1/36
Looks like no tags are added yet.
a form of energy; electromagnetic radiation or photons of energy emitted as waves
light
What unit is light measured in?
nanometers (10^-9)
What are the wavelengths associated with visible light, UV light, and infrared light?
UV light: 200-400nm
Visible light: 400-700nm
Infrared light: 760-3000nm
visible light
400-700nm
infrared light
760-3000nm
Which area of the visible light spectrum will provide the greatest absorbance for a substance?
A colored solution has the greatest absorbance in its *complimentary* color range
(Red with green, yellow with violet, blue with orange)
What are the two main principles of spectrophotometric methods?
1) Substances absorb light at specific wavelengths, therefore the wavelength absorption pattern can be the basis for the identification of an analyte (*each analyte has a specific wavelength*)
2) The amount of light absorbed is proportional to the amount of substance that is present (*absorption is directly proportional to substance present*)
What is the difference between absorbance and trasmittance?
Absorption is the amount of light taken up by a substance (directly proportional)
Transmittance is the light passed through a substance (inversely proportional)
Transmittance
is the light passed through a substance (inversely proportional)
What is Beer's Law and what does each variable represent?
*A=abc*
A is absorbance
a is absorptivity constant for the substance (constant for each substance)
b is the length of light path through a substance (constant set by spect. 1cm)
c is the concentration of the substance
Note: since a and b are constant, absorbance (A) is equal to c (concentration) which makes sense because they are directly proportional.
What is the formula based on Beer's Law that can be used to solve for an unknown concentration or absorbance?
A1/C1 = A2/C2
the extent to which a range over a measurement, such as absorbance, is directly proportional to the concentration
linearity
Absorbance vs. concentration is a (linear/non-linear) function.
A= linear
% Transmittance vs. concentration is a (linear/non-linear) function.
%T= non-linear
provides wavelengths of light in the visible spectrum or UV range
light source/lamp
What types of lamps are used for visible light and UV light?
Tungsten lamp for visible light
hydrogen deuterium or mercury-arc for UV
Newer specs have a laser bean
used to eliminate unwanted wavelengths of light and allow the desired light to reach the sample; use diffraction grating
monochromator
smaller bandwidth leads Tom
(in regards to resolution, sensitivity, intensity of light)
greater resolution, increased sensitivity, decreased intensity of light
A larger bandwidth leads to _________________.
decreased resolution, decreased sensitivity, increased intensity of light
Bandwidth:
the range of wavelengths exiting the slit is called the bandpass or bandwidth
holds the cuvette containing the test solution; may be round or square and either glass, plastic or quartz
sample holder
Which is better-- square or round sample holder?
square because curved bends the light
Sample holders may be either glass, plastic or quartz. What is each material used for?
glass and plastic used for visible light
quartz used for UV light
converts light energy to electrons to produce an electrical signal
photodetector
What is the most common photodetector? Why?
*photomultiplier tube* because of the rapid response and its ability to amplify the signal
read-out device that presents changes in electrical energy numerically; gives A or %T; may be linked to LIS or HIS/EMR for fast delivery of information to the provider
readout device/meter
gives power to the device; supply must be continuous and stable and required to have a back-up power supply such as batteries
power source
What is the purpose of blanking an instrument?
Zeroing and blanking the instrument set the limits of the instrument from no light transmitted to all light transmitted
a check of an instrument to ensure continued accuracy throughout the reportable range of the test results for the test system
calibration
a material of known value used to calibrate, or adjust a measurement procedure, or to compare the response obtained with the response of a test specimen and/or sample; are commercially prepared from manufacturer
calibrators
reference material whose concentration is determined by assay with a reference method
standard (used interchangeably with calibrator)
the ability of a photometric system to yield a linear relationship between the amount of radiant power absorbed by a substance and the concentration of that substance
linearity
Controls have
Controls have the same base matrix as the substance being tested
Calibrators are
manufactured to contain a specified amount of a pure analyte and adding it to an appropriate solvent for analysis (such as Grade A water), so they contain a known amount of analyte
What 4 situations would require a calibration?
1) Reagent lot number changed
2) Extensive maintenance performed on the instrument
3) Based on manufacturer's guidelines
4) If calibration fails
When does CLIA require calibration verification?
CLIA requires calibration verification every 6 months, after a major preventative maintenance repair, or whenever recommended by the manufacturer
Where are absorbance and concentration plotted on a graph?
A is Y axis
Conc is X axis