Gas Chromatography

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/50

encourage image

There's no tags or description

Looks like no tags are added yet.

51 Terms

1
New cards

Gas Chromatography

  • ANALYTES MUST BE IN GAS PHASE (straight-forward interaction)

  • fundamental limitations = volatility, thermal stability


<ul><li><p>ANALYTES MUST BE IN GAS PHASE (straight-forward interaction)</p></li><li><p>fundamental limitations = volatility, thermal stability</p></li></ul><p></p>
2
New cards

GC Sample Types

1) Gas

  • gas-tight syringe

  • Headspace Sampling (HS)


2) Liquid

  • volatile solvent injected into injection port → if not volatile, will lead to contamination

  • some degree of sample prep needed (matrix dependent)


3) Solid (more limited)

  • Headspace Sampling (HS)


3
New cards

Headspace Sampling (HS)

extraction is from a liquid or solid sample after being heated (it evaporates into the “headspace” — the gas in the headspace is what we use)

  • temp between 60-90 degrees C → won’t get macromolecules like proteins

  • functional group expected with this = alcohols (BAC) or accelerants (organic solvents = very volaile)

  • can do this with MS or FID

    • good for chemical profiling in full scan mode


4
New cards

Good for GC Analytes

  • boiling point up to 400 degrees C

  • non-polar or slightly polar

  • thermally stable


5
New cards

BAD for GC Analytes

  • do not vaporize at GC normal operation conditions → degrade at injection port and contaminate

  • highly reactive

  • thermally unstable

  • strongly polar


6
New cards

Derivatization

if your target analytes are not volatile or thermally stable → process of chemically modifying a compound to produce a new molecule suitable for GC analysis

  • more volatile (non-polar), more thermally stable (no degradation), or more useful detection characteristics (molecular weight information)

  • increases sensitivity and chromatographic performance for specific compounds

  • as soon as you derivatize, fragments can immediately change (happens before injection)


7
New cards

Why are substances with reactive/acidic hydrogens (acids, alchs, amines) difficult to analyze?

they react with the surface of the injection port or the analytical column and result in tailing peaks and low response

  • Acidic Hydrogens = proton donors (lose a proton to the medium)


8
New cards

What are the three reactions used for Derivization?

  1. Silylation

  2. Acylation

  3. Alkylation


9
New cards

Silylation

introduction of a silyl moiety (RCO-) by substitution of an active H (-OH, -COOH, -SH, -NH, -NH2) 

  • Reagents: BSTFA, MSTFA, MTBSTFA 

  • Works with: amides, -OH, -CO2H, -NH2 

    • Amines = harder to derivitize the second time 

  • ex) Amphetamines, benzodiazepines, cannabinoids, BE (cocaine metabolite) 

    • Cocaine does not need to be derivatized 


<p><span style="line-height: 22.0875px;">introduction of a silyl moiety (RCO-) by substitution of an active H (-OH, -COOH, -SH, -NH, -NH2)&nbsp;</span></p><ul><li><p><span style="line-height: 22.0875px;"><strong>Reagents: </strong>BSTFA, MSTFA, MTBSTFA&nbsp;</span></p></li></ul><ul><li><p class="Paragraph SCXW181811248 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;"><strong>Works with: </strong>amides, -OH, -CO2H, -NH2&nbsp;</span></p><ul><li><p class="Paragraph SCXW181811248 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;">Amines = harder to derivitize the second time&nbsp;</span></p></li></ul></li></ul><ul><li><p class="Paragraph SCXW181811248 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;">ex) Amphetamines, benzodiazepines, cannabinoids, BE (cocaine metabolite)&nbsp;</span></p><ul><li><p class="Paragraph SCXW181811248 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;">Cocaine does not need to be derivatized&nbsp;</span></p></li></ul></li></ul><p></p>
10
New cards

Acylation

introduction of an acyl moiety (RCO-) by substitution of an active H (-OH, -COOH, -SH, -NH, -NH2) 

  • Reagents: TFAA, PFPA, HFBA 

  • ex) Amphetamines, cannabinoids, BE (cocaine metabolite) 


<p><span style="line-height: 22.0875px;">introduction of an acyl moiety (RCO-) by substitution of an active H (-OH, -COOH, -SH, -NH, -NH2)&nbsp;</span></p><ul><li><p><span style="line-height: 22.0875px;"><strong>Reagents</strong>: TFAA, PFPA, HFBA&nbsp;</span></p></li></ul><ul><li><p class="Paragraph SCXW153301855 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;">ex) Amphetamines, cannabinoids, BE (cocaine metabolite)&nbsp;</span></p></li></ul><p></p>
11
New cards

Alkylation

substitution of an alkyl or aryl group at an active H (-OH, -COOH, -SH, -NH, -NH2) 

  • Ends with: ethers, esters, thioethers, thioesters, N-alkyl amines or N-alkyl amides 

  • Reagents: BF3/MeOH 


<p><span style="line-height: 22.0875px;">substitution of an alkyl or aryl group at an active H (-OH, -COOH, -SH, -NH, -NH2)&nbsp;</span></p><ul><li><p><span style="line-height: 22.0875px;"><strong>Ends with: </strong>ethers, esters, thioethers, thioesters, N-alkyl amines or N-alkyl amides&nbsp;</span></p></li></ul><ul><li><p class="Paragraph SCXW54276600 BCX0" style="text-align: left;"><span style="line-height: 22.0875px;"><strong>Reagents:</strong> BF3/MeOH&nbsp;</span></p></li></ul><p></p>
12
New cards

Oven

three components of GC under temp control

  • sample injection chamber/port

  • column

  • detector


13
New cards

Elements that Affect Separation

  1. Solid Phase (type and thickness)

  2. Carrier Gas (type and linear velocity) = He or H

  3. Column Internal Diameter

  4. Column length

  5. Oven temperature → only parameter you can change in terms of separation outside of the column


14
New cards

Why is carrier gas important?

key aspect for the determining performance of the system and method → efficiency is a function of linear velocity

  • affects column efficiency and resolution

  • must be chemically inert (why O2 is bad)

  • performance (want a smaller H)

  • cost

  • safety

  • van deemter plot

  • He or H more comparable than N2

    • solutes diffuse more rapidly through H2 and He (decreased resistance to mass transfer)


15
New cards

Sample Introduction System

  1. For optimum column efficiency, the sample should be introduced onto the column as a “plug of vapor” → usually inject 1-2 uL

  2. Syringe used to inject sample through a rubber septum into a flash vaporizer port at the head of the column

  3. Injection port temp is ~50 degrees C higher than the boiling point of the least volatile component of the sample


16
New cards

Manual Injection

  • rinse syringe with appropriate solvent

  • rinse syringe with sample

  • slowly draw up desired volume (1-2 uL) & avoid air bubbles

  • carefully insert the syringe through the septum

    • will get less than 1 uL most of the time (some will be air) = LIMITATION

  • let needle breifly heat up

  • QUICKLY depress plunger & pull syringe out


solvent you select for washes should have same polarity of sample your injecting

  • want to make sure any residual solvent is dissolved

  • number of washes up to user


YOU are autosampler


17
New cards

Autosampler Injection

more robust and reproducible compared to manual injection (instantaneous injection)

  • can analyze 100 samples with identical sample volume and injection speed

  • wash and waste vials

  • eliminates human error because its a machine (harder to personally control each sample = coordination of time and when its time to inject sample)

    • autosampler could glitch


18
New cards

Inlet Liner

important part within injection port (glass tube) = increases the inert surface area for vaporization

  • limit sample degradation

  • enhance vaporization

  • guiding syringe needle into the correct position → choice of inlet liner can have a noticeable effect on chromatogram

  • injection port is heated independently of the GC oven (independent)

  • location where sample is volatized


can you liner ~100x before replacing (very general though, depends on what sample is being used) → if more complex sample (keep eye on it & sample inlet temp)


19
New cards

Ways to Inject Liquid Samples

  1. Splitless Injection Mode

  2. Split Injection Mode


20
New cards

Splitless Injection Mode

entire sample is injected into the column → no separation

  • split vent = closed

  • constant inlet flow

  • most commonly used in forensic tox


PROS = lower detection limits because more sample is introduced

CONS = overloading column, band broadening, leak = O2 entering through hole


<p>entire sample is injected into the column → no separation</p><ul><li><p>split vent = closed</p></li><li><p>constant inlet flow</p></li><li><p>most commonly used in forensic tox</p></li></ul><p></p><p><strong>PROS</strong> = lower detection limits because more sample is introduced</p><p><strong>CONS</strong> = overloading column, band broadening, leak = O2 entering through hole</p><p></p>
21
New cards

Split Injection Mode

set ratio of sample is sent through the injection port to the column

  • split vent is open = part of your sample will not enter the column (large portion of sample discarded)

  • you determine what ratio of sample you want


PROS = concentrated samples, protects the column, sharper peaks

CONS = higher detection limit because less sample is introduced


<p>set ratio of sample is sent through the injection port to the column</p><ul><li><p>split vent is open = part of your sample will not enter the column (large portion of sample discarded)</p></li><li><p>you determine what ratio of sample you want</p></li></ul><p></p><p><strong>PROS = </strong>concentrated samples, protects the column, sharper peaks</p><p><strong>CONS = </strong>higher detection limit because less sample is introduced</p><p></p>
22
New cards

Headspace Injection

mechanism to extract volatile components from non-volatile matrices (sample prep as well)

  • set temp & time for equilibrium between liquid and gas phases

  • heated syringe injected directly into injection port (incubation at temps higher than room temp (60-90 degrees C)) → sample is agitating, volatile samp will vaporize

  • GOAL = break equilibrium between liquid and gas phase so we can sample just the gas phase with needle

  • if not volatile = won’t work (ex. sugars, proteins, etc)

  • SYRINGE MUST BE GAS TYPE TO PREVENT LOSS

  • ex) BAC & fire debris analysis


<p>mechanism to extract volatile components from non-volatile matrices (sample prep as well)</p><ul><li><p>set temp &amp; time for equilibrium between liquid and gas phases</p></li><li><p>heated syringe injected directly into injection port (incubation at temps higher than room temp (60-90 degrees C)) → sample is agitating, volatile samp will vaporize</p></li><li><p>GOAL = break equilibrium between liquid and gas phase so we can sample just the gas phase with needle</p></li><li><p>if not volatile = won’t work (ex. sugars, proteins, etc)</p></li><li><p>SYRINGE MUST BE GAS TYPE TO PREVENT LOSS</p></li><li><p>ex) BAC &amp; fire debris analysis</p></li></ul><p></p>
23
New cards

Isothermal Oven Temp Program

best for separation of highly similar compounds

  • unable to provide adequate separtion within reasonable run time for samps containing compounds of widely different boiling points

  • RESULT = decreasing resolution as a function of retention time

  • less common for most analytical apps


PROS = when analytes are chemically similar they have similar boiling points (can separate without changing temp)


24
New cards

Temperature Oven Temp Program

column temp raised in a linear manner during separation (can control how long oven stays at a temp and rate

  • increases analyte vapor pressure

  • changes retention factor (k)

  • it sharpens peaks

  • decreases retention times of late-eluting components

  • start with slower manner for temp raising and increase over time

  • more common for analytical apps


PROS = can separate compounds with different chemical properties (gives better resolution and sharp peak shapes), impacts retention factor (can measure column efficiency)


25
New cards

Types of Columns

  1. Packed Columns

  2. Capillary Columns


26
New cards

Packed Column

liquid stationary phase coated on particles

  • metal columns

  • original GC columns

  • greater sample capacity

  • replaced by capillary columns in 1980s (dont use anymore)


<p>liquid stationary phase coated on particles</p><ul><li><p>metal columns</p></li><li><p>original GC columns</p></li><li><p>greater sample capacity</p></li><li><p>replaced by capillary columns in 1980s (dont use anymore)</p></li></ul><p></p>
27
New cards

Capillary Columns

superior efficiency (>300,000 N)

  • narrower columns, thinner films, and longer columns

  • three types = WCOT, SCOT, PLOT


<p>superior efficiency (&gt;300,000 N)</p><ul><li><p>narrower columns, thinner films, and longer columns</p></li><li><p>three types = WCOT, SCOT, PLOT</p></li></ul><p></p>
28
New cards

Wall-Coated Open Tubular (WCOT)

narrow columns made of fused SiO2 coated on the outside with a polyamide polymer

  • quite common

  • entire coiled onto a cage for easier handling

  • internal diameter: 0.25, 0.32, or 0.53 nm (depends on application)


<p>narrow columns made of fused SiO2 coated on the outside with a polyamide polymer</p><ul><li><p>quite common</p></li><li><p>entire coiled onto a cage for easier handling</p></li><li><p>internal diameter: 0.25, 0.32, or 0.53 nm (depends on application)</p></li></ul><p></p>
29
New cards

Support-Coated Open Tubular (SCOT)

thin film (~30 uM) to support material coats column

  • liquid stationary phase applied to a solid support material

    • analyte retained by liquid film


<p>thin film (~30 uM) to support material coats column</p><ul><li><p><strong>liquid stationary phase </strong>applied to a solid <strong>support</strong> material</p><ul><li><p>analyte retained by liquid film</p></li></ul></li></ul><p></p>
30
New cards

Porous Layer Open Tubular (PLOT)

stationary phase is made up of very small particles coating inside walls of capillary

  • increase surface area = analyte has a lot of opportunties to react with stationary phase


<p>stationary phase is made up of very small particles coating inside walls of capillary</p><ul><li><p><strong>increase surface area</strong> = analyte has a lot of opportunties to react with stationary phase </p></li></ul><p></p>
31
New cards

Column Bleeding

when stationary phase starts to elute from the column

32
New cards

Important Attributes to Selecting Type of Column

  • low volatility

  • thermal stability → wide temp range, if it goes above = column bleeding

  • chemical inertness

  • like dissolves like


33
New cards

Thicker Stationary Film Thickness (greater than or equal to 1.0 um)

PROS

  • increased retention for volatiles → more opportunity for analyte to interact with stationary phase

  • increased capacity for GC-MS and GC-IR


CONS

  • less efficient

  • higher temperatures required — leads to noise

  • higher bleed (with higher temps because more liquid can leave column)

  • reduce resolution for late eluters (especially those with high boiling points)



34
New cards

Thinner Stationary Film Thickness (less than or equal to 0.5 um)

PROS

  • highest efficiency

  • lower elution temperature (less bleed)

  • fast analysis


CONS

  • low capacity → don’t have as many sites for analyte to interact with

    • peak fronting & column overload/saturation

  • limited trace analysis


**if you have low concentration of analyte, want to inject MORE to give better chance of interaction


35
New cards

Solvent Delay

used to avoid a solvent peak → cut the peak from our detection

  • initial time of your run that in a GC-MS your detector is off (anything reaching the source at that time isn’t ionized at detector)


36
New cards

Narrow Column Internal Diameter (0.15 mm)

better for fewer compounds

  • lead to higher hep pressures (gas needed at injection port to push gas/analyte inside the column


PROS = higher resolution

CONS = requires higher operating pressure, less column capacity


37
New cards

Wider Column Internal Diameter (0.50 mm)

better for complex mixtures

PROS = requires lower operating pressure, greater column capacity

CONS = lower resolution

38
New cards

Shorter Column Length (<15 m)

  • faster run times

  • better for chemically active samples

  • better with simple mixtures

  • **shouldn’t have reaction in GC, but maybe expose analyte to shorter run to an analyte that’s stubbornly stuck inside the column


39
New cards

Larger Column Length (>60 m)

  • slower run times

  • better for less active samples

  • better for complex (50-60 compounds) mixtures


40
New cards

Detectors

GOAL = rapid response to an analyte eluting from the column followed by a quick return to its original state in prep for the next compound

  • choice of detector depends on the nature of compounds being separated

  • each type has a different response to compounds

  • each type has different uses based on response and specific detector characteristics


41
New cards

IDEAL Detector Characteristcs

  • adequate sensitivity

  • good stability and reproducibility

  • similar response towards all analytes or at least highly predictable and selective response (selectivity)

  • linear response to solutes that extends over several orders of magnitude throughout overall concentration (dynamic range)

  • temp range from room temp to at least 400 degrees C

  • short response time independent of flow rate

  • high reliability and ease of use

  • non-destructive


42
New cards

5 Most Common GC Detectors

  1. Flame Ionization Detector (FID)

  2. Thermal Conductivity Detector (TCD)

  3. Electron-Capture Detector (ECD)

  4. Nitrogen-Phosphorous Detector (NPD)

  5. Mass Spectrometry Detector (MS)


43
New cards

Flame Ionization Detector (FID)

**most widely used and applicable detector

  • elute from the column directed to a small air-hydrogen flame → combustion leads to formation of ions and electrons (burn all organic components)

    • Carbon atoms (except CO) produce ions (CHO+) = how current is made

    • detection involves monitoring the current by collecting these charge carriers

  • USED FOR BLOOD ALCOHOL IN TOX LABS

  • Mass-sensitive rather than concentration-sensitive (flow rate insensitive)

    • concentration doesn’t matter cuz everything goes to flame

  • Make up gas = N2 → quickly sweeps effluent into flame & reduces band broadening

  • Hydrogen gas = combustion fuel (REQUIRED)


Advantages =

  • high sensitivity

  • high linear dynamic range (~10^7)

  • low detector noise because flame only sensitive with compounds containing C and H

  • quantitative analysis

  • simplicity + ease of operation

  • applications = cocaine


Limitations =

  • destructive (cant combine with another detector)

  • only applicable to compounds with carbon atoms = depends on carbon content

  • non-specific response

  • requires additional gases and flow controllers (H2 gas is dangerous)

  • applications = small analyte with halogens and amines DONT WORK


<p>**most widely used and applicable detector</p><ul><li><p>elute from the column directed to a small <strong>air-hydrogen flame</strong> → combustion leads to formation of ions and electrons (burn all organic components)</p><ul><li><p>Carbon atoms (except CO) produce ions (CHO+) = how current is made</p></li><li><p>detection involves monitoring the current by collecting these charge carriers</p></li></ul></li><li><p>USED FOR BLOOD ALCOHOL IN TOX LABS</p></li><li><p><strong>Mass-sensitive rather than concentration-sensitive</strong> (flow rate insensitive)</p><ul><li><p>concentration doesn’t matter cuz everything goes to flame</p></li></ul></li><li><p>Make up gas = N2 → quickly sweeps effluent into flame &amp; reduces band broadening</p></li><li><p>Hydrogen gas = combustion fuel (REQUIRED)</p></li></ul><p></p><p><strong>Advantages = </strong></p><ul><li><p>high sensitivity</p></li><li><p>high linear dynamic range (~10^7)</p></li><li><p>low detector noise because flame only sensitive with compounds containing C and H</p></li><li><p>quantitative analysis</p></li><li><p>simplicity + ease of operation</p></li><li><p>applications = cocaine</p></li></ul><p></p><p><strong>Limitations = </strong></p><ul><li><p>destructive (cant combine with another detector)</p></li><li><p>only applicable to compounds with carbon atoms = depends on carbon content</p></li><li><p>non-specific response</p></li><li><p>requires additional gases and flow controllers (H2 gas is dangerous)</p></li><li><p>applications = small analyte with halogens and amines DONT WORK</p></li></ul><p></p>
44
New cards

Thermal Conductivity Detector (TCD)

one of the earliest detectors = BASED ON DIFFERENCES IN THERMAL CONDUCTIVITY

  • sample conductivity compared to a reference conductivity from the carrier gas (He or H2)

    • any analyte mixed with helium lowers conductivity of the gas stream

    • detector measures change in voltage from filament (electrically heated tungsten-rhenium filament in temp-controlled cell)

  • when the analyte elutes, thermal conductivity decreases, the filament temperature increases, and the resistance changes (impacts voltage)


Advantages =

  • universal detector = all compounds change thermal conductivity

  • simplicity = no training required

  • large dynamic range (~10^5)

  • non-destructive detection (not in terms of sample, but of analyte)


Limitations =

  • low sensitivty

  • sensitivity affected by flow rate → inversely proportional to flow rate

  • works best with packed GC columns


<p>one of the earliest detectors = BASED ON DIFFERENCES IN THERMAL CONDUCTIVITY</p><ul><li><p>sample conductivity compared to a reference conductivity from the carrier gas (He or H2)</p><ul><li><p>any analyte mixed with helium lowers conductivity of the gas stream</p></li><li><p>detector measures change in voltage from filament (electrically heated <strong>tungsten-rhenium </strong>filament in temp-controlled cell)</p></li></ul></li><li><p><strong><em>when the analyte elutes, thermal conductivity decreases, the filament temperature increases, and the resistance changes (impacts voltage)</em></strong></p></li></ul><p></p><p><strong>Advantages = </strong></p><ul><li><p>universal detector = all compounds change thermal conductivity</p></li><li><p>simplicity = no training required</p></li><li><p>large dynamic range (~10^5)</p></li><li><p><strong>non-destructive detection </strong>(not in terms of sample, but of analyte)</p></li></ul><p></p><p><strong>Limitations = </strong></p><ul><li><p>low sensitivty</p></li><li><p>sensitivity affected by flow rate → inversely proportional to flow rate</p></li><li><p>works best with packed GC columns</p></li></ul><p></p>
45
New cards

Electron-Capture Detector (ECD)

most common in environmental applications due to halogen selectivity

  • eluents pass over a radioactive beta-emitter (63-Ni) that ionizes the sample through electron capture by electronegative analytes

    • without analyte, constant current passed between a pair of electrodes

    • current decreases in the presence of analytes that capture electrons

  • Selective for halogens, peroxides, quinones, organometallics, and nitro groups

    • relatively insenstive to hydrocarbons, alcohols, and ketones

  • Carrier gas = N2


Advantages =

  • selective for halogens (I, Br, Cl, F), nitro-, and sulfur-containing compounds

  • detection of polynuclear aromatic compounds, anhydrides, and conjugated carbonyl compounds


Limitations =

  • targeted applications because of non-universal detector

  • dynamic range only about 2 orders of magnitude

  • contains radioactive source


<p>most common in environmental applications due to halogen selectivity</p><ul><li><p>eluents pass over a <strong>radioactive beta-emitter (63-Ni)</strong> that ionizes the sample through electron capture by electronegative analytes</p><ul><li><p>without analyte, constant current passed between a pair of electrodes</p></li><li><p>current decreases in the presence of analytes that capture electrons</p></li></ul></li><li><p>Selective for <strong>halogens</strong>, peroxides, quinones, organometallics, and <strong>nitro groups</strong></p><ul><li><p>relatively insenstive to hydrocarbons, alcohols, and ketones</p></li></ul></li><li><p>Carrier gas = N2</p></li></ul><p></p><p><strong>Advantages = </strong></p><ul><li><p>selective for halogens (I, Br, Cl, F), nitro-, and sulfur-containing compounds</p></li><li><p>detection of polynuclear aromatic compounds, anhydrides, and conjugated carbonyl compounds</p></li></ul><p></p><p><strong>Limitations = </strong></p><ul><li><p>targeted applications because of non-universal detector</p></li><li><p>dynamic range only about 2 orders of magnitude</p></li><li><p>contains radioactive source</p></li></ul><p></p>
46
New cards

Nitrogen-Phosphorus Detector (NPD)

similar principles as FID, but specific to nitrogen and phosphorus compounds

  • type of thermic ionic detector → rubidium sulfate (Rb2SO4) containing glass/ceramic bead greatly enhances the formation of ions from nitrogen and phosphorus containing compounds

  • about 500-fold more sensitive than an FID in detecting phosphorus-containing compounds & 50-fold more sensitive to nitrogen-containing compounds


Advantages =

  • increased sensitivity for nitrogen & phosphorus containing compounds

  • quantitative analysis with reference standards

  • works well for ketamine!


Disadvantages =

  • destructive

  • very sensitive to hydrocarbon impurities in carrier or make up gas = increased background noise

  • required additional gases and flow controllers

  • bead must be replaced over time because it gets eaten overtime

  • CANNOT USE deuterated internal standards because they will co-elute


<p>similar principles as FID, but specific to nitrogen and phosphorus compounds</p><ul><li><p>type of thermic ionic detector → <strong>rubidium sulfate (Rb2SO4) containing glass/ceramic bead</strong> greatly enhances the formation of ions from nitrogen and phosphorus containing compounds</p></li><li><p>about 500-fold more sensitive than an FID in detecting phosphorus-containing compounds &amp; 50-fold more sensitive to nitrogen-containing compounds</p></li></ul><p></p><p><strong>Advantages = </strong></p><ul><li><p>increased sensitivity for nitrogen &amp; phosphorus containing compounds</p></li><li><p>quantitative analysis with reference standards</p></li><li><p>works well for ketamine!</p></li></ul><p></p><p><strong>Disadvantages = </strong></p><ul><li><p>destructive</p></li><li><p>very sensitive to hydrocarbon impurities in carrier or make up gas = increased background noise</p></li><li><p>required additional gases and flow controllers</p></li><li><p>bead must be replaced over time because it gets eaten overtime</p></li><li><p>CANNOT USE deuterated internal standards because they will co-elute</p></li></ul><p></p>
47
New cards

Mass Spectrometry Detector (MS)

the most powerful & robust GC detector

  • based on measuring mass-to-charge ratio of ions generated from effluent

  • often combined with quadrupole mass analyzers

  • capability to identify analytes based on unique EI mass spectra based on unique EI mass spectra

  • qualitative & quantitative analyses


<p>the most powerful &amp; robust GC detector</p><ul><li><p>based on measuring mass-to-charge ratio of ions generated from effluent</p></li><li><p>often combined with quadrupole mass analyzers</p></li><li><p>capability to identify analytes based on unique EI mass spectra based on unique EI mass spectra</p></li><li><p>qualitative &amp; quantitative analyses</p></li></ul><p></p>
48
New cards

Total Ion Chromatogram (TIC)

sum of all ion abundances as a function of time (all ions being fragmented in instrument)

  • all peaks corresponding to ions in this range

  • click on peak = get mass spectrum (correspond to ions in smaller range)


49
New cards

Extracted Ion Chromatogram (EIC)

selection of specific fragments (m/z) from the TIC

  • “I want to see a chromatogram for these ___ lines from TIC)”

    • ex) have a few for ADB-BUTINACA and its internal standard, can select (examples) 181, 182, 183 => EIC show spectrum with only these three ions

  • Generated FROM mass spectrum to track that ion throughout the run (shows elution of something specific in MS)


50
New cards

SIM Chromatogram

TIC corresponding to only a few fragments monitored throughout the experiment

  • same as TIC but for selected peaks/few fragments

  • EIC chromatogram of MS from this get you back the SIM chromatogram

  • LOOK AT ION RATIOS


51
New cards

Criteria looked at for GC-MS

  • retention time (2%)

  • ions → diagnostic ions (most intense/stable ions)

    • molecular ion best to look for

  • ion ratios (± 20%)

    • some exceptions to rule

  • S/N (>3 for ID)

  • library searches (only in full scan)