EXPOSURE FACTOR FORMULATION

IS A SYSTEMIC PROCESS OF ESTABLISHING THE REQUIRED RADIATION DOSE FOR A PARTICULAR RADIOGRAPHIC EXAMINATION. IT IS A PRODUCT OF FACTORS SUCH AS KVP, MAS, AND GRID RATIO

2
New cards
  • Receptor Exposure Standardization

  • Image Consistency


PURPOSE OF EXPOSURE FACTOR FORMULATION

3
New cards
  • Optimum kVp/ Variable mAs

  • Variable kVp/ Fixed mAs

  • Automated Exposure Control

  • Anatomically Programmed Radiography


TYPES OF EXPOSURE FACTOR FORMULATION

4
New cards

EXPOSURE TECHNIQUE CHART

ARE PRE-ESTABLISHED GUIDELINES USED BY THE RADIOGRAPHER TO SELECT STANDARDIZED MANUAL OR AEC ___________________ FOR EACH TYPE OF RADIOGRAPHIC EXAMINATION

5
New cards

TECHNIQUE CHARTS STANDARDIZE

THE SELECTION OF EXPOSURE FACTORS FOR THE TYPICAL PATIENT SO THAT THE QUALITY OF RADIOGRAPHIC IMAGES IS CONSISTENT

6
New cards

Exposure Technique Charts and Digital Imaging

Exposure Technique Charts are just as important, if not more to go so, when using the IRs. Underexposure or overexposure of a film a film -screen IR can result in a radiograph with decreased or increased density. Because image brightness is controlled by computer processing, the visual cues for overexposure or underexposure are missing. _______________________ are an effective tool in selecting the appropriate exposure techniques for a quality digital image

7
New cards
  • RADIOGRAPHER’S JUDGEMENT

  • PATIENT VARIABLITY

  • INDIVIDUAL TECHNIQUE CHART FOR EACH X-RAY TUBE

  • STANDADIZATION OF DEPARMENTAL PRACTICES

  • PROPER EQUIPMENT PERFORMANCE

  • ACCURATE MEASUREMENT OF PART THICKNESS

  • AVAILABILITY AND PROPER USE OF CALIPER

  • CONSIDERATION FOR DIGITAL IMAGING


Conditions for Effective Technique Chart

8
New cards

RADIOGRAPHER’S JUDGEMENT

TECHNIQUE CHARTS SERVES AS GUIDES AND DO NOT REPLACE THE RADIOGRAPHER’S CRITICAL THINKING

9
New cards

PATIENT VARIABILITY

TECHNIQUE CHARTS ARE DESIGNED FOR AVERAGE 0R TYPICAL PATIENTS

10
New cards

INDIVIDUAL TECHNIQUE CHART FOR EACH X-RAY TUBE

EACH ___________ MUST HAVE ITS OWN TECHNIQUE CHART, EVEN IF THEY SHARE ONE GENERATOR

11
New cards

STANDARDIZATION OF DEPARTMENTAL PRACTICES

THE RADIOLOGY DEPARTMENT MUST ESTABLISH______________ FOR IMAGE QUALITY, EXPOSURE FACTORS, AND ACCESSORY USAGE

12
New cards

PROPER EQUIPMENT PERFORMANCE

RADIOGRAPHIC EQUIPMENT MUST BE FUNCTIONING _____________________ CHARTS TO PRODUCE CONSISTENT RESULTS

13
New cards

ACCURATE MEASUREMENT OF PART THICKNESS

______________________ OF THE ANATOMICAL PART IS ESSENTIAL IN SELECTING THE APPROPRIATE KVP AND MAS

14
New cards

AVAILABILITY AND PROPER USE OF CALIPERS

___________ ARE USED TO MEASURE THE THICKNESS OF THE ANATOMICAL PART

15
New cards

CONSIDERATIONS FOR DIGITAL IMAGING

__________________ HAS A WIDER EXPOSURE LATITUDE, ALLOWING IT TO COMPENSATE FOR SOME EXPOSURE VARIATIONS

16
New cards

VARIABLE KVP/FIXED MAS

  • THE VARIABLE KVP/FIXED MAS TECHNIQUE CHART IS BASED ON THE CONCEPT THAT KVP CAN BE INCREASED AS THE ANATOMIC PART SIZE INCREASES

  • ACCURATE MEASUREMENT OF PART THICKNESS IS CRITICAL TO THE EFFECTIVE USE OF THIS TYPE OF TECHNIQUE CHART


17
New cards
  • 2

  • 1CM (0.4IN)


SPECIFICALLY, THE BASELINE KVP IS INCREASED BY ____ FOR EVERY _________ INCREASE IN PART THICKNESS, WHEREAS MAS IS MAINTAINED

18
New cards

Variable kVp/Fixed mAs Technique Chart

The ________________ chart adjusts the kVp for changes in part thickness while maintaining a fixed mAs

19
New cards
  • FIXED KVP/VARIABLE MAS TECHNIQUE CHART

  • OPTIMAL KVP VALUE

  • ADJUSTING THE MAS


THE ___________________ MAS TECHNIQUE CHART USES THE CONCEPT OF SELECTING AN __________________ THAT IS REQUIRED FOR THE RADIOGRAPHIC EXAMINATION AND _________________ FOR VARIATIONS IN PART THICKNESS

20
New cards

OPTIMAL KVP

____________ CAN BE DESCRIBED AS THE KVP VALUE THAT IS HIGH ENOUGH TO ENSURE PENETRATION OF THE PART BUT NOT TOO HIGH TO DIMINISH RADIOGRAPHIC CONTRAST

21
New cards

THE MAS SHOULD BE ADJUSTED BY A FACTOR OF 2

BECAUSE X-RAYS ARE ATTENUATED EXPONENTIALLY, A GENERAL GUIDELINE IS THAT FOR EVERY 4–5 CM (1.6–2 IN) CHANGE IN PART THICKNESS ___________________________

22
New cards

Fixed kVp/ Variable mAs Technique Charts

___________________ Identify optimal kVp and alter the mAs for variations in part thickness

23
New cards

Knee

Anatomical Part?

<p>Anatomical Part?</p>
24
New cards

Midpatella

Measuring point?

<p>Measuring point?</p>
25
New cards

Bucky

Tabletop/Bucky?

<p>Tabletop/Bucky?</p>
26
New cards

Small

Focal spot size?

<p>Focal spot size?</p>
27
New cards
  • 10-13

  • 14-17

  • 18-21


1-3 Cm?

28
New cards
  • kVp is changed

  • mAs remains Constant

  • Patient thickness affects kVp


Variable kVp/ Fixed mAs

<p>Variable kVp/ Fixed mAs</p>
29
New cards
  • Optimum kVp is sleceted and maintained

  • mAs is adjusted

  • Patient Thickness affects mAs


Fixed kVp/Variable mAs

<p>Fixed kVp/Variable mAs</p>
30
New cards

AUTOMATIC EXPOSURE CONTROL

IS A SYSTEM USED TO CONSISTENTLY CONTROL THE AMOUNT OF RADIATION REACHING THE IR BY TERMINATING THE LENGTH OF EXPOSURE

31
New cards
  • AUTOMATIC EXPOSURE DEVICES

  • PHOTOTIMING


ALSO ARE CALLED _____________________ OR _____________

32
New cards

Principle of Automatic Exposure Control Operation

Once a predetermined amount of radiation is transmitted through the patient, the xray exposure is terminated. This defines the exposure time and therefore the total of radiation exposure to the IR

33
New cards
  • PHOTOTIMERS

  • IONIZATION CHAMBERS


TWO TYPES OF AEC SYSTEMS HAVE BEEN USED

34
New cards

PHOTOTIMERS

  • FIRST GENERATION AEC

  • EXIT TYPE DEVICE

  • COINS THE TERM “PHOTOTIMING”

  • REFERS TO THE USE OF AEC DEVICES THAT USES PHOTOMULTIPLIERS TUBES OR PHOTODIODES


35
New cards

IONIZATION CHAMBERS

  • ENTRANCE TYPE DEVICE

  • IS A HOLLOW CELL THAT CONTAINS AIR AND IS CONNECTED TO THE TIMER CIRCUIT VIA AN ELECTRICAL WIRE


36
New cards

MAS READOUT AND ITS IMPORTANCE

USED DURING THE EXPOSURE IS CALCULATED BY THE SYSTEM AND DISPLAYED POSTEXPOSURE, PROVIDING VALUABLE FEEDBACK TO RADIOLOGIC TECHNOLOGISTS


37
New cards

NOT APPLICABLE

UNDERSTANDING MAS READOUT IS ESSENTIAL FOR LEARNING MANUAL TECHNIQUES, ESPECIALLY IN SCENARIOS WHERE AEC IS ________________

38
New cards

Automatic Exposure Control and mAs Readout

If the radiographic unit has an mAs display readout display, The radiographer should take note of the reading after an exposure is made. This information can be invisible

39
New cards

kVp and Automatic Exposure Control Response

THE SELECTION OF KVP IS CRITICAL; HIGHER KVP RESULTS IN GREATER PENETRATION, LEADING TO SHORTER EXPOSURE TIMES AND REDUCED PATIENT DOSE

40
New cards

mA and Automatic Exposure Control Response

MA VALUE SELECTION WILL AFFECT THE EXPOSURE TIME NEEDED BY THE AEC DEVICE: HIGHER MA IS REQUIRED TO LESSEN THE EXPOSURE TIME FOR A PARTICULAR EXAM

41
New cards

MINIMUM RESPONSE TIME

  • 1ms (0.001s)


THE ______________________ OF MODERN AEC SYSTEMS IS APPROXIMATELY_____ _______, WHICH IS IMPORTANT FOR RAPID IMAGING SCENARIOS

42
New cards

LONGER

MINIMUM RESPONSE TIME IS USUALLY __________ WITH AEC SYSTEMS WITH OTHER TYPES OF RADIOGRAPHIC TIMERS

43
New cards

BACK UP TIME

________________ REFERS TO THE MAXIMUM LENGTH OF TIME FOR WHICH THE X-RAY EXPOSURE WILL CONTINUE WHEN USING AN AEC SYSTEM

44
New cards
  • OVEREXPOSURE

  • THE X-RAY TUBE FROM OVERHEATING


BACKUP TIME IS A SAFETY FEATURE THAT LIMITS THE MAXIMUM DURATION OF EXPOSURE, PROTECTING BOTH THE PATIENT FROM ______________ AND __________________________________

45
New cards

Function of Back up time

Backuptime, The maximum exposure time allowed during an AEC examination serves as a safety mechnanism when AEC is not used properly or is not functioning properly

46
New cards

DENSITY ADJUSTMENTS AND CONTROLS

AEC DEVICES ARE EQUIPPED WITH DENSITY CONTROLS THAT ALLOW THE RADIOGRAPHER TO ADJUST THE AMOUNT OF PRESET RADIATION DETECTION VALUES

47
New cards
  • 25%

  • (+25% OR −25%)

  • (+50% OR −50%)


A COMMON INCREMENT IS _____, MEANING THAT THE PREDETERMINED EXPO-SURE LEVEL NEEDED TO TERMINATE THE TIMER CAN BE EITHER INCREASED OR DECREASED FROM NORMAL IN ONE INCREMENT___________ OR TWO INCREMENTS _______________

48
New cards

TRUE

DENSITY CONTROLS SHOULD ONLY BE UTILIZED WHEN THE IMAGE BRIGHTNESS IS INADEQUATE DESPITE PROPER POSITIONING, TO AVOID UNNECESSARY REPEAT EXPOSURES

49
New cards
  • Proper Detection

  • Poor Positioning


Practical applications and Limitation of AEC: Detector Selection and positioning

50
New cards

PROPER DETECTOR

_______________ SELECTION IS CRUCIAL; DETECTORS SHOULD BE SUPERIMPOSED BY THE ANATOMY OF INTEREST TO ENSURE ACCURATE EXPOSURE READINGS

51
New cards

POOR POSISTIONING

______________ CAN LEAD TO OVEREXPOSURE OR UNDEREXPOSURE, INCREASING THE LIKELIHOOD OF REPEAT IMAGING DUE TO INADEQUATE QUALITY

52
New cards

PATIENT CENTERING

knowt flashcard image
53
New cards

IMPACT OF PATHOLOGY

  • CERTAIN PATHOLOGIES CAN SIGNIFICANTLY AFFECT AEC PERFORMANCE; FOR EXAMPLE, FLUID ACCUMULATION (LIKE PNEUMONIA) MAY REQUIRE LONGER EXPOSURE TIMES, WHILE CONDITIONS LIKE EMPHYSEMA MAY NECESSITATE SHORTER EXPOSURES

  • THE AEC SYSTEM IS DESIGNED TO COMPENSATE FOR CHANGES IN PATIENT THICKNESS


54
New cards

WHEN TO AVOID AED

  • AEC SHOULD NOT BE USED WHEN THE ANATOMY DOES NOT ADEQUATELY COVER THE DETECTORS OR WHEN PATHOLOGICAL CONDITIONS COULD MISLEAD THE DETECTORS, POTENTIALLY RESULTING IN INCORRECT EXPOSURE SETTINGS.

  • RADIOLOGIC TECHNOLOGISTS MUST ASSESS EACH SITUATION TO DETERMINE THE APPROPRIATENESS OF AEC USAGE.

  • SOMETIMES, THE BEST SOLUTION IS A MANUAL TECHNIQUE DETERMINED THROUGH USE OF A TECHNIQUE CHART. AEC IS NOT A REPLACEMENT FOR A KNOWLEDGEABLE RADIOGRAPHER USING CRITICAL THINKING SKILLS


55
New cards

ANATOMICALLY PROGRAMMED TECHNIQUE

REFERS TO A SYSTEM THAT ALLOWS THE RADIOGRAPHER TO SELECT A PARTICULAR BUTTON ON THE CONTROL PANEL THAT REPRESENTS AN ANATOMIC AREA; A PREPROGRAMMED SET OF EXPOSURE FACTORS IS DISPLAYED AND SELECTED FOR USE

56
New cards

Anatomically Programmed Technique and Patient Exposure

knowt flashcard image