Digital calibration
Each individual electrode is compared to a system reference
Analog calibration verifies the 70 Hz high filter standard setting by assessing which of the following?
The sharpness of the tip of the calibration signal
Performing square wave and biological calibration on a digital EEG instrument does which of the following?
Verifies standard instrument settings are correct before beginning the recording
The energy internal to the EEG instrument used to perform square wave calibration is called ____________.
Input voltage
If the activity of input 1 is more negative than input 2:
The EEG deflection will go up
If the EEG amplitude is high, you should
Decrease your sensitivity
The purpose of the differential amplifier is:
cancel out external noise in both inputs
The bandwidth is the...
The frequencies inside the set filter settings
The function of the HFF is:
Attenuate the high frequencies without affecting the low frequencies
What causes Phase Lead?
Low Frequency Filter
What causes Phase Lag?
High Frequency Filter
The cutoff frequency
the frequency at which there will be a 30% reduction in amplitude of the signal
The time constant in the HFF
the time it takes for the calibration signal to rise 63%
A notch filter
removes only the frequency inside the notch filter
Lowering the HFF could result in:
decrease muscle artifact
To better visualize spikes:
increase the LFF
What is called a sensor on a digital EEG system?
the scalp electrode
A signal that is continuous without breaks is called:
an analog signal
The second component of the digital EEG system is:
an analog amplifier
The purpose of the analog amplifier is to:
boost the waveform for further processing
The purpose of the ADC circuit is to:
transform the signal into a number
The introduction of a virtual signal into the waveform is called
aliasing
Current EEG systems have a _____ ADC processor.
16 bit
The benefit to having a large range of amplitude for a digital EEG system is:
you can see all high and low amplitude signals with good resolution
Dwell time is:
the length of time between to sample points
The Nyquist Theorem
the fastest signal that can be recorded is the frequency that is half the sampling rate of the digitizer.
What happens when you record a signal faster than the Nyquist frequency?
your signal has less peaks than the original signal
ACNS recommends a sampling rate of at least:
256
Extra receptacles in the jackbox may used to record _____________.
any signal of cerebral or artifactual origin
If the inputs of the differential amplifier are equipotential what do you expect the output signal to be?
no output signal (flat line)
If the inputs of the differential amplifier are in phase, but not equipotential what do you expect the output signal to be?
A small output signal
Reducing (going to a lower number) the high filter setting does which of the following?
attenuates fast frequencies
Changing the low filter setting to a lower number is the equivalent of which of the following?
changing to a longer time constant
Reducing the setting of the low pass filter will affect which of the following?
attenuates spikes
Which of the following frequencies is most attenuated by a high filter setting of 35 Hz?
100 Hz
Which of the following is NOT affected by a high filter setting of 15 Hz?
theta
Which of the following is affected by the Notch Filter?
60 Hz
Time Constant
square wave calibration
Which of the following accentuates focal slow activity?
long time constant
What is the standard paper speed for the recording of EEG?
10 pages per second
Which of the following is NOT a reason to use fast (5 seconds per page) paper speed?
better visualize periodic activity
The EEG is being recorded with standard sensitivity and filter settings. When excessive muscle cannot be reduced or eliminated by patient relaxation the technologist should do which of the following?
change the high (low pass) filter from 70 Hz to 35 Hz or to 15 Hz