Vocabulary flashcards detailing ultrasound display formats (A-mode, B-mode, and M-mode), their respective axes, operational principles, and clinical applications.

22 Terms

1
New cards

Ultrasound Display Modes

The three basic formats used to present ultrasound data: Amplitude mode (A-mode), Brightness mode (B-mode), and Motion mode (M-mode).

2
New cards

A-mode (Amplitude Mode)

A display format that appears as a series of upward spikes; when a sound pulse is emitted and its reflection returns to the transducer, it is processed and displayed as an upward deflection.

3
New cards

A-mode Spike Height

The height of the upward deflection on an A-mode display, which is directly proportional to the amplitude (strength) of the returning echo; strong reflections produce tall spikes, while weak reflections produce short spikes.

4
New cards

X-axis of A-mode

The horizontal axis on an A-mode display, which represents reflector depth and is derived from the time-of-flight information of the sound pulse.

5
New cards

Y-axis of A-mode

The vertical axis on an A-mode display, which represents reflection amplitude (the strength of the reflected signal).

6
New cards

Ophthalmology (A-mode Use)

The primary clinical field where A-mode is utilized due to its high accuracy in determining the depth of anatomical reflectors.

7
New cards

Anechoic Area (A-mode Display)

An area with no spikes on an A-mode display, indicating the absence of returning echoes, characteristic of structures such as a simple cyst.

8
New cards

B-mode (Brightness Mode)

A display format that appears as a line of dots of varying brightness; when a reflection returns to the transducer, it is processed and an invisible dot is turned on.

9
New cards

Dot Brightness (B-mode)

The visual brightness of a dot that indicates echo strength; weaker reflections appear as darker gray dots, whereas stronger reflections appear as brighter white dots.

10
New cards
<p>Highest Amplitude Reflection (B-mode)</p>

Highest Amplitude Reflection (B-mode)

The reflection with the greatest amplitude wave (labeled Wave E), which produces the brightest white dot on a B-mode display.

11
New cards

X-axis of B-mode

The horizontal axis of B-mode (parallel to the transducer) that represents reflector depth, derived from time-of-flight information.

12
New cards

Z-axis of B-mode

The axis on an oscilloscope onto which the reflection amplitude information of B-mode is routed.

13
New cards

Y-axis of B-mode

An axis that does not exist on a standard two-dimensional (2D) B-mode display.

14
New cards

Gray Scale Imaging (Historical Basis)

B-mode was the first form of gray scale imaging and serves as the foundation for all other gray scale modalities, including real-time imaging.

15
New cards

M-mode (Motion Mode)

A display format appearing as a group of horizontal wavy lines; produced by moving photosensitive paper at a constant speed across a B-mode display to track changing reflector depths.

16
New cards

Moving Reflector Line (M-mode)

A wavy line moving up and down on the M-mode display, indicating that a reflecting surface is moving closer to and farther away from the transducer.

17
New cards

Stationary Reflector Line (M-mode)

A straight horizontal line on the M-mode display, indicating that a reflector is not moving relative to the transducer.

18
New cards

X-axis of M-mode

The horizontal axis on an M-mode display, which represents time.

19
New cards

Y-axis of M-mode

The vertical axis on an M-mode display, which represents reflector depth derived from time-of-flight information.

20
New cards

Sampling Rate of M-mode

A very high sampling rate that is directly equal to the pulse repetition frequency (PRF\text{PRF}) of the ultrasound system.

21
New cards

Cardiac Imaging (M-mode Use)

The primary clinical application of M-mode, used to evaluate cardiac walls and moving structures of the heart such as valve leaflets.

22
New cards

Distance Calculation (Display Modes)

A capability shared by all three ultrasound display modes (A-mode, B-mode, and M-mode) to determine anatomical measurements based on acoustic reflection data.