Why are instruments needed to detect ionizing radiation?
The human eye does not perceive ionizing radiation.
What do instruments use to detect ionizing radiation?
Radiation-induced physical effects in matter.
What do instruments convert radiation-induced physical effects into?
Measurable signals.
what are the 3 most relevant effects for instrumentation in diagnostic radiology are
production of electric charge by gas ionization
scintillation (luminescent excitation of solids)
electron-hole pairs (semiconductors)
What is the first relevant physical effect used for detection in diagnostic radiology?
Production of electric charge by gas ionization.
What is scintillation in radiation detection?
Luminescent excitation of solids.
What physical process enables radiation detection in semiconductors?
Creation of electron–hole pairs and charge transport.
What signals can be collected during radiation detection?
Current, pulses or light.
How is the signal collected during radiation detection?
Through an electronic chain.
What does the electronic chain in radiation detection allow?
Reliable measurement and metrological traceability.
What is the useful phenomenon in most radiation detectors?
The appearance of electric charge in an active volume.
What charges are generated by radiation in the active volume of most detectors?
Positive ions and electrons (or holes).
What separates the charges generated by radiation in most detectors?
An applied electric field.
What collects the charges generated in a detector's active volume?
The electrodes.
What is generated when charges are collected by detector electrodes?
A current or pulses.
What determines the relationship between a detector's current or pulses and the incident radiation?
Their amplitude or number relates to the incident radiation.
What are the two possible functions of a radiation detector, depending on its design?
A counter or a spectrometer.
What does a radiation detector acting as a counter measure?
The presence and rate of radiation.
What information does a radiation detector acting as a spectrometer provide?
Energy information.
Gaseous ionization detectors
Simple explanation
Gaseous ionization detectors contain a volume of gas between polarized electrodes. When radiation enters the gas, it creates ion–electron pairs. An electric field collects these charges, producing a pulse.
The pulse amplitude depends on the applied voltage. The PDF identifies three operating regions of interest:
Ionization chamber: all pairs reach the electrodes without multiplication.
Proportional counter: electrons cause secondary ionizations, multiplying the charge.
Geiger–Müller counter: a gaseous avalanche propagates throughout the volume, producing pulses of nearly the same amplitude regardless of energy.
done
Gaseous ionization detectors
What do gaseous ionization detectors contain?
A gas volume between polarized electrodes.
Gaseous ionization detectors
What does radiation create in the gas of a gaseous ionization detector?
Ion–electron pairs.
Gaseous ionization detectors
What collects the ion–electron pairs in a gaseous ionization detector?
The electric field.
Gaseous ionization detectors
What does charge collection produce in a gaseous ionization detector?
A pulse.
Gaseous ionization detectors
What does the pulse amplitude depend on in a gaseous ionization detector?
the applied voltage.
Gaseous ionization detectors
How many operating regions of gaseous ionization detectors are of interest
3
Gaseous ionization detectors
what are the operating regions
ionization chamber (saturation region)
proportional counter (charge multiplication)
Geiger-Muller counter
Gaseous ionization detectors
In which operating region does an ionization chamber work?
The saturation region.
Gaseous ionization detectors
What happens to the ion–electron pairs in an ionization chamber?
All pairs reach the electrodes without multiplication.
Gaseous ionization detectors
What is the signal proportional to in an ionization chamber?
the primary charge
Gaseous ionization detectors
What is the typical bias voltage of an ionization chamber?
200–250 V.
Gaseous ionization detectors
In which operating region does a proportional counter work?
The proportional region.
Gaseous ionization detectors
What causes charge multiplication in a proportional counter?
Electrons gain enough energy to cause secondary ionizations.
Gaseous ionization detectors
What is the pulse amplitude proportional to in a proportional counter?
The initial ionization.
Gaseous ionization detectors
What can a proportional counter be used as?
A spectrometer.
Gaseous ionization detectors
What is the approximate bias voltage of a proportional counter?
About 500 V, depending on geometry.
Gaseous ionization detectors
In which operating region does a Geiger–Müller counter work?
The Geiger region.
Gaseous ionization detectors
What happens in the gas of a Geiger–Müller counter?
A gaseous avalanche propagates throughout the volume.
Gaseous ionization detectors
How do Geiger–Müller counter pulse amplitudes compare?
All pulses have nearly the same amplitude, regardless of energy.
Gaseous ionization detectors
What sensitivity is noted for Geiger–Müller counters?
Excellent sensitivity for charged particles.
Gaseous ionization detectors
Do Geiger–Müller counters provide spectrometric information?
no
Gaseous ionization detectors
What happens in the recombination region of a gaseous ionization detector?
Charges annihilate in the gas before reaching the electrodes.
What determines the exact operating-region voltage values of gaseous ionization detectors?
Detector design, including the gas used and geometry.

read
done
i want you to name all the gaseous ionization detectors again !
ionization chamber
proportional counter
Geiger-Muller counter
what do Scintillation detectors do
collect electrons and holes (positions of free electrons)
Environmental and personal dosimetry
Dosimetry quantifies the energy deposited by radiation in matter. For personal dosimetry, the important quantity is the personal dose equivalent,
The value of d represents the depth in millimetres:
Hp(10): deep dose
External personal dosimetry is mandatory for category A workers and recommended for category B workers. Internal dosimetry applies when radionuclide intake is possible.
Continuous radiological control includes periodic measurement of the doses accumulated by each individual using calibrated, traceable personal dosimeters
done
Environmental and personal dosimetry
What does personal dosimetry evaluate?
The personal dose equivalent,
Environmental and personal dosimetry
What does d represent in the personal dose equivalent Hp(d)
depth in millimetres
Environmental and personal dosimetry
Which personal dose equivalent corresponds to deep dose?
Hp(10)
Environmental and personal dosimetry
For which workers is external personal dosimetry mandatory?
category A workers
Environmental and personal dosimetry
For which workers is external personal dosimetry recommended?
Category B workers.
Environmental and personal dosimetry
When does internal dosimetry apply?
When radionuclide intake is possible.
Environmental and personal dosimetry
How are the doses accumulated by each individual measured during continuous radiological control?
Periodically, using calibrated, traceable personal dosimeters.
Environmental and personal dosimetry
Thermoluminescent dosimeters (TLDs)
What is thermoluminescence?
The emission of light upon heating a previously irradiated material.
Environmental and personal dosimetry
Thermoluminescent dosimeters (TLDs)
What causes light emission during thermoluminescence?
Electrons trapped in discrete levels (traps) created by defects or dopants.
Environmental and personal dosimetry
Thermoluminescent dosimeters (TLDs)
What is a TL glow curve?
The emitted light as a function of temperature.
Environmental and personal dosimetry
Thermoluminescent dosimeters (TLDs)
What are the four advantages of TLDs
Reusable,
good performance over a wide dose range,
tissue equivalence
and small size.
Environmental and personal dosimetry
Thermoluminescent dosimeters (TLDs)
What are the three drawbacks of TLDs
Destructive readout, fading and the need for tight thermal control.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What are direct-reading electronic dosimeters based on?
Geiger tubes or Si diodes.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What electronic features do direct-reading electronic dosimeters have?
Miniaturized electronics, alarms and non-volatile memories.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What do direct-reading electronic dosimeters provide in real time?
Estimates of rate and dose.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What connectivity do direct-reading electronic dosimeters provide?
Computer connectivity.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What power-related limitation do direct-reading electronic dosimeters have?
Batteries are required.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What weight and cost limitations do direct-reading electronic dosimeters have?
Higher weight and cost.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
What calibration do direct-reading electronic dosimeters require?
Individual calibration.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
How are direct-reading electronic dosimeters accepted in some countries?
As official dosimeters.
Environmental and personal dosimetry
Direct-reading electronic dosimeters
How are direct-reading electronic dosimeters used in other countries?
As operational or alarm dosimeters.
Measurement of dose in the direct beam
What do measurements in the direct X-ray beam allow?
Optimization of procedures and estimation of risks to justify techniques.
Measurement of dose in the direct beam
Which two types of dosimeters are used for direct-beam measurements?
Suitable ionization chambers and TLDs.
Measurement of dose in the direct beam
Where are ionization chambers and TLDs placed for direct-beam measurements?
On the patient or phantom.
Measurement of dose in the direct beam
Which recommended dose magnitude, other than DAP, is used to monitor patient dose?
Entrance-surface dose (ESD).
Measurement of dose in the direct beam
Is entrance-surface dose (ESD) directly measurable?
yes
Measurement of dose in the direct beam
What is entrance-surface dose (ESD)?
Absorbed dose in air on the beam axis at the patient’s or phantom’s entrance surface, including the backscattered contribution.
Measurement of dose in the direct beam
How is entrance-surface dose (ESD) determined?
with TLDs placed on the skin surface.
Measurement of dose in the direct beam
Why does the small size of TLDs help when measuring entrance-surface dose?
Their small size minimizes artefacts on the image.