Artifact
An error in imaging.
Artifacts are incorrect representations of anatomy or function.
Echoes that are not representative of actual anatomic structures, or echoes of structures that exist but are not represented on the display monitor.
Artifact types include reflections that are:
not real, not seen on the image, not of the correct shape or size, not in correct position, not of correct brightness
Artifacts result from:
violation of assumptions, equipment malfunctions or poor design, the physics of ultrasound, operator error
Six Assumptions of Imaging Systems
1. Sound travels in a straight line
2. Sound travels directly to a reflector and back
3. Sound travels in soft tissue at exactly 1,540 m/s.
4. Reflections arise only from structures positioned in the beam's main axis
5. The imaging plane is very thin
6. The strength of a reflection is related to the characteristics of the tissue creating the reflection.
Artifacts appear in some views and disappear in others. True or false?
True.
Characteristics of reverberation
Appears in multiples
Appears as equally spaced
Located parallel to the sound beam's main axis
Located at ever-increasing depths
What do reverberations resemble?
A ladder or Venetian blind
Reverberation
Appear on the display as multiple, equally spaced echoes caused by the bouncing of the sound wave between two strong reflectors positioned parallel to the ultrasound beam.

Comet Tail artifact
AKA Ring Down Artifact
Often described as "reverberation with the spaces squeezed out," appears as a solid hyperechoic line directed downward.
It is created when closely spaced reverberations merge. More likely to appear when the reflecting surfaces are located in a medium with a very high propagation speed, such as a mechanical heart valve. Can also arise from the resonance, or vibration, of small structures such as gas bubbles after having been bombarded by an ultrasound pulse.

Characteristics of Comet Tail
Appears as a single long hyperechoic echo
Located parallel to the sound beam's main axis
Shadow Artifact
Appears as a hypoechoic or anechoic region extending downward from a very strong attenuating medium. Same color as the image background. When sound cannot pass through a structure, deeper reflecting surfaces are absent on the image. Appear when the attenuation is higher in the tissue above the shadow than in the surrounding tissue.

Characteristics of Shadowing
Hypo- or anechoic (background color)
The result of too much attenuation
Located beneath the structure with abnormally high attenuation
Prevents visualization of true anatomy on the scan.
Edge Shadow Artifact
AKA
Shadowing by Refraction
A special form of shadowing that appears as a hypoechoic region extending along the edge of a curved reflector. Prevents the display of true anatomic structures that are positioned within the extended hypoechoic region. The sound beam refracts (transmits with a bend) at the edge of a curved reflector. The beam simultaneously diverges, resulting in a drop in intensity.

Characteristics of Edge Shadowing
Hypo- or anechoic (background color)
Results when the beam spreads after striking a curved reflector
Extends downward from the curved reflector's edge, parallel to the beam
Prevents visualization of true anatomy on the scan
Enhancement Artifact
Appears as a hyperechoic region beneath tissues with abnormally low attenuation. Hyperechoic regions are the same color as the foreground (or echoes) of the image. The opposite of shadowing.

Characteristics of Enhancement
Hyperechoic (appears the same color as the foreground color)
The result of too little attenuation
Located beneath the structure with abnormally low attenuation
Focal Enhancement Artifact
AKA
Focal Banding
A special form of enhancement in which a side-to-side region of an image appears brighter than tissues at other depths. Most prominent at the focus. When a beam is strongly focused, the intensity in the focal zone is increased and reflections from the focal zone may be abnormally intense. Therefore, structures at the focus appear brighter than those at other depths. An entire horizontal region of tissue appears hyperechoic.

Characteristics of Focal Enhancement
A hyperechoic side-to-side region (appears the same color as the foreground color)
The result of increased intensity at the focus.
Mirror Image Artifact
Created when sound reflects off of a strong reflector, and is redirected toward a second structure. This redirection causes a replica, or a second copy, of the structure to incorrectly appear on the image. The artifact is placed deeper than the real structure.

Characteristics of Mirror Image
A second copy of a true reflector
The copy appears deeper than the true reflector
A bright reflector, the mirror, lies on a straight line between the artifact and the transducer.
True reflector and artifact are equal distances from the transducer
Crosstalk Artifact
AKA
Spectral Mirroring
A mirror image artifact that appears on a spectral Doppler display.

Speed Error Artifact
AKA Propagation Speed Errors
AKA Range Error Artifact
Created when a sound wave propagates through a medium at a speed other than that of soft tissue (1.54 km/s). The correct number of reflectors are displayed, but they appear at incorrect depths.

Characteristics of Speed Error Artifact
Correct number of reflectors
Improper depth
Appears as a step off
Lobe Artifact
Appear when sound energy is transmitted in a direction other than along the beam's main axis. Degrades lateral resolution.
Characteristics of Lobe Artifact
A second copy of the true reflector.
The artifact and the true reflector are located side by side at the same depth.
Side Lobes Artifact
Created by a single crystal transducer, such as a mechanical probe.
Grating Lobes Artifact
Created by array transducers. Reduced by dividing each PZT element into small pieces - a process called subdicing. Can also reduce by apodization- exciting the subdiced elements at different voltages.
Refraction Artifact
Occurs when a sound pulse changes direction during transmission. Occurs when a sound wave strikes a boundary obliquely and the media on either side of the boundary have different propagation speeds. Transmission with a bend. Degrades Lateral resolution

Characteristics of Refraction artifact
A second copy of the reflector.
The copy is side by side, or at the same depth as the true reflector.
Slice Thickness Artifact
(Partial Volume Artifact
Elevational Resolution
Section thickness
Volume averaging)
Related to the dimension of the beam perpendicular to the imaging plane. True reflector lies outside of the assumed imaging plane. Fills in hollow structures such as cysts. Appears when the beam dimension is greater than the reflector size.

Elevational resolution
Determined by the thickness of the imaging plane.
Lateral Resolution Artifact
Occurs when a beam is wider than the distance between two reflectors that are located side by side or perpendicular to the beam. The two objects appear as one reflection on the image. May display a small reflector as a wide line. Least likely to occur at the focus of the sound beam, where the beam diameter is the smallest.
Axial Resolution Artifact
Created when a long pulse strikes two closely spaced structures, where one is in front of the other, or parallel to the main axis of the beam. Only one reflection will appear on the image if the structures are closer together than one-half the spatial pulse length.
Multipath Artifact
Created when sound pulses glance off a second structure on the way to or from the primary reflector. As a result, the transmit path length differs from the return path length.
Temporal resolution artifact
Low frame rates. Less accurate positioning of reflectors in motion.
Acoustic Speckle
The granular appearance of images that is caused by the interference of echoes from the distribution of scatters in tissue. Arises from small amplitude sound waves interfering with each other.

Aliasing
The appearance of Doppler spectral information on the wrong side of the baseline. Caused by under sampling of the Doppler shifts.

Methods of Reducing or Eliminating Aliasing
Shift the baseline
Increase the PRF
Increase the Doppler angle
Use a lower operating frequency
Use a continuous wave device
Nyquist Limit
Half the PRF
Range Ambiguity
Occurs when a pulse is emitted before all the echoes from the previous pulse have been received. When this happens, early echoes from the last pulse are received simultaneously with late echoes form the previous pulse. The system is unable to determine whether an echo is an early one from the last pulse or a late one from the previous one. Can be caused by a high PRF