Tube Malposition / Blockage
Cause: Kinking, dependent loops, clots, improper placement
Tube Malposition / Blockage
Sign: No tidaling
Tube Malposition / Blockage
Sign: Absent drainage
Tube Malposition / Blockage
Sign: Worsening dyspnea
Tube Malposition / Blockage
Sign: Absent breath sounds
Tube Malposition / Blockage
Management: Reposition patient; unkink tubing
Tube Malposition / Blockage
Management: Milk/squeeze gently ONLY if policy allows
Tube Malposition / Blockage
Management: Notify physician; order chest X-ray
Tube Malposition / Blockage
Management note: Routine milking/stripping is NOT recommended — risk of high negative pressure & tissue damage
Persistent Air Leak
Cause: Lung injury or a leak in the system itself
Persistent Air Leak
Sign: Continuous bubbling in the water-seal chamber
Persistent Air Leak
Management: Check and tighten/tape all connections; if the leak persists, notify the physician (possible bronchopleural fistula)
Accidental Disconnection
Cause: Tubing separates from the drainage system
Accidental Disconnection
Sign: Bubbling stops; risk of air entry into the chest
Accidental Disconnection
Management: Place the tube end in sterile water; reconnect using sterile technique; notify the provider
Accidental Tube Removal / Dislodgment
Cause: Tube pulls out of the insertion site
Accidental Tube Removal / Dislodgment
Sign: Open chest wound, sucking sound, respiratory distress
Accidental Tube Removal / Dislodgment
Management: Cover the site with a sterile occlusive dressing taped on 3 sides; provide oxygen; notify the physician urgently
Subcutaneous Emphysema
Cause: Air leaking into the surrounding soft tissue
Subcutaneous Emphysema
Sign: Swelling with a "crackling" (crepitus) sensation on palpation
Subcutaneous Emphysema
Management: Assess extent and airway; ensure the tube is not kinked; notify the physician
Rapid Recall — Unit 2 Exam Cues: No tidaling
No rise/fall of fluid in the tubing with respiration
Rapid Recall — Unit 2 Exam Cues: No tidaling
Think blockage, kink, or lung re-expansion
Rapid Recall — Unit 2 Exam Cues: No tidaling
Always check blockage, kink, or lung re-expansion in that order
Rapid Recall — Unit 2 Exam Cues: Continuous bubbling in the water-seal chamber
AIR LEAK
Rapid Recall — Unit 2 Exam Cues: Continuous bubbling in the water-seal chamber
Different from gentle bubbling in the SUCTION chamber, which is normal for wet suction
Rapid Recall — Unit 2 Exam Cues: Crepitus/crackling on palpation around the tube site
Subcutaneous emphysema
Rapid Recall — Unit 2 Exam Cues: Golden rule
NEVER clamp a chest tube that is bubbling/has an air leak unless specifically ordered
Rapid Recall — Unit 2 Exam Cues: Golden rule
Clamping can cause a tension pneumothorax
Pneumothorax
Air in the pleural space
Hemothorax
Blood in the pleural space
Pleural Effusion
Excess (usually serous) fluid in the pleural space
Empyema
Accumulation of infected fluid or pus in the pleural space
Chylothorax
Accumulation of chyle (a milky-white lymphatic fluid) in the pleural space
Chylothorax
Chyle contains fat (especially triglycerides)
Chylothorax
Chyle contains lymph
Chylothorax
Chyle contains proteins
Chylothorax
Chyle contains lymphocytes
Chylothorax
Chyle contains fat-soluble vitamins
"Milky-White" Drainage Fluid
Chylothorax (chyle), NOT infection
"Pus / Infected Fluid"
Empyema
Triangle of Safety
The most important landmark for chest tube insertion, located on the lateral chest wall
Triangle of Safety
Keeps the tube away from major vessels, nerves, and muscle
Triangle of Safety Posterior Border
Anterior border of the latissimus dorsi
Triangle of Safety Anterior Border
Lateral border of the pectoralis major
Triangle of Safety Inferior Border
Horizontal line at the level of the nipple (roughly the 5th intercostal space)
Triangle of Safety Apex
Below the axilla (base of the axilla)
Three-Bottle (or 3-Chamber) System
Most commonly used traditional setup
Three-Bottle (or 3-Chamber) System
Now built into a single modern disposable unit, but conceptually still 3 chambers
Three-Bottle (or 3-Chamber) System
Combines a fluid collection chamber, a one-way water-seal chamber, and a suction control chamber
Three-Bottle (or 3-Chamber) System
Safer and more effective than simpler systems when controlled suction is needed
Wet Suction Control
Suction regulation uses a water column
Wet Suction Control
Uses water for suction control: Yes
Wet Suction Control
Example setting is −20 cm H2O
Wet Suction Control
Bubbling in the suction-control chamber is usually gentle, continuous bubbling
Wet Suction Control
Requires a suction source when ordered: Yes
Wet Suction Control
Main nursing concern is maintaining the correct water level (evaporation changes the suction pressure)
Dry Suction Control
Suction regulation uses a mechanical regulator / dial
Dry Suction Control
Uses water for suction control: No
Dry Suction Control
Example setting is −20 cm H2O
Dry Suction Control
Bubbling in the suction-control chamber is usually not required
Dry Suction Control
Requires a suction source when ordered: Yes
Dry Suction Control
Main nursing concern is verifying the correct dial/indicator setting
Wet Suction (Rapid Recall)
Gentle continuous bubbling in the SUCTION chamber is expected and normal
Vigorous/Violent Bubbling in the Suction Chamber
Suction is turned up too high, not a good thing — turn down until gentle
Bubbling in the Water-Seal Chamber
Reflects an air leak, not suction strength
Tube Comes Out of the Patient (Accidental Removal)
First priority is to seal the insertion site to prevent air from entering the chest
Tube Comes Out of the Patient (Accidental Removal) Step 1
Ask the client to cough and exhale immediately
Tube Comes Out of the Patient (Accidental Removal) Step 2
Apply a sterile occlusive dressing (e.g., Vaseline/petroleum gauze)
Tube Comes Out of the Patient (Accidental Removal) Step 3
Tape on 3 sides only — allows air to escape, prevents air from entering, and reduces the risk of a tension pneumothorax
Tube Comes Out of the Patient (Accidental Removal) Step 4
Call the provider immediately
Tube Comes Out of the Patient (Accidental Removal) Step 5
Monitor closely for respiratory distress
Tube Disconnects From the Drainage System
First priority is to prevent air from being sucked back into the pleural space
Tube Disconnects From the Drainage System (Simple Disconnection)
Apply an antiseptic swab to the tubing ends and reconnect
Tube Disconnects From the Drainage System (Damaged Bottle/Unit)
Place the end of the chest tube into sterile water or saline (~2 cm depth) to create a temporary water seal, then notify the provider
3-Sides Tape Rule
Taping an occlusive dressing on 3 sides creates a one-way "flutter valve"
3-Sides Tape Rule
Air can still escape out the open side (prevents tension pneumothorax) but cannot be sucked back in
4-Sided Taping
Never tape all 4 sides over an open pneumothorax wound — this traps air in and can cause a tension pneumothorax
Tube Out of the Patient vs. System
Tube out of the PATIENT → cover the site
Tube Out of the Patient vs. System
Tube out of the SYSTEM → protect the open tubing end (sterile water)