88 Terms

1
New cards

What was an older form of ultrasound that used a static photograph?

Static B scanners were an older form of ultrasound that produced a static photograph.

2
New cards

What were the potentiometers used for in static B scanners?

Potentiometers were manipulated at the joints of static B scanners to determine the depth of the ultrasound beam into the body.

3
New cards

What were the main pros and cons of static B scanners?

The main pro of static B scanners was a large, expanded field of view; the cons were that they took a long time to get an image due to requiring slow, coordinated movements, had to store information in memory before display, were bistable (eventually moving to a few shades of grey), used entirely mechanical transducers, had no ability to display moving structures, and were not ergonomic.

4
New cards

What type of ultrasound imaging scans without storing the image and displays the picture live or "on the fly"?

Real-time imaging (or dynamic imaging) scans without storing the image and displays the picture live or "on the fly" directly from the receiver to the display.

5
New cards

What is another term used to describe real-time imaging?

Dynamic imaging is another term used to describe real-time imaging.

6
New cards

Why is real-time imaging compared to a flip book?

Real-time imaging is compared to a flip book because it creates the illusion of movement by lining up consecutive frames with incremental movements.

7
New cards

Does real-time imaging bypass the scan converter?

Yes, real-time imaging bypasses the scan converter to display live images directly from the receiver to the display.

8
New cards

Why does a real-time ultrasound system still require a scan converter if it bypasses it during live scanning?

A real-time ultrasound system still requires a scan converter to allow "freeze frame" capabilities.

9
New cards

What resolution is introduced by the reliance on frame rate in real-time imaging?

Temporal resolution is introduced by the reliance on frame rate in real-time imaging.

10
New cards

What minimum frame rate must be exceeded to avoid the flicker rate of the human eye?

The frame rate must exceed 15 frames per second (fps) to avoid the flicker rate of the human eye and appear moving.

11
New cards

What is the desirable frame rate range for real-time ultrasound imaging?

A frame rate of 25 to 30 frames per second (or higher) is desirable for real-time ultrasound imaging.

12
New cards

Why is real-time imaging possible in ultrasound applications?

Real-time imaging is only possible because ultrasound has a high velocity in soft tissue.

13
New cards

What is the definition of frame rate in ultrasound?

Frame rate is defined as the number of frames displayed or scanned per second.

14
New cards

What is the average frame rate range in diagnostic ultrasound?

The average frame rate range in diagnostic ultrasound is between 10 and 60 frames per second.

15
New cards

What visual artifact occurs if the frame rate drops below 20 frames per second?

If the frame rate drops below 20 frames per second, image flickering occurs, causing eye fatigue and a choppy movie appearance.

16
New cards

What constitutes a single frame in ultrasound imaging?

A single frame is an individual image composed of multiple scan lines created by one complete scan of the ultrasound beam.

17
New cards

What is the definition of temporal resolution?

Temporal resolution is the ability of an instrument to separate two events closely spaced in time, or to accurately locate the position of moving structures at any given moment.

18
New cards

How does frame rate affect temporal resolution?

Higher frame rates improve temporal resolution, whereas lower frame rates degrade temporal resolution.

19
New cards

What is the formula for frame rate involving PRF, number of focal zones, and scan lines per frame?

The formula is FR = PRF / (n × N), where n is the number of focal zones and N is the number of scan lines per frame.

20
New cards

What is the frame rate formula when using a single focal zone?

When using a single focal zone, the formula is FR = PRF / Lines per Frame.

21
New cards

What is frame time, and how is it calculated?

Frame time is the amount of time it takes to create a single image, calculated as Frame Time = 1 / FR.

22
New cards

What is the formula for frame rate based on propagation speed, depth, and number of scan lines?

The formula is FR = C / (2 × Depth × N), where C is propagation speed and N is lines per frame.

23
New cards

What equation represents the maximum limit relationship between depth, focal zones, lines per frame, and frame rate?

The equation representing this relationship is Depth × n × N × FR ≤ 77,000 cm/s.

24
New cards

How is frame rate related to penetration depth?

Frame rate is inversely related to penetration depth.

25
New cards

Why does an increase in depth lead to a decrease in frame rate?

An increase in depth increases listening time and pulse repetition period (PRP), which decreases pulse repetition frequency (PRF) and leads to a decreased frame rate.

26
New cards

How is frame rate related to frame time?

Frame rate is inversely related to frame time.

27
New cards

Why does an increased number of focal zones decrease the frame rate?

An increased number of focal zones decreases frame rate because more focal zones require more than one pulse per scan line, taking longer to create an image.

28
New cards

How does line density affect frame rate and lateral resolution?

Increased line density requires more scan lines per frame, which decreases frame rate and degrades temporal resolution, but improves lateral resolution by creating a narrower, highly packed beam.

29
New cards

How does an increased field of view impact the frame rate?

An increased field of view creates a larger screen requiring more scan lines, which increases the time to generate an image and decreases the frame rate.

30
New cards

How does sending more pulses per scan line affect the frame rate?

Sending more pulses increases the waiting time before sending subsequent echoes, which lowers the frame rate.

31
New cards

How is frame rate related to pulse repetition frequency (PRF) and pulse repetition period (PRP)?

Frame rate is directly related to PRF and inversely related to PRP.

32
New cards

How is frame rate related to the propagation speed of sound in a medium?

Frame rate is directly related to propagation speed because faster returning echoes allow for faster image creation.

33
New cards

What is detail resolution?

Detail resolution describes the overall image resolution of a transducer, including lateral and axial resolutions.

34
New cards

What is lateral resolution, and what are its alternative names?

Lateral resolution is the ability to resolve objects perpendicular to the ultrasound beam, and its alternative names are azimuthal, transverse, and angular resolution (ATLA).

35
New cards

What adjustments can be made to narrow the beam and improve lateral resolution?

Lateral resolution is improved by decreasing the field of view, using a smaller crystal, increasing the number or depth of focal zones, or increasing line density.

36
New cards

What is axial resolution, and what are its alternative names?

Axial resolution is the ability to resolve objects parallel to the ultrasound beam, and its alternative names are longitudinal, radial, range, and depth resolution (LAARD).

37
New cards

What are the formulas used to calculate axial resolution?

The formulas are Axial Resolution = SPL / 2 or Axial Resolution = (n × λ) / 2, where n is the number of cycles in a pulse and λ is wavelength.

38
New cards

What factors can be modified to improve axial resolution by reducing the spatial pulse length?

Axial resolution is improved by anything that reduces spatial pulse length (SPL), such as increasing damping or increasing operating frequency.

39
New cards

What is contrast resolution, and how is it calculated based on bit depth?

Contrast resolution is the system's ability to distinguish between different shades of grey, calculated using the formula Shades of grey = 2^(number of bits in a pixel).

40
New cards

Why is temporal resolution critical for fast-moving structures in real-time imaging?

Temporal resolution is critical for fast-moving structures because it determines the system's ability to distinguish between two events occurring at slightly different times.

41
New cards

What is slice thickness resolution, and what are its alternative names?

Slice thickness resolution is the ability to distinguish structures in a three-dimensional plane, and its alternative names are elevational resolution and partial voluming.

42
New cards

Which crystal shape produces the best slice thickness resolution?

Disc-shaped crystals produce the best slice thickness resolution.

43
New cards

Why do linear, convex, and phased array transducers have worse elevational resolution than disc-shaped transducers?

Linear, convex, and phased array transducers have rectangular elements that can only be phased/focused side-to-side (lateral resolution), leaving elevational thickness un-phased across its height.

44
New cards

What is the order of resolution quality from best to worst in standard linear array transducers?

The order of resolution quality from best to worst in standard linear array transducers is axial, then lateral, then elevational resolution.

45
New cards

How do matrix array transducers improve elevational resolution?

Matrix array transducers feature numerous crystals arranged both side-to-side and top-to-bottom, allowing electronic phasing in all directions to thin the beam elevationally.

46
New cards

What defines spatial resolution in display systems?

Spatial resolution refers to display accuracy, which improves with an increased number of lines per frame or increased pixel density.

47
New cards

What is a transducer, and what physical principle does it utilize to operate?

A transducer is a device that utilizes piezoelectric principles (pressure electricity) to convert electrical energy into mechanical energy and vice versa.

48
New cards

What are the primary components of an ultrasound transducer?

The primary components are crystals, damping material, matching layer, electric shield, acoustic insulator, electrodes, and casing.

49
New cards

What component is the active element responsible for transmitting and receiving ultrasound signals?

The crystal (also called the active element or piezoelectric element) is responsible for transmitting and receiving ultrasound signals.

50
New cards

How do crystal thickness and propagation speed determine the operating frequency of a transducer?

Operating frequency is determined by the formula Operating Frequency = (propagation speed of crystal) / (2 × crystal thickness), meaning thinner crystals yield higher frequencies.

51
New cards

What crystal thickness is considered optimal relative to the wavelength?

The optimal crystal thickness is equal to λ / 2 (one-half of the wavelength).

52
New cards

How does crystal diameter influence the near zone length (NZL) and penetration depth?

A wider crystal diameter yields a longer near zone length (NZL = diameter² × frequency / 6) and deeper focus with a wider beam, while a narrower crystal yields a shorter NZL, shallower focus, and narrower beam.

53
New cards

What are examples of naturally occurring piezoelectric materials?

Examples of naturally occurring piezoelectric materials include quartz, rochelle salts, sugar cane, and topaz.

54
New cards

What is the most common synthetic piezoelectric material used in ultrasound transducers?

Lead zirconate titanate (PZT) is the most common synthetic piezoelectric material used.

55
New cards

What is the useful frequency range for standard diagnostic ultrasound applications?

The useful frequency range for standard diagnostic ultrasound applications is 2 to 20 MHz.

56
New cards

What ultrasound application frequencies can reach up to 50 MHz?

Frequencies up to 50 MHz may be used in ophthalmologic, dermatologic, and intravascular imaging.

57
New cards

What is the main function of the damping material in a pulsed-wave transducer?

The main function of the damping material is to reduce the number of cycles in a pulse by absorbing extra ringing energy.

58
New cards

What materials are used to construct the damping layer?

Damping material is made of epoxy resin mixed with tungsten powder.

59
New cards

What are the pros and cons of adding damping material to a transducer?

The pros of damping material are improved axial resolution (via reduced SPL and pulse duration) and broadened bandwidth; the con is decreased transducer sensitivity due to reduced signal amplitudes.

60
New cards

What is bandwidth, and what is the formula to calculate it?

Bandwidth is the range of frequencies emitted by a transducer, calculated using the formula Bandwidth = f₂ - f₁.

61
New cards

What defines a wide bandwidth transducer?

A wide bandwidth transducer has a fractional bandwidth of at least 70%.

62
New cards

What is the quality factor (Q-factor), and what is its formula?

Quality factor (Q-factor) measures the degree of damping/ringing in a sound wave, calculated as Q-factor = Resonant frequency / Bandwidth.

63
New cards

Why do pulsed-echo ultrasound systems require a low Q-factor?

Pulsed-echo ultrasound requires a low Q-factor because low Q-factor corresponds to high bandwidth and short spatial pulse length.

64
New cards

What is the purpose of the matching layer in an ultrasound transducer?

The matching layer protects the active element and reduces reflection at the transducer/skin interface by decreasing impedance mismatch.

65
New cards

What is the optimal thickness of the matching layer relative to wavelength?

The optimal thickness of the matching layer is λ / 4 (one-quarter of the wavelength).

66
New cards

What is the order of impedance from highest to lowest between the crystal, matching layer, gel, and skin?

The order of impedance from highest to lowest is crystal, matching layer, gel, and skin.

67
New cards

How do mechanical and electronic steering and focusing methods differ?

Mechanical steering and focusing rely on physical motors, curved crystals, or lenses, whereas electronic steering and focusing rely on phased time delays across multiple electrical circuits.

68
New cards

What is the difference between internal and external mechanical focusing?

Internal mechanical focusing is achieved by curving the active crystal itself, whereas external mechanical focusing is achieved by placing an acoustic lens or mirror in front of the crystal.

69
New cards

How do transmit focusing and receive focusing operate electronically?

Transmit focusing delays electrical signal pulses during transmission to curve the beam wavefront, while receive focusing electronically delays returning echo signals as they arrive at the transducer.

70
New cards

What is dynamic focusing?

Dynamic focusing occurs when an ultrasound system applies electronic focusing during both transmission and reception.

71
New cards

What is variable aperture, and how does it maintain a constant focal width?

Variable aperture selectively activates smaller or larger groups of array elements to adjust beam diameter, increasing the active aperture size for deeper focal lengths to maintain a consistent beam width.

72
New cards

How is line density calculated for rectangular image formats versus sector image formats?

For rectangular formats, Line Density = Lines per frame / Display width (cm); for sector formats, Line Density = Lines per frame / Sector angle (degrees).

73
New cards

What are the primary characteristics, pros, and cons of single-element mechanical transducers?

Single-element mechanical transducers use one circular disc crystal rotated or rocked by a motor to produce sector images with fixed mechanical focus; their pros are low cost and wide field of view, while their cons are slow frame rates, lack of ergonomics, and complete image loss if the single crystal breaks.

74
New cards

How do mechanical rotating transducers steer and focus the sound beam?

Mechanical rotating transducers steer the beam using a motor-driven wheel mounted with several crystals suspended in protective oil, and focus the beam mechanically.

75
New cards

How do mechanical oscillating transducers create a sector image?

Mechanical oscillating transducers create a sector image by rocking a single crystal back and forth in a side-to-side motion via a motor.

76
New cards

How does a mirror-driven mechanical transducer steer the ultrasound beam?

A mirror-driven mechanical transducer holds its crystal stationary and steers the sound beam using a motor-driven acoustic mirror.

77
New cards

Why are annular array transducers described as the link between electronic and mechanical designs?

Annular array transducers are described as an electronic/mechanical link because they steer the beam mechanically with a motor while focusing the beam electronically using concentric ring elements.

78
New cards

What image shape is produced by an annular array transducer?

An annular array transducer produces a sector-shaped image.

79
New cards

What is an array transducer, and how do phasing and sequencing operations differ?

An array transducer contains multiple active elements; phasing fires most or all elements with small time delays (elements ¼λ wide), whereas sequencing fires successive small groups of elements in order (elements 1λ wide).

80
New cards

What image shapes are generated by linear, convex, phased, vector, and annular array transducers?

Linear arrays produce rectangular images, convex arrays produce sector images with a scooped top, phased arrays produce sector images starting at a point, vector arrays produce trapezoidal (blunted sector) images, and annular arrays produce sector images.

81
New cards

What are the main clinical applications and operational characteristics of linear sequential array transducers?

Linear sequential arrays use 200-400 elements fired in groups sequentially to produce rectangular images, typically operating at high frequencies for vascular, musculoskeletal, and small parts imaging.

82
New cards

How are elements arranged and fired in a convex array transducer?

Elements in a convex array transducer are arranged in a curved arc and fired sequentially using phasing delays.

83
New cards

How does a linear phased array steer and focus the beam electronically?

A linear phased array steers the beam by applying an electronic slope (sequential time delays across all elements) and focuses the beam by applying an electronic curvature (arc-shaped time delays).

84
New cards

What image shape is produced by a vector array transducer?

A vector array transducer produces a trapezoidal (blunted sector) image by combining linear sequencing with phased steering.

85
New cards

Why do invasive transducers use higher frequencies than external transducers?

Invasive transducers use higher frequencies because placing the transducer inside a body orifice shortens the sound path length, requiring less penetration and providing higher image resolution.

86
New cards

What is subdicing in matrix array transducers, and how does it affect beam quality?

Subdicing divides each active element into 3 to 8 smaller sections, which eliminates side lobe and grating lobe artifacts and improves short-axis resolution.

87
New cards

How does apodization assist in reducing transducer artifacts?

Apodization reduces side lobe and grating lobe artifacts by varying the electrical voltage amplitudes applied to inner versus outer transducer elements.

88
New cards

What additional plane of resolution is provided by 2D and 1.5D matrix arrays?

2D and 1.5D matrix arrays provide improved elevational (slice thickness) resolution across the 3D plane.