IMAGE FORMATION PULSE SEQUENCES MRI QUIZ

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/85

encourage image

There's no tags or description

Looks like no tags are added yet.

86 Terms

1
New cards

What is an MRI pulse sequence?

Timed instructions controlling how MRI images represent sampled tissue.

2
New cards

What creates the echo in gradient echo (GRE)?

Gradient coils refocus protons; an RF pulse and gradient field are used.

3
New cards

Which pulse is absent in GRE?

The 180° refocusing RF pulse.

4
New cards

Why is GRE more sensitive to field inhomogeneities?

It lacks a 180° RF refocusing pulse.

5
New cards

Why are GRE sequences generally fast?

They can use shorter TR because they lack 180° refocusing pulses.

6
New cards

What contrast can GRE produce?

T1 or T2* characteristics.

7
New cards

What can influence GRE contrast?

Susceptibility, chemical shift, and field inhomogeneities.

8
New cards

When does steady state occur in GRE?

When TR is shorter than T2 or T2*, leaving residual transverse magnetization.

9
New cards

Which GRE type provides high fluid signal?

T2* steady-state, coherent, or steady-state gradient echo.

10
New cards

What does spoiled GRE do?

Removes residual transverse magnetization before the next excitation.

11
New cards

How does RF spoiling work?

Additional or phase-shifted RF pulses spoil residual transverse magnetization.

12
New cards

What is BTFE?

A balanced steady-state sequence with high T2 signal and mixed T2/T1 contrast.

13
New cards

What are BTFE image characteristics?

High SNR and bright fluids; useful for cranial nerve evaluation.

14
New cards

What is GE's vendor name for BTFE-like imaging?

Fiesta.

15
New cards

What is Siemens' vendor name for BTFE-like imaging?

TrueFISP.

16
New cards

What is Hitachi's vendor name for BTFE-like imaging?

BASG.

17
New cards

What is Toshiba's vendor name for BTFE-like imaging?

True SSFP.

18
New cards

What is Philips' name for the sequence?

BTFE.

19
New cards

What is a common use of T1 spoiled GRE?

Dynamic organ imaging and contrast-enhanced MRA.

20
New cards

What is one example of T1 spoiled GRE acquisition?

SPGR coronal oblique imaging to evaluate the hippocampus in a patient with seizures.

21
New cards

What pulse begins conventional spin echo?

A 90° RF excitation pulse.

22
New cards

What does the 180° pulse do in spin echo?

Refocuses the decaying spin echo.

23
New cards

How many k-space lines does conventional spin echo acquire per TR?

One line.

24
New cards

How many k-space lines can fast spin echo acquire per TR?

Multiple lines, approximately 2 to 50 or more.

25
New cards

What is tau in spin echo?

The interval between the 90° and 180° RF pulses; half of TE.

26
New cards

How is a spin echo produced?

By a combination of two or more RF pulses.

27
New cards

What is dual-echo spin echo?

Two images per slice with different TEs but the same TR.

28
New cards

What weightings are commonly obtained with dual-echo spin echo?

Short TE: proton density; long TE: T2 weighting.

29
New cards

With TEs of 25 ms and 90 ms, which has more T2 contrast?

The second echo at 90 ms.

30
New cards

How does the second dual echo compare in SNR?

It has lower SNR than the first echo.

31
New cards

If 20 slices are planned for dual-echo imaging, how many images result?

40 images: one at each TE for every slice.

32
New cards

How does fast spin echo reduce scan time?

It collects several echoes and fills multiple k-space lines during each TR.

33
New cards

What is TSE factor?

Echo train length in fast spin echo; the term used by Philips.

34
New cards

If ETL increases by a factor of 3, how does scan time change?

It becomes about three times faster, as described in the document.

35
New cards

What are two fast spin echo trade-offs?

Lower SNR and increased patient SAR absorption.

36
New cards

What can increasing ETL do to image appearance?

Increase blurring.

37
New cards

What is effective TE in fast spin echo?

The echo encoded closest to the center of k-space (k0).

38
New cards

How are fast spin echo shots calculated?

Number of phase encodings divided by ETL.

39
New cards

What does a 'shot' mean in fast spin echo?

The number of non-redundant RF excitations.

40
New cards

What is single-shot imaging?

One excitation and an echo train fill the entire k-space matrix.

41
New cards

Why can single-shot imaging help in abdominal MRI?

It can reduce respiratory motion or improve temporal resolution.

42
New cards

What is multi-shot imaging?

Each excitation fills only part of k-space, allowing opportunity for higher spatial resolution.

43
New cards

What pulse begins inversion recovery (IR)?

A 180° inversion pulse followed by a 90° excitation pulse.

44
New cards

What is the purpose of the initial 180° IR pulse?

To null fat or fluid signal, depending on TI and field strength.

45
New cards

What is inversion time (TI)?

The interval between the 180° inversion pulse and the 90° excitation pulse.

46
New cards

What IR sequence types are listed?

STIR, FLAIR, and 3D-IR.

47
New cards

What TI relationship is given for nulling a tissue?

TI is approximately 69% of that tissue's T1 relaxation time.

48
New cards

What does STIR stand for?

Short TI inversion recovery.

49
New cards

At 1.5 T, what is the approximate STIR fat null point?

About 140–160 ms.

50
New cards

What STIR applications are listed?

Compression fractures, retro-orbital fat lesions, musculoskeletal contusions, and fat suppression.

51
New cards

What does FLAIR stand for?

Fluid-attenuated inversion recovery.

52
New cards

What TI is given to null CSF in FLAIR at 1.5 T?

Approximately 2000 ms.

53
New cards

What is T2-weighted FLAIR commonly used to evaluate?

Periventricular white matter.

54
New cards

What does SPGR stand for?

Spoiled Gradient Recalled echo.

55
New cards

How does SPGR spoil transverse magnetization?

It phase-shifts the RF excitation pulse to prevent residual transverse magnetization from accumulating.

56
New cards

Why does SPGR use a short TE?

To reduce T2 effects so T1 effects predominate.

57
New cards

What is Siemens' name for the SPGR-type sequence?

FLASH (Fast Low Angle SHot).

58
New cards

What is Philips' name for the SPGR-type sequence?

T1-FFE (Fast Field Echo).

59
New cards

What is Hitachi's name for the SPGR-type sequence?

SARGE (Steady state Acquisition Rewound Gradient Echo).

60
New cards

What is TR?

Time between two RF excitation pulses; described in the document as between two 90° pulses.

61
New cards

What is TE?

Time from the 90° RF pulse to the peak signal in the receiver coil.

62
New cards

What happens to PD weighting when TR lengthens and TE stays short?

Proton-density weighting increases.

63
New cards

What happens to available slices when TE increases?

The number decreases because fewer slices fit within TR.

64
New cards

What happens when TR increases?

SNR and the number of available slices increase, according to the document.

65
New cards

What is the basic scan-time formula?

TR × phase matrix × NEX gives milliseconds; divide by 1000 for seconds.

66
New cards

What scan-time formula includes ETL?

TR × matrix × NEX ÷ ETL.

67
New cards

What controls image contrast through pulse sequence timing?

The timing of RF pulses.

68
New cards

Why are spin echo sequences less sensitive to inhomogeneity?

Their 180° RF refocusing pulses help correct local field inhomogeneities.

69
New cards

What happens when flip angle is reduced?

Images contain less T1 information.

70
New cards

What determines scan plane and slice thickness?

The slice-selection gradient.

71
New cards

How can oblique slice orientations be acquired?

By using two gradients for slice selection.

72
New cards

Which gradient is active during echo production?

The frequency-encoding (readout) gradient.

73
New cards

What is echo-planar imaging (EPI)?

A very fast technique that can acquire an entire MR image in a fraction of a second.

74
New cards

Which sequence is described as the fastest commonly available?

Echo-planar imaging (EPI).

75
New cards

What is the basic spin echo pulse pattern?

90° RF followed by 180° RF.

76
New cards

What is the basic fast spin echo pulse pattern?

90° RF followed by a train of 180° RF pulses.

77
New cards

What is the basic inversion recovery pulse pattern?

180° inversion pulse followed by a 90° excitation pulse.

78
New cards

How does gradient echo create an echo?

A variable flip-angle RF pulse followed by gradient-based refocusing.

79
New cards

Which GRE settings are listed for T2* weighting?

TR 450 ms with a 30° flip angle.

80
New cards

Which GRE settings are listed for T1* weighting?

TR 45 ms with a 90° flip angle.

81
New cards

What happens to T2 information when TE is reduced?

T2 information decreases.

82
New cards

How does increasing flip angle affect SNR?

SNR increases up to the Ernst angle.

83
New cards

What changes tend to lengthen scan time?

Increase TR, decrease ETL, increase NEX, and decrease parallel-imaging factor.

84
New cards

What changes tend to shorten scan time?

Decrease TR, increase ETL, decrease NEX, enable half-scan, and increase parallel-imaging factor.

85
New cards

Which changes can improve SNR?

Increase TR, reduce phase matrix, increase FOV, narrow receive bandwidth, increase NEX/NSA, and reduce ETL.

86
New cards

Which changes can reduce SNR?

Decrease TR or pixel size/FOV, widen receive bandwidth, decrease NEX, or increase ETL.