what are the two ways to capture a digital radiograph

computed radiograph and digital radiograph

2
New cards

when was the first clinical scan

1971

3
New cards

when was the first commercial storage phosphorus system

Early 1980s

4
New cards

when was the first large-area panels for radiography (flat panel DR)

late 1990s

5
New cards

what is the 1st step of CR

expose the plate. storage phosphor plate(BaFBr:Eu) in a cassette. electrons are trapped: a hidden latent image

6
New cards

what is the 2nd step of CR

Carry to the reader. the cassette goes to a separate CR reader, which removes the plate

7
New cards

what is the 3rd step of CR

Laser readout. a red laser frees the trapped electrons; the plate glows violet (about 400nm). a PMT detects it

8
New cards

what is the 4th step of CR

Digitize. the PMT signal is digitized (ADC) and sent to the workstation as an image

9
New cards

what is the 5th step of CR

Erase and reuse, bright white light clears the plate, which goes back in the cassette for the next exam.

10
New cards

what is the 1st step of DR

expose the detector. fixed in the table or wall stand, or a cassette size wireless panel

11
New cards

what is the 2nd step of DR

Xrays to charge. indirect: scintillator light to photodiodes. direct: a-Se turns xrays straight to charge

12
New cards

what is the 3rd step of DR

read out in place. the TFT array reads the charge row by row inside the detector

13
New cards

what is the 4th step of DR

image at the console. digitalized and displayed within seconds. No cassette to carry or process.

14
New cards

indirect conversion goes to what two DR detectors

TFT flar panel (Csl:Ti or Gd202S:Tb) and CCD or CMOS

15
New cards

direct conversion goes to which DR detector

TFT flat panel a-Se

16
New cards

what are the parts of a Csl indirect flat-panel detector

  1. protective front cover

  2. reflector +moisture seal

  3. Csl:TI scintillator- needle crystals

  4. a-Si:H photodiode- one per DEL

  5. TFT switch- one per DEL

  6. glass substrate

  7. edge electronics


17
New cards

how does a Csl indirect detector work

xrays absorbed in a Csl needle

Scintillation xray to light

needles guide the light down

light to stored charge

row by row readout

amplify to multiplex to digitize

18
New cards

what are the parts of a GOS indirect flat-panel detector

protective front cover

support + reflective layer

Gd202S: Tb phosphor (GOS, “Gadox”)

optical coupling

a-Si:H photodiode- one per DEL

TFT switch- one per DEL glass substrate

edge electronics

19
New cards

How does a GOS indirect detector work

xray absorbed in the phosphor grain

scintillation: xray to light

light scatters in every direction

light to stored charge

row by row readout

amplfy to multiplex to digitize

20
New cards

what are the parts of a CCD-based indirect detector

protective front coverprotective front cover

scintillator screen

mirror (angled 45 degrees)

lens

CCD chip (charge-coupled device)

thermoelectric cooler

lead sheilding

camera electronics

21
New cards

what are the steps of a CCD based indirect detector work

xray to light

mirror turn the light to 90 degrees

lens focuses and shrinks

light to stored electrons

parallel shift

serial shift to amplifier to ADC

software corrections

22
New cards

The parts of a direct conversion a-Se detector

protective front cover

top(bias) electrode

high voltage supply

blocking layers

amorphous selenium photoconductor

pixel (charge-collection) electrode

storage capacitor (Cs)

TFT switch

glass substrate

23
New cards

how does a direct conversion a-Se detector work

xray absorbed in the selenium

photoelectrons to electron-hole pairs

field pulls the charges apart

charge stored in the DEL

row by row readout

amplify to multiplex to digitize

24
New cards

general indirect conversion process

xrays are absorbed in the scintillator

it glows: visible light spreads toward the array

photodiodes turn the light into charge; TFTs hold it

electronics read each DEL and build the image

25
New cards

Gadox: turbid phosphor causes what

high light spread: light bounces off the grains and spreads sideways into neighboring DELs

26
New cards

True or False: Csl:Tl results in low light spread because the needles guide the light straight down by total internal reflection

True

27
New cards

SNR detector performance

Indirect CCD- low to moderate

indirect TFT -very high

Direct TFT- moderate

28
New cards

MTF detector performance

Indirect CCD- low to moderate

Indirect TFT- high

Direct TFT- very high

29
New cards

CNR detector performance

indirect CCD- low to moderate

indirect TFT- very high

Direct TFT- moderate

30
New cards

DQE detector performance

indirect CCD- low

indirect TFT- very high

direct TFT- moderate

31
New cards

what is a DEL

a detector element: one tiny square detector. thousands of rows and columns of them make up a flat panel

32
New cards

what is inside a DEL

a sensing area (photodiode or pixel electrode in a-Se panels). it reflects the Xray exposure that reached that small square

33
New cards

what is pixel pitch

center to center spacing of DELs. about 100-200um in general radiography. About 50-100um in mammography

34
New cards

True or False: a DEL can record detail smaller than itself

False

35
New cards

what detectors does general radiography use

Indirect Csl and indirect GOS

36
New cards

what detectors does mammography use

Direct a-Se, Indirect Csl and photon counting

37
New cards

what detectors does fluoroscopy, angiography use

Indirect Csl, image intensifier + camera, and Direct a-Se

38
New cards

what detector does stereotactic breast biopsy use

CCD

39
New cards

what detector does slot-scan systems use

CCD

40
New cards

Large field, wide kVp range

indirect Csl (or GOS to save cost)

41
New cards

moving images at low does

indirect Csl

42
New cards

Lowest kVp, finest detail

direct a-Se

43
New cards

smaller field or scanning slot

CCD

44
New cards

what is DR

a system whose detector coverts xrays to charge and reads the image out electronically, so it appears at the console within seconds cassette to process

45
New cards

what is indirect conversion

Two steps: a scintillator turns xrays into visible light then a light sensor (photodiode, CCD, or CMOS) turns the light into charge

46
New cards

what is Csl:Tl (cesium iodine)

a scintillator grown as needle like columns that guide light straight down by total internal reflection: low light spread high DQE

47
New cards

what is amorphous selenium (a-Se)

the photoconductor in direct flat panels. under a high voltage field it turns xrays into electon hole pairs; mainly used in mammography

48
New cards

what is charge-coupled device (CCD)

a small silicon camera chip. charge is passed pixel to pixel (bucket brigade) to one output amplifier, then digitized

49
New cards

what is computed radiography (CR)

a cassette based system: xrays trap electrons in a storage phosphor plate, and a separate reader releases them with a red laser to form the image

50
New cards

what is direct conversion

one step: a photoconductor (amorphous selenium) turns xrays straight into electrical charge with no light produced

51
New cards

what is Gd2O2:Tb (GOS, Gadox)

a powder phosphor in a binder, light scatters among the grains and spreads sideways. lower lost, moisture-stable

52
New cards

what is thin-film transistor (TFT)

the switch in each DEL. it stays OFF during exposure and turns ON one row at a time so the stored charge can be read

53
New cards

what is demagnification

shrinking the large screen image into a small CCD with a lens or fiber-optic taper. more shrinking means less light is collected

54
New cards

what is photostimulable phosphor

the CR plate material (BaFBr:Eu). it stores a latent image as trapped electrons and glows violet (about 400nm) when a laser frees them

55
New cards

what is a scintillator

a layer that glows when it absorbs xrays. in DR it is Csl:TI or Gadox both emit green light

56
New cards

what is amorphous silicon photodiode

the a-Si:H light sensor in each DEL of an indirect flat panel. it turns scintillator light into charge and stores until readout

57
New cards

what is flat-panel detector (FPD)

a detector built on a TFT array on glass: indirect or direct, CCD and CMOS systems are not flat panels

58
New cards

what is fiber-optic taper

a bundle of glass fibers that narrows from screen size to chip size. collects more light than a lens; used for small fields

59
New cards

Detector element (DEL)

one tiny square detector in the panels grid. it holds a sensing area, a TFT and charge storage, and it becomes one pixel in the image

60
New cards

Fill factor

the fraction of each DEL that can sense signal. the TFT and wiring take the rest; smaller DELs have a lower fill factor

61
New cards

True or False: smaller DELs have a lower fill factor

True

62
New cards

Analog to digital converter

turns each DELs amplified charge into a number, one value per pixel

63
New cards

spatial resolution

the ability to show small, closely spaced objects as seperate. limited by DEL size and by light or charge spread

64
New cards

SNR and CNR

signal to noise and contrast to noise ratios. at the same dose both rise when a detector uses more of an incoming xrays

65
New cards

storage capacitor

holds each DELs charge until readout. in indirect panels the photodiode itself acts as the capacitor; direct panels have a separate one

66
New cards

Nyquist frequency

the finest detail a detector can record: 1/(2 x pitch)

67
New cards

contrast resolution

the ability to show small difference in soft tissues. xray attenuation, such as soft tissues. limited mainly by image noise

68
New cards

DQE (detective quantum efficiency)

how efficiently a detector turns incoming xrays into image signal to noise. higher DQE gives the same image quality at lower dose

69
New cards

matrix

number of DELs in rows and columns

70
New cards

gate line and data line

gate lines (rows) switch the TFTs on; data lines (columns) carry each DELs charge to a charge amplifier at the panels edge

71
New cards

light spread

sideways travel of scintillator light before it reaches the sensor. more spread puts signal into neighboring DELs and blurs the image

72
New cards

MTF (modulation transfer function)

how well a detector passes detail at each spatial frequency. higher MTF means sharper images

73
New cards

secondary quantum sink

a stage where signal carries drop to about one per xray, adding noise. in lens-coupled CCDs, most light is lost at the lens

74
New cards

What are the five stages that carry the signal from the detector to the display

  1. detector capture

  2. Segmentation and Histogram

  3. Automatic rescaling

  4. VOI lookup table

  5. GSDF+ display


75
New cards

what is Grayscale Standard Display Function (GSDF)

the DICOM standard for displaying non-color medical images on a medical grade monitor

76
New cards

only the … branch reports the exposure that rescaling hides

EI/DI

77
New cards

what does the “four orders of magnitude” mean

This is an exposure range, not a bit depth or a number of gray shades

78
New cards

your DD can image … of exposure

“four orders of magnitude”

79
New cards

The “four orders of magnitude” exposure ratio

10,000:1

80
New cards

what does exposure dynamic range measure

max+min usable detector exposure

81
New cards

what does bit depth measure

number of discrete pixel values

82
New cards

what does subject (anatomic) range measure

intensity range inside one patient

83
New cards

what does display range measure

gray levels a monitor renders

84
New cards

what is the typical value for exposure dynamic range

about 10^4 (10,000:1)

85
New cards

what is the typical value of bit depth

10-16 bit (1,024-65,536 values)

86
New cards

what is the typical value of subject (anatomic) range

about 100:1 (6cm breast, 20 kev)

87
New cards

what is the typical value of display range

8 bit= 256; 10 bit=1,024

88
New cards

what is the relation to the four decades of exposure dynamic range

this is the four orders of magnitude

89
New cards

what is the relation to the four decades of bit depth

must resolve all four decades

90
New cards

what is the relation to the four decades of subject (anatomic) range

uses only part of the four decades

91
New cards

what is the relation to the four decades of display range

far smaller; needs rescaling and LUTs

92
New cards

does raw detector data look like a finished radiograph

no

93
New cards

what does the raw detector image appear like

open beam is white; nearly all anatomy is crushed into near black

94
New cards

what is the For-presentation image

real radiograph after log, segmentation, rescaling, equalization, and LUT

95
New cards

in screen film what is the brightness set by

detector exposure

96
New cards

in screen film, what does underexposure show as

light film

97
New cards

in screen film, what does overexposure show as

dark film

98
New cards

in screen film, exposure indicator shows as

optical density

99
New cards

In digital brightness is set by

postprocessing

100
New cards

in digital underexposure shows as

more quantum noise