EXPOSURE FACTOR FORMULATION
IS A SYSTEMIC PROCESS OF ESTABLISHING THE REQUIRED RADIATION DOSE FOR A PARTICULAR RADIOGRAPHIC EXAMINATION. IT IS A PRODUCT OF FACTORS SUCH AS KVP, MAS, AND GRID RATIO
Receptor Exposure Standardization
Image Consistency
PURPOSE OF EXPOSURE FACTOR FORMULATION
Optimum kVp/ Variable mAs
Variable kVp/ Fixed mAs
Automated Exposure Control
Anatomically Programmed Radiography
TYPES OF EXPOSURE FACTOR FORMULATION
EXPOSURE TECHNIQUE CHART
ARE PRE-ESTABLISHED GUIDELINES USED BY THE RADIOGRAPHER TO SELECT STANDARDIZED MANUAL OR AEC ___________________ FOR EACH TYPE OF RADIOGRAPHIC EXAMINATION
TECHNIQUE CHARTS STANDARDIZE
THE SELECTION OF EXPOSURE FACTORS FOR THE TYPICAL PATIENT SO THAT THE QUALITY OF RADIOGRAPHIC IMAGES IS CONSISTENT
Exposure Technique Charts and Digital Imaging
Exposure Technique Charts are just as important, if not more to go so, when using the IRs. Underexposure or overexposure of a film a film -screen IR can result in a radiograph with decreased or increased density. Because image brightness is controlled by computer processing, the visual cues for overexposure or underexposure are missing. _______________________ are an effective tool in selecting the appropriate exposure techniques for a quality digital image
RADIOGRAPHER’S JUDGEMENT
PATIENT VARIABLITY
INDIVIDUAL TECHNIQUE CHART FOR EACH X-RAY TUBE
STANDADIZATION OF DEPARMENTAL PRACTICES
PROPER EQUIPMENT PERFORMANCE
ACCURATE MEASUREMENT OF PART THICKNESS
AVAILABILITY AND PROPER USE OF CALIPER
CONSIDERATION FOR DIGITAL IMAGING
Conditions for Effective Technique Chart
RADIOGRAPHER’S JUDGEMENT
TECHNIQUE CHARTS SERVES AS GUIDES AND DO NOT REPLACE THE RADIOGRAPHER’S CRITICAL THINKING
PATIENT VARIABILITY
TECHNIQUE CHARTS ARE DESIGNED FOR AVERAGE 0R TYPICAL PATIENTS
INDIVIDUAL TECHNIQUE CHART FOR EACH X-RAY TUBE
EACH ___________ MUST HAVE ITS OWN TECHNIQUE CHART, EVEN IF THEY SHARE ONE GENERATOR
STANDARDIZATION OF DEPARTMENTAL PRACTICES
THE RADIOLOGY DEPARTMENT MUST ESTABLISH______________ FOR IMAGE QUALITY, EXPOSURE FACTORS, AND ACCESSORY USAGE
PROPER EQUIPMENT PERFORMANCE
RADIOGRAPHIC EQUIPMENT MUST BE FUNCTIONING _____________________ CHARTS TO PRODUCE CONSISTENT RESULTS
ACCURATE MEASUREMENT OF PART THICKNESS
______________________ OF THE ANATOMICAL PART IS ESSENTIAL IN SELECTING THE APPROPRIATE KVP AND MAS
AVAILABILITY AND PROPER USE OF CALIPERS
___________ ARE USED TO MEASURE THE THICKNESS OF THE ANATOMICAL PART
CONSIDERATIONS FOR DIGITAL IMAGING
__________________ HAS A WIDER EXPOSURE LATITUDE, ALLOWING IT TO COMPENSATE FOR SOME EXPOSURE VARIATIONS
VARIABLE KVP/FIXED MAS
THE VARIABLE KVP/FIXED MAS TECHNIQUE CHART IS BASED ON THE CONCEPT THAT KVP CAN BE INCREASED AS THE ANATOMIC PART SIZE INCREASES
ACCURATE MEASUREMENT OF PART THICKNESS IS CRITICAL TO THE EFFECTIVE USE OF THIS TYPE OF TECHNIQUE CHART
2
1CM (0.4IN)
SPECIFICALLY, THE BASELINE KVP IS INCREASED BY ____ FOR EVERY _________ INCREASE IN PART THICKNESS, WHEREAS MAS IS MAINTAINED
Variable kVp/Fixed mAs Technique Chart
The ________________ chart adjusts the kVp for changes in part thickness while maintaining a fixed mAs
FIXED KVP/VARIABLE MAS TECHNIQUE CHART
OPTIMAL KVP VALUE
ADJUSTING THE MAS
THE ___________________ MAS TECHNIQUE CHART USES THE CONCEPT OF SELECTING AN __________________ THAT IS REQUIRED FOR THE RADIOGRAPHIC EXAMINATION AND _________________ FOR VARIATIONS IN PART THICKNESS
OPTIMAL KVP
____________ CAN BE DESCRIBED AS THE KVP VALUE THAT IS HIGH ENOUGH TO ENSURE PENETRATION OF THE PART BUT NOT TOO HIGH TO DIMINISH RADIOGRAPHIC CONTRAST
THE MAS SHOULD BE ADJUSTED BY A FACTOR OF 2
BECAUSE X-RAYS ARE ATTENUATED EXPONENTIALLY, A GENERAL GUIDELINE IS THAT FOR EVERY 4–5 CM (1.6–2 IN) CHANGE IN PART THICKNESS ___________________________
Fixed kVp/ Variable mAs Technique Charts
___________________ Identify optimal kVp and alter the mAs for variations in part thickness
Knee
Anatomical Part?

Midpatella
Measuring point?

Bucky
Tabletop/Bucky?

Small
Focal spot size?

10-13
14-17
18-21
1-3 Cm?
kVp is changed
mAs remains Constant
Patient thickness affects kVp
Variable kVp/ Fixed mAs

Optimum kVp is sleceted and maintained
mAs is adjusted
Patient Thickness affects mAs
Fixed kVp/Variable mAs

AUTOMATIC EXPOSURE CONTROL
IS A SYSTEM USED TO CONSISTENTLY CONTROL THE AMOUNT OF RADIATION REACHING THE IR BY TERMINATING THE LENGTH OF EXPOSURE
AUTOMATIC EXPOSURE DEVICES
PHOTOTIMING
ALSO ARE CALLED _____________________ OR _____________
Principle of Automatic Exposure Control Operation
Once a predetermined amount of radiation is transmitted through the patient, the xray exposure is terminated. This defines the exposure time and therefore the total of radiation exposure to the IR
PHOTOTIMERS
IONIZATION CHAMBERS
TWO TYPES OF AEC SYSTEMS HAVE BEEN USED
PHOTOTIMERS
FIRST GENERATION AEC
EXIT TYPE DEVICE
COINS THE TERM “PHOTOTIMING”
REFERS TO THE USE OF AEC DEVICES THAT USES PHOTOMULTIPLIERS TUBES OR PHOTODIODES
IONIZATION CHAMBERS
ENTRANCE TYPE DEVICE
IS A HOLLOW CELL THAT CONTAINS AIR AND IS CONNECTED TO THE TIMER CIRCUIT VIA AN ELECTRICAL WIRE
MAS READOUT AND ITS IMPORTANCE
USED DURING THE EXPOSURE IS CALCULATED BY THE SYSTEM AND DISPLAYED POSTEXPOSURE, PROVIDING VALUABLE FEEDBACK TO RADIOLOGIC TECHNOLOGISTS
NOT APPLICABLE
UNDERSTANDING MAS READOUT IS ESSENTIAL FOR LEARNING MANUAL TECHNIQUES, ESPECIALLY IN SCENARIOS WHERE AEC IS ________________
Automatic Exposure Control and mAs Readout
If the radiographic unit has an mAs display readout display, The radiographer should take note of the reading after an exposure is made. This information can be invisible
kVp and Automatic Exposure Control Response
THE SELECTION OF KVP IS CRITICAL; HIGHER KVP RESULTS IN GREATER PENETRATION, LEADING TO SHORTER EXPOSURE TIMES AND REDUCED PATIENT DOSE
mA and Automatic Exposure Control Response
MA VALUE SELECTION WILL AFFECT THE EXPOSURE TIME NEEDED BY THE AEC DEVICE: HIGHER MA IS REQUIRED TO LESSEN THE EXPOSURE TIME FOR A PARTICULAR EXAM
MINIMUM RESPONSE TIME
1ms (0.001s)
THE ______________________ OF MODERN AEC SYSTEMS IS APPROXIMATELY_____ _______, WHICH IS IMPORTANT FOR RAPID IMAGING SCENARIOS
LONGER
MINIMUM RESPONSE TIME IS USUALLY __________ WITH AEC SYSTEMS WITH OTHER TYPES OF RADIOGRAPHIC TIMERS
BACK UP TIME
________________ REFERS TO THE MAXIMUM LENGTH OF TIME FOR WHICH THE X-RAY EXPOSURE WILL CONTINUE WHEN USING AN AEC SYSTEM
OVEREXPOSURE
THE X-RAY TUBE FROM OVERHEATING
BACKUP TIME IS A SAFETY FEATURE THAT LIMITS THE MAXIMUM DURATION OF EXPOSURE, PROTECTING BOTH THE PATIENT FROM ______________ AND __________________________________
Function of Back up time
Backuptime, The maximum exposure time allowed during an AEC examination serves as a safety mechnanism when AEC is not used properly or is not functioning properly
DENSITY ADJUSTMENTS AND CONTROLS
AEC DEVICES ARE EQUIPPED WITH DENSITY CONTROLS THAT ALLOW THE RADIOGRAPHER TO ADJUST THE AMOUNT OF PRESET RADIATION DETECTION VALUES
25%
(+25% OR −25%)
(+50% OR −50%)
A COMMON INCREMENT IS _____, MEANING THAT THE PREDETERMINED EXPO-SURE LEVEL NEEDED TO TERMINATE THE TIMER CAN BE EITHER INCREASED OR DECREASED FROM NORMAL IN ONE INCREMENT___________ OR TWO INCREMENTS _______________
TRUE
DENSITY CONTROLS SHOULD ONLY BE UTILIZED WHEN THE IMAGE BRIGHTNESS IS INADEQUATE DESPITE PROPER POSITIONING, TO AVOID UNNECESSARY REPEAT EXPOSURES
Proper Detection
Poor Positioning
Practical applications and Limitation of AEC: Detector Selection and positioning
PROPER DETECTOR
_______________ SELECTION IS CRUCIAL; DETECTORS SHOULD BE SUPERIMPOSED BY THE ANATOMY OF INTEREST TO ENSURE ACCURATE EXPOSURE READINGS
POOR POSISTIONING
______________ CAN LEAD TO OVEREXPOSURE OR UNDEREXPOSURE, INCREASING THE LIKELIHOOD OF REPEAT IMAGING DUE TO INADEQUATE QUALITY
PATIENT CENTERING

IMPACT OF PATHOLOGY
CERTAIN PATHOLOGIES CAN SIGNIFICANTLY AFFECT AEC PERFORMANCE; FOR EXAMPLE, FLUID ACCUMULATION (LIKE PNEUMONIA) MAY REQUIRE LONGER EXPOSURE TIMES, WHILE CONDITIONS LIKE EMPHYSEMA MAY NECESSITATE SHORTER EXPOSURES
THE AEC SYSTEM IS DESIGNED TO COMPENSATE FOR CHANGES IN PATIENT THICKNESS
WHEN TO AVOID AED
AEC SHOULD NOT BE USED WHEN THE ANATOMY DOES NOT ADEQUATELY COVER THE DETECTORS OR WHEN PATHOLOGICAL CONDITIONS COULD MISLEAD THE DETECTORS, POTENTIALLY RESULTING IN INCORRECT EXPOSURE SETTINGS.
RADIOLOGIC TECHNOLOGISTS MUST ASSESS EACH SITUATION TO DETERMINE THE APPROPRIATENESS OF AEC USAGE.
SOMETIMES, THE BEST SOLUTION IS A MANUAL TECHNIQUE DETERMINED THROUGH USE OF A TECHNIQUE CHART. AEC IS NOT A REPLACEMENT FOR A KNOWLEDGEABLE RADIOGRAPHER USING CRITICAL THINKING SKILLS
ANATOMICALLY PROGRAMMED TECHNIQUE
REFERS TO A SYSTEM THAT ALLOWS THE RADIOGRAPHER TO SELECT A PARTICULAR BUTTON ON THE CONTROL PANEL THAT REPRESENTS AN ANATOMIC AREA; A PREPROGRAMMED SET OF EXPOSURE FACTORS IS DISPLAYED AND SELECTED FOR USE
Anatomically Programmed Technique and Patient Exposure