type flat panel detectors obtain digital information?

  • x rays strike the scintillator crystal (Csl)

  • scintillator crystal releases light

  • light interacts with a photodetector (a-Si) releasing electrons

  • electrons migrate to capacitors where they are stored

  • address driver stimulates TFT switch

  • TFT switch opens allowing electrons to flow from DELs to the ADC


87
New cards

how do non-scintillator type flat panel detectors obtain digital information

  • x rays interacts with amorphous silicon photoconductor, releasing electrons

  • electrons migrate to capacitors where they are stored

  • address driver stimulates TFT switch

  • TFT switch opens allowing electrons to flow from DELs to the ADC


88
New cards

How do CCDs obtain digital information

  • x rays strike scintillator creating light

  • light bounces off mirror, through lens to the CCD

  • silicon layer in CCD converts light to electron

  • electonrs captured by CCD pixels and sent to ADC

  • ADC assigns numeric values based on the number of electrons caputred by that pixel