Types of Digital Radiolography System
1) Computer Radiography (CR)
2) Digital Radiography (DR)
Computed Radiography (CR) (Basic info)
Used photostimulable imaging plates (Pst, IP), inside cassette which is then procedded
Digital Radiography (DR) (types)
Direct conversions: No scientillator, ONLY photodector
Indirect Conversions: uses CR and DR equipment. Includes a charged-coupled device or amorphous silicon with a film transistor.
Digital Formation formation
Electronic devices that produce radiographic images use analog language. However computers use binary language and would therefore need to be processed over to binary using:
ADC
ADC (Analog-To-Digital) (Steps)
1) Sampling
2) Quantification
Sample (ADC)
First step
Analog values measureed at a chosen frequency on an analog wave form
Quanification (ADC)
Second step
Analog values assigned a discrete analog value
Computer memory/power expressed in total bytes
Megabytes
Gigabytes
Terabytes
etc.
Matrix
Made up of pixels and voxels
Field of view (FOV)
Spatial resolution is dependent of this (size)(greater the matrix size, the better the spatial resolution)
Pixel
Picture element
Inversely related to spatial resolution
Sample frequency (pixel/mm)
Has bit depth
Pixel dependent on
Matrix size
Image receptor
Size
Bit depth
How much gray scale is it capable of showing
Greater bit depth = increases the size of the data set/volume of data for the image
EQUATION:
2^(bit depth)
DICOM
Digital Imaging and Communications in Medicine:
Standard for imaging pictures
Voxel
Most important for CT or MRI
The volume of elements
Image File Size (Affected by)
Pixel size
Matrix
Bit depth shades
Image Processing information (types)
Point Processing operation
Local Processing operations
Geometric Processing Operations
Point Processing Operations (Includes what)
Formation of histograms (histogram equalization)
Look-up table & windowing (for contrast and brightness)
LUT (Look-up-Table)
Adjustments to image contrast
Values assigned to data points (pixels) in the histogram
Produces contrast. Look, according to the reference contrast for exam view
Varys with exam/manufactures
Can NOT be compensated for with LUTs
Image display can be adjusted from the LUT
Adjustment similar to changing the D(log)E curve of film emulsion.
Histogram
Shape depends on anatomy/exposure
Used to determine pixel values of interest (VOI) and exposed
Arranges signals from min to max values
Graphical computation of signal values of a data set
VOI
Values of interest
Histogram Equalization
Rescaling if overexposed or underexposed to appear like a good histogram
Wide histogram = higher contrast
Narrow histogram = low contrast
If higher comparative # = bright/light
If lower comparative # = dark/dim
ACR
American college of radiology
Forms standard procedure for histogram radiation exposure
Local Processing Operations (
Kernal
Processing which mathematical calculation are applied to pixels until all pixels are subjected
Local processing operations (include)
High-pass filtering
Low-pass filtering
Spatial location filtering
High-pass filtering (local processing operations)
Removes low spatial frequency and produce sharper image (edge enhancing)
Low-pass filtering (local processing operations)
Intentionally blurs (reducing noise)
Should NOT be viewed in normal or high pass mode.
Spatial location filtering (local processing operations)
Kernel applied repeatedly to each pixel in the matrix to weight values or apply a coefficient across the matrix.
Geometric Processing Operations
Changing the position or orientation of pixels
Allows rotation/magnification
Digital Image quality (Types of resolution)
Spatial resolution
Density resolution
Contrast resolution
Spatial resolution
Matrix size increases, pixel size decrease, spatial resolution increases
Density resolution
Gray scale bit depth increases, sensity resolution increases
Contrast reolution
Low contrast resolution (LCR)
Ability to represent small energy values in data set
CR/DR receptors hav excellent LCR (great dynamic range)
Noise (types)
Electronic (system) noise
Quantum noise
Ambient noise
Electronic (system) noise
Results from undesirable signals from the digital system
Random background info detected. Does NOT contribute to image quality
Quantum Noise
Insufficient quantity of photons from improperly put exposure levels (grainy images)
Ambient Noise
Background radiation
SNR
Signal-to-Noise
To measure noise. A higher value correlates to little noise.
Image noise has an inverse relationship to contrast. Therefore, the more contrast, the less noise.
DQE
Detective Quantum Efficiency
Detective Quantum Efficiency (DQE)
Measure sensitivity and accuracy to which IR converts input to output.
Detective Quantum Efficiency (DQE) (Affected how)
SNR
Quantum noise
System noise
Exposure Index (EI)
Calculated using histogram values and pixel values of interest (PVOI) mid points
Varies between vendors
Responsible for understanding and applying EI values (Assess lmage quaility and ALARA compliance)
No universal system
ALARA
As low as reasonably achievable
Deviation Index (DI)
Created by AAPM and IEC
Standardizes exposure values between manufacturers using Elโs
Uses department ELT values
Comparison between acquired exposure (KIND) and target (KTGT)
Expressed as a scale of DI values