what are the two ways to capture a digital radiograph
computed radiograph and digital radiograph
when was the first clinical scan
1971
when was the first commercial storage phosphorus system
Early 1980s
when was the first large-area panels for radiography (flat panel DR)
late 1990s
what is the 1st step of CR
expose the plate. storage phosphor plate(BaFBr:Eu) in a cassette. electrons are trapped: a hidden latent image
what is the 2nd step of CR
Carry to the reader. the cassette goes to a separate CR reader, which removes the plate
what is the 3rd step of CR
Laser readout. a red laser frees the trapped electrons; the plate glows violet (about 400nm). a PMT detects it
what is the 4th step of CR
Digitize. the PMT signal is digitized (ADC) and sent to the workstation as an image
what is the 5th step of CR
Erase and reuse, bright white light clears the plate, which goes back in the cassette for the next exam.
what is the 1st step of DR
expose the detector. fixed in the table or wall stand, or a cassette size wireless panel
what is the 2nd step of DR
Xrays to charge. indirect: scintillator light to photodiodes. direct: a-Se turns xrays straight to charge
what is the 3rd step of DR
read out in place. the TFT array reads the charge row by row inside the detector
what is the 4th step of DR
image at the console. digitalized and displayed within seconds. No cassette to carry or process.
indirect conversion goes to what two DR detectors
TFT flar panel (Csl:Ti or Gd202S:Tb) and CCD or CMOS
direct conversion goes to which DR detector
TFT flat panel a-Se
what are the parts of a Csl indirect flat-panel detector
protective front cover
reflector +moisture seal
Csl:TI scintillator- needle crystals
a-Si:H photodiode- one per DEL
TFT switch- one per DEL
glass substrate
edge electronics
how does a Csl indirect detector work
xrays absorbed in a Csl needle
Scintillation xray to light
needles guide the light down
light to stored charge
row by row readout
amplify to multiplex to digitize
what are the parts of a GOS indirect flat-panel detector
protective front cover
support + reflective layer
Gd202S: Tb phosphor (GOS, “Gadox”)
optical coupling
a-Si:H photodiode- one per DEL
TFT switch- one per DEL glass substrate
edge electronics
How does a GOS indirect detector work
xray absorbed in the phosphor grain
scintillation: xray to light
light scatters in every direction
light to stored charge
row by row readout
amplfy to multiplex to digitize
what are the parts of a CCD-based indirect detector
protective front coverprotective front cover
scintillator screen
mirror (angled 45 degrees)
lens
CCD chip (charge-coupled device)
thermoelectric cooler
lead sheilding
camera electronics
what are the steps of a CCD based indirect detector work
xray to light
mirror turn the light to 90 degrees
lens focuses and shrinks
light to stored electrons
parallel shift
serial shift to amplifier to ADC
software corrections
The parts of a direct conversion a-Se detector
protective front cover
top(bias) electrode
high voltage supply
blocking layers
amorphous selenium photoconductor
pixel (charge-collection) electrode
storage capacitor (Cs)
TFT switch
glass substrate
how does a direct conversion a-Se detector work
xray absorbed in the selenium
photoelectrons to electron-hole pairs
field pulls the charges apart
charge stored in the DEL
row by row readout
amplify to multiplex to digitize
general indirect conversion process
xrays are absorbed in the scintillator
it glows: visible light spreads toward the array
photodiodes turn the light into charge; TFTs hold it
electronics read each DEL and build the image
Gadox: turbid phosphor causes what
high light spread: light bounces off the grains and spreads sideways into neighboring DELs
True or False: Csl:Tl results in low light spread because the needles guide the light straight down by total internal reflection
True
SNR detector performance
Indirect CCD- low to moderate
indirect TFT -very high
Direct TFT- moderate
MTF detector performance
Indirect CCD- low to moderate
Indirect TFT- high
Direct TFT- very high
CNR detector performance
indirect CCD- low to moderate
indirect TFT- very high
Direct TFT- moderate
DQE detector performance
indirect CCD- low
indirect TFT- very high
direct TFT- moderate
what is a DEL
a detector element: one tiny square detector. thousands of rows and columns of them make up a flat panel
what is inside a DEL
a sensing area (photodiode or pixel electrode in a-Se panels). it reflects the Xray exposure that reached that small square
what is pixel pitch
center to center spacing of DELs. about 100-200um in general radiography. About 50-100um in mammography
True or False: a DEL can record detail smaller than itself
False
what detectors does general radiography use
Indirect Csl and indirect GOS
what detectors does mammography use
Direct a-Se, Indirect Csl and photon counting
what detectors does fluoroscopy, angiography use
Indirect Csl, image intensifier + camera, and Direct a-Se
what detector does stereotactic breast biopsy use
CCD
what detector does slot-scan systems use
CCD
Large field, wide kVp range
indirect Csl (or GOS to save cost)
moving images at low does
indirect Csl
Lowest kVp, finest detail
direct a-Se
smaller field or scanning slot
CCD
what is DR
a system whose detector coverts xrays to charge and reads the image out electronically, so it appears at the console within seconds cassette to process
what is indirect conversion
Two steps: a scintillator turns xrays into visible light then a light sensor (photodiode, CCD, or CMOS) turns the light into charge
what is Csl:Tl (cesium iodine)
a scintillator grown as needle like columns that guide light straight down by total internal reflection: low light spread high DQE
what is amorphous selenium (a-Se)
the photoconductor in direct flat panels. under a high voltage field it turns xrays into electon hole pairs; mainly used in mammography
what is charge-coupled device (CCD)
a small silicon camera chip. charge is passed pixel to pixel (bucket brigade) to one output amplifier, then digitized
what is computed radiography (CR)
a cassette based system: xrays trap electrons in a storage phosphor plate, and a separate reader releases them with a red laser to form the image
what is direct conversion
one step: a photoconductor (amorphous selenium) turns xrays straight into electrical charge with no light produced
what is Gd2O2:Tb (GOS, Gadox)
a powder phosphor in a binder, light scatters among the grains and spreads sideways. lower lost, moisture-stable
what is thin-film transistor (TFT)
the switch in each DEL. it stays OFF during exposure and turns ON one row at a time so the stored charge can be read
what is demagnification
shrinking the large screen image into a small CCD with a lens or fiber-optic taper. more shrinking means less light is collected
what is photostimulable phosphor
the CR plate material (BaFBr:Eu). it stores a latent image as trapped electrons and glows violet (about 400nm) when a laser frees them
what is a scintillator
a layer that glows when it absorbs xrays. in DR it is Csl:TI or Gadox both emit green light
what is amorphous silicon photodiode
the a-Si:H light sensor in each DEL of an indirect flat panel. it turns scintillator light into charge and stores until readout
what is flat-panel detector (FPD)
a detector built on a TFT array on glass: indirect or direct, CCD and CMOS systems are not flat panels
what is fiber-optic taper
a bundle of glass fibers that narrows from screen size to chip size. collects more light than a lens; used for small fields
Detector element (DEL)
one tiny square detector in the panels grid. it holds a sensing area, a TFT and charge storage, and it becomes one pixel in the image
Fill factor
the fraction of each DEL that can sense signal. the TFT and wiring take the rest; smaller DELs have a lower fill factor
True or False: smaller DELs have a lower fill factor
True
Analog to digital converter
turns each DELs amplified charge into a number, one value per pixel
spatial resolution
the ability to show small, closely spaced objects as seperate. limited by DEL size and by light or charge spread
SNR and CNR
signal to noise and contrast to noise ratios. at the same dose both rise when a detector uses more of an incoming xrays
storage capacitor
holds each DELs charge until readout. in indirect panels the photodiode itself acts as the capacitor; direct panels have a separate one
Nyquist frequency
the finest detail a detector can record: 1/(2 x pitch)
contrast resolution
the ability to show small difference in soft tissues. xray attenuation, such as soft tissues. limited mainly by image noise
DQE (detective quantum efficiency)
how efficiently a detector turns incoming xrays into image signal to noise. higher DQE gives the same image quality at lower dose
matrix
number of DELs in rows and columns
gate line and data line
gate lines (rows) switch the TFTs on; data lines (columns) carry each DELs charge to a charge amplifier at the panels edge
light spread
sideways travel of scintillator light before it reaches the sensor. more spread puts signal into neighboring DELs and blurs the image
MTF (modulation transfer function)
how well a detector passes detail at each spatial frequency. higher MTF means sharper images
secondary quantum sink
a stage where signal carries drop to about one per xray, adding noise. in lens-coupled CCDs, most light is lost at the lens
What are the five stages that carry the signal from the detector to the display
detector capture
Segmentation and Histogram
Automatic rescaling
VOI lookup table
GSDF+ display
what is Grayscale Standard Display Function (GSDF)
the DICOM standard for displaying non-color medical images on a medical grade monitor
only the … branch reports the exposure that rescaling hides
EI/DI
what does the “four orders of magnitude” mean
This is an exposure range, not a bit depth or a number of gray shades
your DD can image … of exposure
“four orders of magnitude”
The “four orders of magnitude” exposure ratio
10,000:1
what does exposure dynamic range measure
max+min usable detector exposure
what does bit depth measure
number of discrete pixel values
what does subject (anatomic) range measure
intensity range inside one patient
what does display range measure
gray levels a monitor renders
what is the typical value for exposure dynamic range
about 10^4 (10,000:1)
what is the typical value of bit depth
10-16 bit (1,024-65,536 values)
what is the typical value of subject (anatomic) range
about 100:1 (6cm breast, 20 kev)
what is the typical value of display range
8 bit= 256; 10 bit=1,024
what is the relation to the four decades of exposure dynamic range
this is the four orders of magnitude
what is the relation to the four decades of bit depth
must resolve all four decades
what is the relation to the four decades of subject (anatomic) range
uses only part of the four decades
what is the relation to the four decades of display range
far smaller; needs rescaling and LUTs
does raw detector data look like a finished radiograph
no
what does the raw detector image appear like
open beam is white; nearly all anatomy is crushed into near black
what is the For-presentation image
real radiograph after log, segmentation, rescaling, equalization, and LUT
in screen film what is the brightness set by
detector exposure
in screen film, what does underexposure show as
light film
in screen film, what does overexposure show as
dark film
in screen film, exposure indicator shows as
optical density
In digital brightness is set by
postprocessing
in digital underexposure shows as
more quantum noise