Radiation Safety Features of Radiographic Equipment, Devices, and Accessories
All necessary measures to ensure that radiographic equipment operates safely to ensure radiation protection for
Patients
All personnel
Every diagnostic imaging system must have a
Protective tube housing
Correctly functioning control panel
A radiographic exam table and other devices and accessories must be designed to reduce patient radiation dose
Diagnostic Protective Tube Housing
Must protect the patient and personnel from leakage radiation by restricting the emission of x-rays to the area of the primary beam
Must be lead lined so that the radiation leakage through the portion of the housing away from the useful beam, measured at 1 m (3 ft) does not exceed 1mGy an/hr (100 mR/hr) when the tube is operated at its highest voltage at the highest current that allows for continuous operation
The housing also serves as a shield against electric shock from high voltage entering the x-ray tube and facilitates cooling of the tube
Control Panel
Where you select technical factors; mA, mAs, kVp are selected
Must be located behind a suitable barrier that has a radiation absorbent window which protects the RT and allows viewing of patient
Must indicate the conditions of exposure and when the tube is energized
X-ray control panels could be touch screen; they provide audible tone and visible light when exposure begins/ends
Examination Table
Must be strong enough to support pts whose weight exceeds 300 lbs
Must be uniform in thickness (may be curved or flat)
Must be a floating table top
May have x-ray from underneath (fluoro)
Must be as radiolucent as possible
Carbon fiber materials used
SID Indicator
Equipment must have indicator that will accurately measure the distance from the anode focal spot to the image receptor
Examples: Tape measure, lasers, digital readout
Must be accurate to within 2% of the indicated SID
Distance and centering indicators must be accurate within 2% and 1% of the SID distance
X-Ray Beam Limitation Devices
Collimation;
Reduces the quantity of tissue exposed to radiation
Reduce dose to patient
Reduces the amount of scatter production
Improves the overall quality of the image
Light Localizing Variable Aperture Rectangular Collimators
Most versatile, used most often, most accurate collimator
Close to exit window of x-ray tube, permanently mounted
Contains
1st set (upper) shutters: Mounted as close as possible to the tube window to reduce off focus or stem radiation (rad emitted from parts of tube other than focal spot)
2nd set (lower) shutters: Located below level of light source and mirror; consists of two lead plates perpendicular to each other; further restricts beam to area of interest
A light source to illuminate the x-ray field and permit centering
A mirror to deflect the light beam toward the patient
Skin Sparing
Skin surface should be at least 15 cm below the collimator
Luminance (brightness of a surface)
Quantifies the intensity of a light source
The localizing light beam should adequately outline the margins of the radiographic beam on the area of interest on all patients
Primary unit is the candela per square meter (nit)
Taken at 40” a light meter reading will be at least 15 foot candles
A foot candle is about 10.76 candela per square meter (nit)
Coincidence-Between Radiographic Beam and Localizing Light Beam
When using a rectangular collimator, good coincidence between radiation beam and localizing light beam is essential to eliminate collimator cut off of the body structures being irradiated
Alignment and congruence: (Respectively) The sum of the cross table and along the table alignment differences and the sum of the length and width dimensions of the radiographic and light beams must correspond to within 2% of SID
*Coincide: correspond in position; meet or intersect
Positive Beam Limitation
Adjusted to the selected IR size. Border should be visible on all four sides when field is limited to clinical areas of interest; PBL had override button/key
Radiographer must ensure basic principle of radiation protection: collimating the radiographic beam so that it is no larger than the IR
2% of SID with PBL (some states 3%)
Aperture Diaphragm
Simplest Beam Limitation Device
Flat piece of lead with a shaped hold cut in its center
Dimensions of the hold determine size and shape of beam
Openings can be rectangular, square or round
Placed directly below the window of the x-ray tube
Cones
Mostly used for examination of specific areas
Head Sinuses Vertebral column
Circular metal tubes that attach to x-ray tube housing or variable rectangular collimator
Limits the x-ray to a predetermined shape and size
Construction is simple
Primarily the lower rim of cone that governs beam limitation
Total Filtration
Reduces exposures to patient skin
Absorbs most of the lower energy photons which hardens the beam
Therefore, reducing the overall intensity of the beam
Two types make up “Total Filtration”
Inherent Filtration
Glass envelope, insulating oil, and the glass window 0.5 mm thickness of aluminum
The light localizing variable aperture rectangular collimator provides an additional 1mm aluminum equivalent
Added Filtration
Consists of sheets of aluminum or equivalent
Located outside the glass window of the tube housing, above the collimator shutters
With Image Intensified Fluoro - 3.0 mm AL is preferred (patient dose decreased by 1/4)
Patient dose is decreased by 25% during fluoro when Al filtration is increased from 1 mm to 3mm
Requirement of Total Filtrations
2.5 mm aluminum equivalent for fixed x-ray units operating above 70 kVp is the regulatory standard
Manufacturers only need to add a 1.0 mm Al eq filter between the tube housing and the collimator in order to meet the requirement
Stationary (Fixed) Radiographic Equipment Tube Potential
Minimum Total Filtration Required
Above 70 kVp - 2.5 mm Al Eq
50-70 kVp - 1.5 mm Al Eq
Below 50 - 0.5. mm Al Eq
Mobile Diagnostic Units and Fluoro Equipment
Mobile diagnostic units and fluoro equipment require a minimum of 2.5 mm Al Eq total permanent filtration
Filtration for General Diagnostic Radiology
Aluminum is the metal most widely selected as a filter material because it effectively removes low energy (soft) x-rays from a polyenergetic (heterogenous) x-ray beam without severely decreasing the x-ray beam intensity
A diagnostic x ray beam must always ben adequately filtered
Half value layer (HVL) of the beam must be measured to verity this. A radiologic physicist should obtain this measurement at least once a year and also after and x ray tube is replaced or repairs have been made of the diagnostic x ray tube housing or collimation
HVL is the thickness of an absorber required to decrease the intensity of the primary beam to 50% of its initial value
HVL is expressed in millimeters of aluminum
HVL is a measure of beam quality or effective energy of the x ray beam; a certain minimal HVL is required at a given kVp
Compensating Filters
May be constructed of aluminum, lead acrylic or other suitable material
Partially attenuated x rays that are directed toward thinner or less dense areas
Provides uniform density of part
Ex: Trough (used in dedicated chest units; think in the center thick laterally) or wedge (Used in foot xrays; thickest part towards toes)
Exposure Reproducibility
mA reproducibility of +/- 5%
If any or all technique factors are changed and then returned to previous value, precisely the same radiation exposure should occur of the exposure should be reproducible
Exposure Linearity
mA linearity +/- 10% between adjacent stations and +/- 10% overall
Ability of unit to produce constant radiation output for multiple combinations of mA and exposure time
The ratio of the difference in mR/mAs values between two successive generator setting to the sum of the mR/mAs values. It must be less than 0.1
Inspected annually or when a significant change or repair in operating console or high voltage generator components
kVp Accuracy
Must be consistent within +/- 4% (inspected annually or when a significant change in high voltage generator components)
Grids
Construction: Made of parallel radiopaque strips alternated with low attenuation strips of aluminum
Removes scattered x-ray photons that emerge from the patient before they reach the IR
Used when part is greater than 10cm thick
Technical factor requirements
Bucky factor: Required increase in exposure
Dose increases as a grid is used
Pt dose also increases as grid ratio increases
Mobile Radiography
Patient Protection
Same design and recommendations as Fixed Rad. Equipment
Use gonadal shielding
Vacate area during exposure
Announce when exposure is made
Provide lead aprons for others
Source-to-Skin Distance (SSD) for patient
Must be at least 12 in. (30 cm); Should not be < 15 in.
Use low ratio grids
Use collimation
Use high kVp
Avoid repeats
Radiation Safety Features of Digital Imaging Equipment, devices and accessories
Digital Imaging
Use of the computer in almost all imaging modalities:
Conventional radiography: Analog image
Process of producing the conventional radiographic image
Disadvantages to the use of this technology?
Waiting time for processing
Storage
Manually retrieving radiographs was time consuming
Human error with film jackets and being misfiled or misplaced
Digital radiography (DR)
Process of producing a digital radiographic image:
Components of the digital image
Resolution (detail) of the digital radiographic image
Composition and function of digital radiography image receptors
Function of charge-coupled devices (CCDs)
Access of DR images
Retake Rates in Digital Radiography
DR eliminates the need for almost all retakes required because of improper technical selection because image contrast and overall brightness may be manipulated after image acquisition.
Retake rates for mis-positioning will not be affected.
Mispositioning retakes should be monitored by an independent quality control technologist at a separate monitor, or a quality control system should be used, whereby the number of images per examination is compared with the number ordered for each technologist.
Fluoroscopy
Dynamic, active motion for select anatomic structures that produces a real-time image of those structures on a television monitor that works with an imaging intensification system.
Fluoroscopic Imaging Systems
X-ray tube under table and image intensifier and spot film system mounted on a linked system over the patient (C-arm like)
Tube above the table: image receptor beneath table
Remote control system; tech can fluoro from behind window
Benefits of Modern Image Intensification
Fluoroscopy overall brightness is increased~ 10,000 times the brightness of the discontinued non-image intensifier system
Time saver for radiologist-allows them to use normal light as opposed to having to sit in a dark room for 30 mins to adjust their eyes before preforming any fluoro procedure
Reduces pt dose because it requires less mA (1.5-2 mA)
Exposure Rate Limitation
Limits on entrance skin exposure
Intensity should not exceed 20 mGy/min for each mA of operation at 80 kVp
Image intensified total limit of Intensity - limited the table top exposure rate of fluoro equipment for routine procedures to 100 mGy/min (10 R/min) when no HLC (high level control) was present. If HLC is present, the max TT intensity allowed is 200 mGy/min
High Dose level controls
Allows unlimited exposures
Commonly called HLC (High Level Control) fluoroscopy
Operating exposure rates are higher
Provides for visualization of smaller and lower contrast objects
Used for interventional procedures
Requires continuous, positive pressure manual operation, continuous audible signal (reminds everyone HLC is in use)
Pulsed/Intermittent Fluoroscopy
Allow pulsed fluoro (short burst) rather than continuous exposure; patient dose is decreased
Digital form of fluoro
Last image hold: stores the image from the last time fluoro button was depressed
Exposure Control
Exposure Switch should be the dead-man-type (if the operator drops dead, the exposure will terminate unless he or she falls on the switch); this applies to the foot pedal or pressure switch on the image intensified tower
Exposure Factors
↑ kVp, ↓ mA reduces pt dose
Control Grids
Patient dose ↑ when fluoro grid is used.
Focuses electron beam into the output fluorescent screen by external coils
Bucky-slot Cover
The opening should automatically be covered with at least .25 mm lead equivalent; must be part of the equipment.
Magnification (electron optical magnification)
During Mag. Fluoro. (9 in to 6 in. mode) , mA automatically increases and patient dose ↑ and personnel dose ↑.
Beam Restriction
Collimate to area of interest, it reduces the integral dose (product of dose & volume of tissue
An unexposed border must be visible on the TV monitor when the input phosphor of the Image Intensifier is positioned 35 cm above the TT and the collimator is fully open. For automatic collimation, this unexposed border should be visible at all heights above the TT.
Source-to-Skin Distance
Fixed - Must not be < 15 in. (38 cm)
Mobile - Must not be < 12 in. (30 cm); Should not be < 15 in. (38 cm)
Half-Value Layer
The HVL must be measured annually and performed at initial installation by physicist
For Image Intensified Fluoro. a HVL of 3 - 4 mm Al is considered acceptable at kVp ranges of 80 - 100
Cumulative Timer
Purpose: To make the radiologist aware of the length of the fluoroscopy time
The audible/visual (or both) signal indicates a predetermined period of fluoro. beam-on time not to exceed 5 mins. Activated by exposure switch
The audible/visual (or both) signal should last at least 15 sec. at which time the timer must be manually reset.
Primary Protective Barrier (Direct Beam Absorber)
Permanently built into unit
The Image Intensifier Assembly (input phosphor) acts as a primary protective barrier (see lead equivalency below). It must be coupled with the x-ray tube and interlocked so that the fluoroscopic x-ray tube cannot be energized when the image intensifier is in the parked position. The barrier must ensure that the exposure dose does not exceed 2 mrad/hr at 10 cm from the image receptor. The exposure must terminate when the primary protective barrier is removed.
Lead Equivalency: 1.5 mm - Up to 100 kVp
Lead equivalency: 1.8 mm - > 100 kVp
Lead Equivalency: *2.0 mm - Up to 125 kVp (required for image intensifier unit)
C-Arm Fluoroscopy (Mobile)
0.5 mm lead Equivalent aprons must be worn by RT
Minimal Source-to-end-of-Collimator-Distance of 12 inches (30 cm)
Patient-to-I I-Distance (PID) should be as short as possible
Digital Fluoroscopy
Decreases patient dose
Beam-on time must be limited
Intermittent fluoro
Collimation
Pulse progressive System is used-Pulse of rad creates entire image on output phosphor (where electrons interact and produce light)
Last Image hold
Potential Radiation Responses of Skin from High Dose Fluoro: Early Transient Erythema
Approx Threshold Dose (rad)
200 (2Gy)
Approx Time of Onset
Hours
Potential Radiation Responses of Skin from High Dose Fluoro: Main Eythema
Approx Threshold Dose (rad)
600 (6Gy)
Approx Time of Onset
10 Days
Potential Radiation Responses of Skin from High Dose Fluoro: Temporary Epilation
Approx Threshold Dose (rad)
300 (3Gy)
Approx Time of Onset
3 weeks
Potential Radiation Responses of Skin from High Dose Fluoro: Permanent Epilation
Approx Threshold Dose (rad)
700 (7Gy)
Approx Time of Onset
3 weeks
Potential Radiation Responses of Skin from High Dose Fluoro: Moist Desquamation
Approx Threshold Dose (rad)
1500 (15 Gy)
Approx Time of Onset
4 weeks
Dose area product (DAP)
The quantity that reflects not only the patient radiation dose but also the amount of tissue irradiated.
DAP increases with increased radiation field size
Smaller size = lower DAP
DAP is usually positioned just after the xray beam collimator before the beam enters the patient
The DAP detector is a transmission ionization chamber, parallel plate type, larger than the largest area xray beam
Using DAP to monitor radiation intensity is a good way to implement radiation management procedures and keep patient radiation dose low.