Why do we need to verify lenses?
Laboratories make mistakes
The office makes mistakes in ordering
Both in combination
Verification for toric RGPs is even more important considering the complexity of the design
How are RGP lenses received?
Most labs / manufacturers ship RGPs in solution in either flat lens cases or vials
If shipped dry, should hydrate for about 24 hours before handling / verification and patient pick-up or delivery at appointment
What are the steps of RGP toric verification?
1) Determine BVP With Lensometer
2) Determine if There is Prism With Lensometer
3) Determine BOZRs with Radiuscope
4) Compare BVP Astigmatism with BOZR Astigmatism, to Determine BST, SPE, or CPE Bitoric
What is a lensometer?
A device that measures the power of a lens
How do you use a lensometer to measure BVP for RGPs?
Dry lens
Concave side down
Focus mires as for a spectacle lens
Remember, for bitoric lenses, record the powers in the two principal meridians (do NOT record in sphero-cyl form)
What is a radiuscope?
Measures the base curve(s) of the lens
What are the steps of RGP lens verification with a radiuscope?
Place your lens and 1 drop of solution of the appropriate lens holder
Focus a star
Zero your scale (at 0 or next closest whole number)
Turn your dial towards you
Star → Spiral → Star (if entire star is not clear, possibly toric or warped)
Take your final reading when the last star is clear
Read dial to the nearest 0.01mm
What is unique about BST, FST, and bitoric RGP verification?
Presence of cylinder in BVP indicates FST, BST, or bitoric
Presence of unequal BOZR on radiuscope indicates BST or bitoric
What does the ratio of lensometry and radiuscope measurement mean for BST?
Compare the cylinder on lensometry and radiuscope
BST will have a ratio of 3:2 cylinder reading in lensometer to radiuscope (for n = 1.49, PMMA material)
In modern lenses (lower n), this ratio is closer to 2.25:2
What does the ratio of lensometry and radiuscope measurement mean for SPE bitorics?
Compare the cylinder on lensometry and radiuscope
SPE bitorics will have a 1:1 ratio between radiuscope and lensometry cylinder
What does the ratio of lensometry and radiuscope measurement mean for CPE bitorics?
If the radiuscope and lensometry readings do not follow the previous ratios, and there is cylinder on the radiuscope and lensometry measurement, the lens is likely a CPE bitoric H
How can you identify an FST from lensometry?
If there is both prism and cylinder verified by lensometry, then the lens is very likely an FST
It can be verified by measuring a spherical BOZR
The axis can also be specified by orienting the prism to be base down
How do you verify diameter?
Measure with magnifying loupe or diameter gauge
Recreate the RGP verification flowchart?
Cylinder on Lensometry
Prism?
No
Measure two BOZR with radiuscope
Yes
FST
Confirm with spherical BOZR on radiuscope
3/2 or 2/25 relationship between lensometer cylinder and radiuscope cylinder?
No, so bitoric
Is lensometry cylinder equal to radiuscope cylinder?
Yes, so SPE bitoric
No, so CPE bitoric
Yes, so BST
Provide a summary of this lecture?
Lens verification is necessary to confirm what is ordered versus that is received
It is also necessary in some cases to determine what lens a patient is actually wearing